10 series Blood Pressure Monitor
Model BP786CANN
Instruction Manual
2298848-9B

Advertisement

Table of Contents
loading

Summary of Contents for Omron BP786CANN

  • Page 1 English Français 10 series Blood Pressure Monitor Model BP786CANN Instruction Manual 2298848-9B...
  • Page 2 Product includes: El producto incluye: AC Adapter Adaptateur CA ComFit™ Cuff Brassard ComFit™ Monitor Tensiomètre Battery Set Juego de Pilas BP786CANN Instruction Manual Quick Start Guide Guide de l’utilisateur Guide de démarrage rapide...
  • Page 4: Table Of Contents

    Contents Thank you for purchasing the OMRON ® P786CANN Blood Pressure Monitor. Your new blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood movement through your brachial artery and converts the movements into a digital reading. An oscillometric monitor does not need a stethoscope so the monitor is simple to use.
  • Page 5: Important Safety Information

    Important Safety Information Warning: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. (General Usage) DO NOT adjust medication based on measurement results from this blood pressure monitor. Take medication as prescribed by your physician. Only a physician is qualified to diagnose and treat High Blood Pressure. The monitor is not intended to be a diagnostic device.
  • Page 6 Important Safety Information Caution: Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property. (General Usage) Always consult your physician. Self-diagnosis of measurement results and self-treatment are dangerous. Consult your physician before using the device for any of the following conditions: •...
  • Page 7 Wipe the dust off from the power plug. Unplug monitor when not in use. Disconnect the power plug before cleaning. Use only an OMRON AC adapter designed for this device. Use of unsupported adapters may damage and/or may be hazardous to the device. (Battery Usage) Do not insert the batteries with their polarities incorrectly aligned.
  • Page 8: Know Your Device

    1. Know Your Device Open the rear cover page to read the following: The letter identifiers on the rear cover page correspond to those in the body of this page. Monitor Display Air jack Memory symbol Blood pressure color indicator USER ID symbol Connection button Average value symbol...
  • Page 9: Display Symbols

    1. Know Your Device 1.1 Display Symbols Irregular Heartbeat Symbol ( Normal Heartbeat When the monitor detects an irregular rhythm two or more times during the Pulse measurement, the irregular heartbeat symbol will appear on the display with Blood pressure the measurement values.
  • Page 10: Before Taking A Measurement

    1. Know Your Device Blood Pressure Color Indicator If your Systolic Blood Pressure is 135 mmHg or above and/or the Diastolic Blood Pressure is 85 mmHg or above, the blood pressure color indicator will light in “orange” when the measurement result is displayed. If the measurements are within the standard range, the blood pressure color indicator will light in “green”.
  • Page 11: Preparation

    2. Preparation 2.1 Battery Installation Remove the battery cover. Insert 4 “AA” batteries as indicated in the battery compartment. Replace the battery cover. Notes: • When the low battery symbol ( ) appears on the display, turn the monitor off, then replace all batteries at the same time.
  • Page 12: Using The Ac Adapter

    AC Adapter is disconnected. The measurement results will not be deleted. Use only OMRON authorized parts and accessories. Parts and accessories not approved for use with the device may damage the device.
  • Page 13 2. Preparation Insert the AC adapter plug into the AC adapter jack on the rear side of the monitor. Plug the AC adapter into an electrical outlet. To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first and then remove the AC adapter plug from the monitor.
  • Page 14: Setting The Trureadtm Mode

    2. Preparation 2.3 Setting the TruRead Mode The TruRead Mode takes 3 consecutive measurements. The monitor will inflate, take a measurement, and deflate - 3 times, separated by a short interval between each measurement. The TruRead Mode is set “oFF” by default. Select your USER ID (1 or 2).
  • Page 15: Setting The Date And Time

    2. Preparation Open the rear cover page to read the following: The letter identifiers on the rear cover page correspond to those in the body of this page. 2.4 Setting the Date and Time The date and time will be automatically set when paired with your smartphone. To set manually, follow the instructions below.
  • Page 16: Using The Device

    • The blood pressure can differ between the right arm and the left arm, and the measured blood pressure values can be different. OMRON recommends to always use the same arm for measurement. If the values between both arms differ substantially, please check with your physician which arm to use for your...
  • Page 17: How To Sit Correctly

    3. Using the Device 3.2 How to Sit Correctly To take a measurement, you need to be relaxed and comfortably seated, at a comfortable room temperature. Avoid bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement. •...
  • Page 18: Taking A Measurement

    3. Using the Device 3.3 Taking a Measurement Notes: • To stop a measurement, press the START/STOP button once to deflate the arm cuff. • Remain still and quiet while taking a measurement. The monitor is designed to take measurements and store the measurement values in the memory for 2 people using USER ID 1 and USER ID 2.
  • Page 19: Using The Truread Tm Mode

    3. Using the Device Remove the arm cuff. Press the START/STOP button to turn the monitor off. The monitor automatically stores the measurement result in its memory. It will automatically turn off after 2 minutes. Note: Wait 2-3 minutes before taking another measurement. Waiting between measurements allows the arteries to return to the condition prior to taking a measurement.
  • Page 20 3. Using the Device The first measurement is complete, waiting to start the next measurement. The interval will appear on the display. The second measurement is complete, waiting to start the next measurement. The interval will appear on the display. The third measurement is complete, the average for the 3 measurements appears on the display.
  • Page 21: Using The Guest Mode

    3. Using the Device Using the Guest Mode The monitor stores measurement values for 2 users in the memory. The guest mode can be used to take a single measurement for another user. No measurement values are stored in the memory, and the TruRead mode is not available when the guest mode is selected.
  • Page 22: Using The Memory Function

    3. Using the Device 3.4 Using the Memory Function The monitor automatically stores the results up to 100 readings for each user (1 and It can also calculate an average value based on the last 3 measurement values taken within 10 minutes. Notes: •...
  • Page 23 3. Using the Device Press the button to view the values stored in the memory. : To view the older values : To view the more recent values Note: The value with the TruRead symbol is the average for the consecutive 3 measurements. To display the individual measurement values, press the button while the average is displayed.
  • Page 24 3. Using the Device To View the Average Value Select your USER ID (1 or 2). Press and hold the button for more than 3 seconds. Notes: • If the previous measurement was taken without setting the date and time, the average value is not calculated.
  • Page 25 3. Using the Device To Pair This Device With a Smartphone To begin, download and install the free "Omron Wellness" App onto your smartphone. Open the App on your smartphone and follow set-up and pairing instructions. Note: Please make sure that you do not currently have an active reading screen on. If your reading screen is on, press the START/STOP button to clear the screen to allow you to begin to transfer your readings.
  • Page 26 3. Using the Device Open the "Omron Wellness" App on your smartphone and follow the device pairing instructions shown on your smartphone, while the connection symbol is flashing on the device’s display. Confirm if the device is connected successfully. When the device is connected successfully to the smartphone, “OK”...
  • Page 27 Clear your reading screen by pressing the "START/ STOP" button on your monitor. Next, tap the "transfer" icon in the Omron Wellness app to start the transfer. * If the data transfer is not starting, refer to “Connection failure. Data cannot be sent.”...
  • Page 28 3. Using the Device TO DELETE ALL RESULTS AND SETTINGS To delete all the information stored in your monitor, follow the instructions below. Make sure that the monitor’s power has been off. While holding the button down, press the START/STOP button for more than 5 seconds.
  • Page 29 3. Using the Device To Delete All the Values Stored in Memory The values stored in the memory are deleted by USER ID. Select your USER ID (1 or 2). Press the Memory button, while the memory symbol ( ) appears.
  • Page 30: Turning "Off" The Bluetooth

    3. Using the Device 3.5 Turning “OFF” the Bluetooth ® ® Turn off the Bluetooth of the monitor in the following areas where use of wireless equipment is prohibited. • On aircraft • In hospitals •While abroad Make sure that the power has been off before starting the instructions below. Press and hold the connection button for more than 10 seconds.
  • Page 31 3. Using the Device ® To turn “ON” the Bluetooth Make sure that the power has been off before starting the instructions below. Press and hold the connection button for more than 10 seconds. Release the button when “ON” appears on the display. Press the START/STOP button to turn the monitor off.
  • Page 32: Error Messages And Troubleshooting

    4. Error Messages and Troubleshooting 4.1 Error Messages Display Cause Solution Remove the arm cuff. Wait 2-3 minutes and then take another measurement. Irregular heartbeats are detected. Repeat the steps in section 3.3. If this error continues to appear, contact your physician. Carefully read and repeat the steps in section Movement during measurement.
  • Page 33 4. Error Messages and Troubleshooting Display Cause Solution Insert the air plug securely. Air plug disconnected. Refer to section 3.1. Apply the arm cuff tighter. Arm cuff is applied too loosely. Refer to section 3.1. Replace the arm cuff with a new one. Air is leaking from the arm cuff.
  • Page 34: Troubleshooting

    Data cannot be sent. Refer to "To Pair This Device With a Smartphone" in section 3.4 or visit the "Help" section in the Omron Wellness app for further help in pairing your smart device. The blood pressure monitor is not ready, follow the step “1.
  • Page 35: Maintenance And Storage

    5. Maintenance and Storage 5.1 Maintenance To protect your device from damage, please observe the following: • Store the device and the components in a clean, safe location. • Do not use any abrasive or volatile cleaners. • Do not wash the device and any components or immerse them in water. •...
  • Page 36: Storage

    5. Maintenance and Storage 5.2 Storage Unplug the air plug from the air jack. Gently fold the air tube into the arm cuff. Note: Do not bend or crease the air tube excessively. Do not store the device in the following situations: •...
  • Page 37: Optional Medical Accessories

    5. Maintenance and Storage 5.3 Optional Medical Accessories Arm Cuff AC Adapter Arm circumference 9” - 17” (22 - 42 cm) HEM-ADPTW5 ComFit Cuff (CFX-WR17) (Model: HEM-FL31)
  • Page 38: Specifications

    6. Specifications BP786CANN HEM-7321T-CA Model Display LCD digital display Measurement range Pressure: 0 to 299 mmHg Pulse: 40 to 180 beats / min. Accuracy Pressure: 3 mmHg or 2% of reading Pulse: 5% of display reading Inflation Fuzzy-logic controlled by electric pump...
  • Page 39: Fcc/Ic Statement And Trademarks

    7. FCC/IC Statement and Trademarks FCC CAUTION Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. Note: This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules.
  • Page 40 The Bluetooth Smart word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by OMRON HEALTHCARE Co.,Ltd. is under license. Other trademarks and trade names are those of their respective owners. Apple, the Apple logo, iPad, and iPhone are trademarks of Apple Inc., registered in the U.S. and other countries.
  • Page 41: Limited Warranty

    8. Limited Warranty Your BP786CANN Automatic Blood Pressure Monitor, excluding the arm cuff and batteries are warranted to be free from defects in materials and workmanship appearing within 5 years from the date of purchase, when used in accordance with the instructions provided with the monitor. The arm cuff is warranted to be free from defects in materials and workmanship appearing within 1 year from the date of purchase when the monitor is used in accordance with the instructions provided with the monitor.
  • Page 42: Guidance And Manufacturer's Declaration

    • The MEDICAL ELECTRICAL EQUIPMENT BP786CANN including AC-adapter needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in this documentations. • The Essential Performance of the BP786CANN including AC-adapter is to measure a blood pressure and a pulse rate and using the memory function.
  • Page 43 Guidance and manufacturer’s declaration - electromagnetic immunity OMRON BP786CANN including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON BP786CANN including AC-adapter should assure that it is used in such environment.
  • Page 44 Guidance and manufacturer’s declaration - electromagnetic immunity OMRON BP786CANN including AC-adapter is intended for use in the electromagnetic environment specified below. The customer or the user of this OMRON BP786CANN including AC-adapter should assure that it is used in such environment.
  • Page 45 OMRON BP786CANN including AC-adapter OMRON BP786CANN including AC-adapter is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of this OMRON BP786CANN including AC-adapter can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the OMRON BP786CANN including AC-adapter as recommended below, according to the maximum output power of the communications equipment.
  • Page 48 Manufactured for/Fabriqué pour : OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN Distributed by/Distribué par : OMRON HEALTHCARE, INC. 1925 West Field Court Lake Forest, IL 60045 U.S.A. omronhealthcare.ca © 2015 OMRON HEALTHCARE, INC. Made in China/Fabriqué en Chine...