Care After Use In Nitrous Oxide; Table 2-1: Oxygen Reading Error In A Mixture Of Anesthetic Gas - Teledyne MX300-I Operating/Service Instructions Manual

Portable oxygen monitor
Hide thumbs Also See for MX300-I:
Table of Contents

Advertisement

Portable Oxygen Monitor

Table 2-1: Oxygen Reading Error in a Mixture of Anesthetic Gas

(Balance: Mixture of 30% O
Gas or Vapor
Helium
Nitrous Oxide
Carbon Dioxide
Halothane
Enflurane
Isoflurane
Sevoflurane
Desflurane
* Errors are approximate and may vary based on exposure times and
concentrations.
These performances meet or exceed the requirements of ISO 7767:
1997 (E).
CAUTION:
THE MX300-I SHOULD NOT BE USED IN THE
PRESENCE OF FLAMMABLE ANESTHETICS SUCH
AS DIETHYL ETHER OR CYCLOPROPANE.
CAUTION:
THE MX300-I, OXYGEN SENSOR AND ASSOCIATED
HARDWARE ARE NON-STERILE DEVICES. DO NOT
AUTOCLAVE THE INSTRUMENT OR SENSOR, AS
THIS WILL DAMAGE THE EQUIPMENT.
2.3.5.2 C
A
ARE
FTER
CAUTION:
THE R17MED SENSOR SHOULD NOT BE LEFT IN
NITROUS OXIDE MIXTURES ANY LONGER THAN
ABSOLUTELY NECESSARY.
Teledyne Analytical Instruments
Gas or Vapor Level
/ 70% N
2
Test Level
50%, balance O
2
80%, balance O
2
10%, balance O
2
4%
5%
5%
5%
15%
U
N
O
SE IN
ITROUS
XIDE
Operation
O, except where noted)
2
Oxygen Reading
Error
0%
0%
0%
< 1.5% O
*
2
< 1.5% O
*
2
< 1.5% O
*
2
< 1.5% O
*
2
< 1.5% O
*
2
29