-150V to +150V output
DC to 5MHz @ -3dB large signal
bandwidth
2000V/µ µ µ µ s slew rate typical
±300mA Output current limit
Stable with all capacitive loads,
generates no overshoot
Short-circuit protected output
For MEMS devices, EO-
modulators, PZT (piezo) positioning,
beam steering, ultrasonics, ...
About this manual
This user manual is an integral part of the Falco Systems
WMA-300 amplifier product. Please read it carefully and
pay attention to the recommendations and the instructions
for safe use.
General description
The Falco Systems WMA-300 is a high speed, high-
voltage
linear
laboratory
applications. Its wide bandwidth and large voltage range
make it an excellent choice for use with MEMS devices,
EO-modulators, PZT positioning systems, beam steering,
ultrasonics, dielectric studies, and many others.
Through the extensive use of computer modelling and
state-of-the-art design concepts, this amplifier can be used
in many applications, including those where conventional
amplifiers are difficult to apply. It is designed to be
absolutely stable and free of spurious signals with any
capacitive load.
The amplification is 50x (fixed). With a range of –150V to
+150V, a large signal bandwidth of 5MHz @ -3dB and a
typical slew rate of 2000V/µs, this amplifier will enable
many
measurements
in
frequency response is essentially flat up to 2MHz.

USER MANUAL

amplifier
for
demanding
different
applications.
DC - 5MHz High Voltage Amplifier WMA-300
A fast current limit of ±300mA (typical) and a fast short-
circuit protection make this amplifier suitable for both the
normal daily laboratory use (and abuse), and automated
measurement systems.
Safety
- This product is able to produce over 150V at
!
more than 300mA at its output, which is a
dangerously high level (risk of electric shock). Safety
measures should be taken accordingly.
- Only use this appliance with a mains connection with
protective earth
- The internal circuitry of the amplifier operates at high
voltage. Only qualified personnel from Falco Systems
should service this amplifier.
- When the amplifier is turned on or off, a short voltage
spike may appear at the output which may damage
circuitry already connected to it.
- Replace fuses with 250V 630mA 5x20mm slow blow
rated fuses only.
- The Falco Systems WMA-300 is only suitable for indoor
use in a class II environment (domestic, light industrial).
- This product should only be cleaned with a soft, slightly
moist cloth. Unplug the WMA-300 from the mains power
Its
before cleaning.
www.falco-systems.com

Advertisement

loading
Need help?

Need help?

Do you have a question about the WMA-300 and is the answer not in the manual?

Questions and answers

Summary of Contents for Falco Systems WMA-300

  • Page 1: User Manual

    About this manual circuit protection make this amplifier suitable for both the normal daily laboratory use (and abuse), and automated This user manual is an integral part of the Falco Systems measurement systems. WMA-300 amplifier product. Please read it carefully and pay attention to the recommendations and the instructions for safe use.
  • Page 2 Never apply more than +30V (-30V) to the amplifier input! Output protection The WMA-300 has been designed to be fully stable with all capacitive loads. It has been optimized for its step- response, but is also a very good linear and sine-wave amplifier.
  • Page 3 Amplifier characteristics In the following pages, several amplifier characteristics are shown: Frequency response (Fig. 2, 3) Square wave response (Fig. 4, 5, 6) Slew rate (Fig. 7) Capacitive load dependency square waveoutput (Fig. 8) Triangle and sine response (Fig. 9, 10) Noise and offset (Fig.
  • Page 4 300V step up in 125ns 300V step down in 121ns Figure 9. Triangle wave 300V 1kHz 0V to 150V in 123ns 150V to 0V in 120ns Figure 7. Slew rate (definition: step response from 10% to 90% signal). 240V in 125ns ≈ 2000V/µs 150pF Figure 10.
  • Page 5 WEEE Do not use non-sinusoidal mains power generators, indoor use only Do not dispose of the WMA-300 as standard waste, but Operating temperature: 15 – 30°C bring it to a WEEE electronic waste collection point. Storage temperature: 0 – 60 °C Relative humidity: 30 –...