5.Handling Errors and Problems

5. Handling Errors and Problems
5.1

Error Messages

Error Display
Blinks or
appears
continuously
24
Cause
Cuff is under inflated.
Movement during
measurement
Air tube disconnected.
Arm cuff not applied
correctly.
Clothing is interfering
with the arm cuff.
Air is leaking from the
arm cuff.
The arm cuff was
inflated above
299 mmHg.
Battery power is low.
Remedy
Press air release button and
restart the measurement with a
higher inflation level. Carefully
read steps under section 3.3.
Repeat measurement. Remain
still and do not talk during
measurement.
Refer to section 3.3.
Insert the air tube securely.
Refer to section 3.2.
Apply the arm cuff correctly.
Refer to section 3.2.
Remove any clothing
interfering with the arm cuff.
Refer to section 3.2.
Replace cuff with new one.
Refer to Chapter 7.
Release your hand from the air
inflation bulb before the
pressure reaches 299 mmHg.
Refer to section 3.3.
Replace all four "AAA"
batteries with new ones.
Refer to section 2.1.