(ocsi) oxygen concentrator

Advertisement

USER'S GUIDE
Mark 5 Nuvo 8
(OCSI)

OXYGEN CONCENTRATOR

[Original language is English]
Federal Law (US) restricts this device to sale
by, or on the order of, a licensed physician.
This oxygen concentrator should be used
only under the supervision of a licensed
physician.
GLOSSARY OF SYMBOLS..........................................1
GENERAL SAFETY GUIDELINES ............................2
I.
DESCRIPTION .......................................................2
I.1. Front panel (Fig. I. 1) ........................................... 3
I.2. Rear panel (Fig. I. 2)..................................................3
II. STARTING-UP / INSTALLATION .......................3
II.1. Use in direct oxygen therapy.....................................3
III. CLEANING-MAINTENANCE .............................4
III.1.Cleaning ....................................................................4
III.2.Everyday disinfection....................................................4
III.3.Maintenance..............................................................4
: ON (power switched on)
: Off (power switched off)
: Type B device
: Class II protection
: Do not expose to open  ames
2010-2238CE Rev C
0413: Complies with the 93/42/ EEC directive
certi ed by the approved organization no 0413.
Danger: Do not smoke when using oxygen or when
near this device.
CONTENTS
IV.USEFUL INFORMATION ......................................... 4
IV.1. Accessories and spare parts....................................... 4
IV.3. Operating principles .................................................. 5
IV.4. Alarms - Safety devices ............................................ 5
IV.5. Oxygen Monitor (OCSI) function ........................... 5
IV.6. Technical characteristics ........................................... 5
IV.7. Standards ................................................................... 6
IV.8. Method for disposing of waste.................................. 6
IV.9. Method for disposing of device................................. 6
IV.10.Troubleshooting......................................................... 7
V. EMC, Electromagnetic statements...........................8-9

GLOSSARY OF SYMBOLS

: Do not use oil or grease
: Technical information
: Consult the accompanying documents
: Keep in the vertical position
: Fragile - handle with care
: Oxygen concentration warning light
May 2011
1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the OCSI and is the answer not in the manual?

Questions and answers

Summary of Contents for Nidek Medical OCSI

  • Page 1: Table Of Contents

    IV.3. Operating principles ..........5 IV.4. Alarms - Safety devices ..........5 I.1. Front panel (Fig. I. 1) ........... 3 IV.5. Oxygen Monitor (OCSI) function ......5 I.2. Rear panel (Fig. I. 2)..........3 IV.6. Technical characteristics ........... 5 II. STARTING-UP / INSTALLATION .......3 IV.7.
  • Page 2: General Safety Guidelines

    Mark 5 Nuvo 8 with the accessories recommended by move away from it must be determined to suit the Nidek Medical Products, Inc. surrounding noise level. May 2011 2010-2238CE Rev C...
  • Page 3: Front Panel (Fig. I.1)

    Flow adjustment knob (l/min) Note: After turning the unit off, Circuit Breaker exceeds the set point. OCSI (Oxygen Concentration the user must wait 3-5 minutes before turning it back Status Indicator) Indicator Lights on. System pressure must dissipate before the unit will properly restart.
  • Page 4: Cleaning-Maintenance

    II.2. Daily disinfection The accessories with a Nidek Medical part number reference, Because there is a  nal product  lter inside the device, daily or included in the set of accessories supplied with the device, disinfection concerns only the external oxygen therapy comply with these requirements.
  • Page 5: Iv.3. Operating Principles

    IV.3. Operating principle IV.5. OCSI (oxygen concentration status indicator) func- tion The compressor sends  ltered ambient air to an electronic valving system, which allows compressed air to pass to the IV.5.1. Operating principle column in production. The columns contain a molecular...
  • Page 6: Iv.7. Standards

    The variation of the maximum recommended  ow does not the Nidek Medical technical service department if the unit exceed ± 10 % of the indicated value when a back pressure has the marking.
  • Page 7: Iv.10.Troubleshooting

    IV. 10. Troubleshooting. Observations Possible Causes Solutions The I-0 (ON/OFF) button is in the Power cable (9) is not correctly Check the cable connection. “I” (ON) position but the device plugged into the wall outlet. does not operate. The audible alarm sounds continu- Power failure.
  • Page 8: Emc, Electromagnetic Statements

    V. EMC, Electromagnetic Statements May 2011 2010-2238CE Rev C...
  • Page 9 2010-2238CE Rev C May 2011...
  • Page 10 NOTES ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- May 2011 2010-2238CE Rev C...
  • Page 11 NOTES ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- ------------------------------------------------------------------------------------------------------------------- 2010-2238CE Rev C May 2011...
  • Page 12 EU Representative mdi Europa GmbH Langenhagener Str. 71 30855 Hannover-Langenhagen Germany Tel: +49-511-39-08 95 30 Nidek Medical Products, Inc. 3949 Valley East Industrial Drive Fax: +49-511-39-08 95 39 Birmingham, Alabama 35217 U.S.A. info@mdi-europa.com Tel: 205-856-7200 Fax: 205-856-0533 www.mdi-europa.com May 2011...

This manual is also suitable for:

Mark 5 nuvo 8

Table of Contents