BEFORE TAKING A MEASUREMENT
To ensure a reliable reading follow these recommendations:
INSTRUCTION MANUAL
1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes
before taking a measurement. Rest for at least 15 minutes before taking the
measurement.
Automatic
2. Stress raises blood pressure. Avoid taking measurements during stressful times.
Blood Pressure Monitor
3. Measurements should be taken in a quiet place.
4. Remove tight-fitting clothing from your left arm.
Model 7400REL
5. Sit on a chair with your feet flat on the floor. Rest your left arm on table so that the
cuff is at the same level as your heart.
6. Remain still and do not talk during the measurement.
7. Wait 2-3 minutes between measurements. The wait time allows the arteries to return
to the condition prior to taking the blood pressure measurement. You may need to
9494942-9A
increase the wait time depending on your individual physiological characteristics.
8. Keep a record of your blood pressure and pulse readings for your physician. A
INTRODUCTION
single measurement does not provide an accurate indication of your true blood
pressure. You need to take and record several readings over a period of time. Try to
Thank you for purchasing the ReliOn
®
7400REL Automatic Blood Pressure Monitor.
measure your blood pressure at the same time each day for consistency.
Fill in for future reference.
KNOW YOUR UNIT
DATE PURCHASED: __________________
Main Unit:
Display
SERIAL NUMBER: ____________________
USER ID
• Staple your purchase receipt here
Selection Switch
Air Jack
Your new blood pressure monitor uses the oscillometric method of blood pressure
measurement. This means the monitor detects your blood's movement through your
brachial artery and converts the movements into a digital reading. An oscillometric
SET
Up/Down (
monitor does not need a stethoscope so the monitor is simple to use.
button
The 7400REL comes with the following components:
• Monitor • Wide Range Cuff • AC Adapter
Components:
• Instruction Manual printed in English and Spanish • Quick Start Guide
Please read this instruction manual thoroughly before using the unit.
Please keep for future reference.
For specific information about your own blood pressure, CONSULT YOUR
Air tube
DOCTOR.
Air plug
SAVE THESE INSTRUCTIONS
Wide Range Cuff
SAFETY INFORMATION
INTENDED USE
This monitor is designed to measure blood pressure and pulse rate of adults within the
range of the designated arm cuff, following the directions provided in this instruction
manual. The monitor is designed for home use. Read the Safety Information in this
instruction manual before using this monitor.
UNIT DISPLAY
To assure the correct use of the product, basic safety measures should always be
followed including the warnings and cautions listed in this instruction manual.
Average value symbol
SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL
Memory symbol
Indicates a potentially hazardous situation which, if not avoided,
WARNING
Systolic blood pressure
could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury to the user or patient or
Diastolic blood pressure
CAUTION
damage to the equipment or other property.
Deflation symbol
OPERATING THE DEVICE
Battery low indicator
Contact your physician for specific information about your blood pressure.
Self-diagnosis and treatment using measured results may be dangerous. Follow the
instructions of your physician or licensed healthcare provider.
DO NOT adjust medication based on measurement results from this blood pressure
DISPLAY SYMBOLS
monitor. Take medication as prescribed by your physician. Only a physician is
qualified to diagnose and treat High Blood Pressure.
IRREGULAR HEARTBEAT SYMBOL (
The monitor is not intended to be a diagnostic device.
When the monitor detects an irregular rhythm two or more times during the
If battery fluid should get in your eyes, immediately rinse with plenty of clean
measurement, the Irregular Heartbeat Symbol (
water. Contact a physician immediately.
measurement values.
Read all of the information in the instruction manual and any other literature in the
An irregular heartbeat rhythm is defined as
box before operating the unit.
a rhythm that varies by less than 25% from
the average rhythm or 25% more than the
Do not use this device on infants or persons who cannot express their intentions.
average rhythm detected while the monitor
Read the Special Conditions section of the instruction manual if your systolic
is measuring the systolic and diastolic
pressure is known to be more than 220 mmHg. Inflating to a higher pressure than
blood pressure.
necessary may result in bruising where the cuff is applied.
If the irregular heartbeat symbol (
Operate the device only as intended. Do not use the device for any other purpose.
displays with your measurement results,
Dispose of the device, components and optional accessories according to applicable
we recommend you consult your
local regulations. Unlawful disposal may cause environmental pollution.
physician. Follow the directions of your
doctor.
Do not use a cellular phone near the device. It may result in an operational failure.
MOVEMENT ERROR SYMBOL (
Use only authorized parts and accessories. Parts and accessories not approved for
use with the device may damage the unit.
The Movement Error Symbol (
) is displayed if you move your body during the
measurement. Please remove the arm cuff, and wait 2-3 minutes. Reapply the arm cuff
Use only 1.5V Alkaline batteries with this device. Other types of batteries may
and take another measurement.
damage the unit.
AVERAGE VALUE SYMBOL (
Do not use in a location with moisture, or a location where water may splash on the
unit.
The Average Value Symbol (
) is displayed when you press the memory button.
The most recent average reading appears on the display screen.
RISK OF ELECTRICAL SHOCK
HEARTBEAT SYMBOL (
)
Do not plug or unplug the power cord into the electrical outlet with wet hands.
The Heartbeat symbol flashes on the display at every heartbeat during the
Do not overload power outlets. Plug the device into the appropriate voltage outlet.
measurement.
CARE AND MAINTENANCE
Do not subject the monitor to strong shocks, such as dropping the unit on the floor.
Do not submerge the device or any of the components in water.
Store the device and the components in a clean, safe location.
Changes or modification not approved by the manufacturer will void the user
warranty. Do not disassemble or attempt to repair the unit or components.
USING THE AC ADAPTER
IMPORTANT SAFETY INSTRUCTIONS
SAVE THESE INSTRUCTIONS
- DANGER -
TO REDUCE THE RISK OF
FIRE OR ELECTRIC SHOCK,
CAREFULLY FOLLOW THESE INSTRUCTIONS.
For connection to a power supply not in the U.S.A., use an
attachment plug adapter of the proper configuration for the power
outlet.
The power unit (AC Adapter) is intended to be correctly oriented
in a vertical or floor mount position.
NOTES:
• Use only the authorized AC Adapter included with this monitor.
• We recommend you install the batteries even if the AC Adapter is
used. If no batteries are installed, you may need to reset the date
and time if the AC Adapter is disconnected. The measurement
results will not be deleted.
Battery compartment
CAUTION
Use only authorized parts and accessories. Parts and accessories not
approved for use with the device may damage the unit.
ON/OFF button
Memory button
AC Adapter Jack
) buttons
TO CONNECT THE AC ADAPTER
1. Insert the AC Adapter Plug into the
AC Adapter Jack on the right side
Housing
of the monitor as illustrated.
2. Plug the AC Adapter into a 120V
AC outlet (60 cycles).
AC Adapter
TO DISCONNECT THE AC ADAPTER
1. Hold and pull the Housing to remove the AC Adapter from the electrical
INSTRUCTION MANUAL
Automatic
Blood Pressure Monitor
Model 7400REL
Blood Pressure Monitor
outlet. Do not remove by pulling the cord.
Quick Start Guide
9494942-9A
INTRODUCTION
Thank you for purchasing the ReliOn
7400REL Automatic Blood Pressure Monitor.
®
7400REL
Fill in for future reference.
Blood Pressure
Blood Pressure
Blood Pressure
Automatic
DATE PURCHASED: __________________
Monitor
2. Remove the AC Adapter plug from the monitor.
SERIAL NUMBER: ____________________
Instruction
Quick Start
Manual
Guide
WARNING
Do not plug or unplug the power cord into the electrical outlet with
wet hands.
WARNING
Date/Time Display
Do not overload power outlets. Plug the device into the appropriate
USER ID symbol (A/B)
voltage outlet.
Blood pressure level indicator
BATTERY INSTALLATION
Movement error symbol
1. Press the hook on the top of the battery cover and lift the
Irregular heartbeat symbol
cover off in the direction of the arrow.
Pulse display and Memory number
Battery Cover
Heartbeat symbol
2. Install 4 "AA" size batteries so the + (positive) and
- (negative) polarities match the polarities of
the battery compartment as indicated.
)
) will appear on the display with the
3. Replace the battery cover.
Normal Heartbeat
Pulse
Blood
pressure
BATTERY REPLACEMENT
Irregular Heartbeat
)
Low Battery Indicator
Short Long
Pulse
When the Low Battery Indicator appears on the display, turn the monitor off and
remove all the batteries. Replace with four new batteries at the same time. Long-life
Blood
pressure
alkaline batteries are recommended.
)
WARNING
If battery fluid should get in your eyes, immediately rinse with
plenty of clean water. Contact a physician immediately.
NOTES: • Turn the unit off before replacing the batteries. If the batteries are removed
while the unit is still on, the date and time will be reset to that of the
)
previous use. However, the measurement results are not deleted.
• When the batteries are replaced, you may need to reset the date and time. If
"-:--" appears on the display, refer to "Setting the Date and Time".
CAUTION
Dispose of the device, components and optional accessories according to
applicable local regulations. Unlawful disposal may cause environmental
pollution.
SETTING THE DATE AND TIME
Set the monitor to the current date and time before taking a measurement for the first
time.
1. TO START THE SETTING
Press the SET button to adjust the date and time.
The year flashes on the display.
3. Press the ON/OFF button. All symbols appear on the display.
The cuff starts to inflate automatically. As the cuff inflates, the monitor automatically
determines your ideal inflation level.
Remain still and do not move your arm until the entire measurement process is completed.
2. SETTING THE YEAR
The year can be set between 2013 and 2040. When the
display reaches 2040, it will return to 2013.
Press the Up button (
) to advance by one year.
NOTE: To stop the inflation or measurement, push the ON/OFF button. The
Press the SET button to set the current year.
monitor will stop inflating, start deflating, and will turn off.
4. Inflation stops automatically and the measurement is started.
The month flashes on the display.
As the cuff deflates, decreasing numbers appear on the display and the Heartbeat
Symbol (
) flashes.
NOTES: • Press the Down button (
) to decrease the
date and time values.
• Press and hold the button to increase or
decrease the date and time values faster.
3. SETTING THE MONTH
5. When the measurement is complete, the arm cuff completely
Press the Up button (
) to advance by one month.
deflates. Your blood pressure and pulse rate are displayed.
Press the SET button to set the current month.
The day flashes on the display.
4. SETTING THE DAY
Press the Up button (
) to advance by one day.
Press the SET button to set the current day.
The hour flashes on the display.
The 1999 WHO/ISH guidelines for the management of hypertension are as follows:
Category / Blood pressure
5. SETTING THE HOUR
Hypertension
The time is set using AM or PM.
Press the Up button (
) to advance by one hour.
High-Normal
Press the SET button to set the current hour.
Normal
The minute flashes on the display.
This is a general guideline because blood pressure depends on age and morbidity.
A lower home blood pressure goal may be recommended for certain patients,
6. SETTING THE MINUTE
including diabetic patients, pregnant women, and patients with renal failure.
Press the Up button (
) to advance by one minute.
Cord
Press the SET button to set the minute setting.
6. Press the ON/OFF button to turn the monitor off.
7. TO TURN THE MONITOR OFF
Press the ON/OFF button.
NOTE: The monitor will automatically turn off 2 minutes
after the minute has been set.
NOTE: The monitor will automatically turn off after 2
minutes.
APPLYING THE ARM CUFF
WARNING
Contact your physician for specific information about your blood pressure.
Self-diagnosis and treatment using measured results may be dangerous.
CAUTION
Follow the instructions of your physician or licensed healthcare provider.
Read all of the information in the instruction manual and any other
literature in the box before operating the unit.
CAUTION
WARNING
This device is intended for use in measuring blood pressure and
DO NOT adjust medication based on measurement results from this blood
pulse rate in the adult population. Do not use this device on infants
pressure monitor. Take medication as prescribed by your physician. Only a
or persons who cannot express their intentions.
physician is qualified to diagnose and treat High Blood Pressure.
APPLYING THE CUFF ON THE LEFT ARM
WARNING
1. Make sure the air plug is securely inserted
The monitor is not intended to be a diagnostic device.
in the main unit.
SPECIAL CONDITIONS
CAUTION
Inflating to a higher pressure than necessary may result in bruising where
the cuff is applied.
2. Remove tight-fitting clothing from your upper
If your systolic pressure is known to be more than 220 mmHg, press and
arm.
hold the ON/OFF button until the monitor inflates 30 to 40 mmHg higher
than your expected systolic pressure, after the cuff starts to inflate.
3. Apply the cuff to your left upper arm so
Left Arm
1/2 inch (1 - 2 cm)
the arrow is centered on the inside of your arm
and aligned with your middle finger.
The air tube runs down the inside of your arm.
The bottom of the cuff should be
Arrow
Air Tube
approximately 1/2 " above your elbow.
4. Wrap the cuff firmly in place around your
NOTES: • Do not apply more pressure than necessary.
arm using the cloth fastening.
• The monitor will not inflate above 299 mmHg.
When the measurement is complete, the arm cuff completely deflates.
Your blood pressure and pulse rate are displayed.
APPLYING THE CUFF ON THE RIGHT ARM
When taking a measurement using the right arm, use this instruction for Step 3.
USING THE MEMORY FUNCTION
Right Arm
Apply the cuff to your right upper arm so
The monitor automatically stores up to 100 sets of measurement values (blood pressure
the arrow and the air tube runs along the inside
and pulse rate) for each user (A or B). When 100 sets of measurement values are
of your arm. The bottom of the cuff should be
stored, the oldest record is deleted to save the most recent values.
approximately 1/2 " above your elbow.
Arrow
AVERAGING FUNCTION
NOTE: Be careful not to rest your arm on the air tube. This will restrict the flow of air
The monitor calculates the average reading based on the three most recent sets of
to the cuff.
measurement values taken within 10 minutes of the most recent reading.
NOTES: • If two sets of measurement values are stored in the memory for the 10
TAKING A MEASUREMENT
minutes period, the average is based on the two sets of measurement
values. If one set of measurement values is stored, this is displayed as the
The monitor is designed to store the blood pressure and the pulse rate in the memory
average.
for two people (USER A or USER B) every time a measurement is completed.
• When viewing the reading taken without setting the date and time, "-:--" is
1. Sit on a chair with your feet flat on the floor.
displayed instead of the date and time.
Place your arm on a table so the cuff is level
• The average reading will not be calculated without setting the date and time.
with your heart.
TO DISPLAY THE MEASUREMENT VALUES
Keep still and do not talk during measurement.
1. Select your USER ID (A or B).
2. Select your USER ID (A or B).
2. Press the Memory button (
The Average Symbol (
reading is shown.
3. Press the Down button (
) to display the most recent set of
measurement values on the screen. The values are displayed from the most
recent to the oldest.
NOTES:
• The Memory number appears for
a second before the pulse rate is
Alternating display
displayed. The newest set is
number "1".
• The date and time are alternately
displayed with the measurement
values.
Press the Up (
) or Down (
) button
to display the next set of values.
: to display the older readings
: to display the recent readings
NOTE:
Press and hold the Up or Down button to display the values faster.
TO DELETE ALL VALUES STORED IN THE MEMORY
The values stored in the memory are deleted by USER ID.
You cannot partially delete values stored in the memory. All values for
the user you select will be deleted.
Systolic blood pressure
1. Select your USER ID (A or B).
Red
above 139 mmHg
Yellow
130 - 139 mmHg
Green
129 mmHg or less
2. While holding the Memory button (
), press the ON/OFF button down
for more than 2 seconds until the display appears as it does in the
Diastolic blood pressure
Red
above 89 mmHg
following illustration.
Yellow
85 - 89 mmHg
Green
84 mmHg or less
Systolic blood pressure
Diastolic blood pressure
above 139 mmHg
above 89 mmHg
130 - 139 mmHg
85 - 89 mmHg
3. Press the ON/OFF button to turn the monitor off.
129 mmHg or less
84 mmHg or less
NOTE: The monitor will automatically turn off after 2 minutes.
CARE AND MAINTENANCE
To keep your blood pressure monitor in the best condition and protect the unit from
damage, follow the directions listed below:
Do not forcefully bend the arm cuff or air tube. Do not fold tightly.
Clean the monitor with a soft dry cloth.
Do not use any abrasive or volatile cleaners.
Do not attempt to clean the cuff.
CAUTION
Do not submerge the device or any of the components in water. Do
not subject the monitor to extreme hot or cold temperatures,
humidity or direct sunlight.
CAUTION
Store the device and the components in a clean, safe location.
CAUTION
Do not subject the monitor to strong shocks, such as dropping the unit on the floor.
Remove the batteries if the unit will not be used for three months or longer. Always
replace all the batteries with new ones at the same time.
Use the unit in a manner consistent with the instructions provided in this manual.
CAUTION
Use only authorized parts and accessories. Parts and accessories not
approved for use with the device may damage the unit.
CAUTION
Changes or modification not approved by the manufacturer will void the user
warranty. Do not disassemble or attempt to repair the unit or components.
ERROR INDICATORS
SYMBOL
CAUSE
CORRECTION
Insert the plug securely. Refer to
Air plug disconnected.
"Applying the Arm Cuff".
Arm cuff not applied
Apply the arm cuff correctly. Refer to
correctly.
"Applying the Arm Cuff".
Air is leaking from the
Replace the cuff with a new one.
arm cuff.
Repeat measurement. Remain still and do
not talk during measurement. Refer to
"Taking a Measurement".
Cuff over or under
inflated.
If "E2" appears repeatedly, inflate the cuff
manually until it is 30 to 40 mmHg above
your previous measurement result. Refer
to "Taking a Measurement".
The arm cuff was inflated
Do not inflate the cuff above 299 mmHg.
above 299 mmHg when
Refer to "Taking a Measurement".
inflating the cuff manually.
Repeat measurement. Remain still and do
Movement during
not talk during measurement. Refer to
measurement.
"Taking a Measurement".
Remove any clothing interfering with the
Clothing is interfering
arm cuff. Refer to "Applying the Arm
with the arm cuff.
Cuff".
Device error.
Contact Customer Service.
Replace the four batteries.
Batteries are worn.
Refer to "Battery Installation".
TROUBLESHOOTING TIPS
).
) appears on the display when the average
PROBLEM
CAUSES AND SOLUTIONS
No power.
Replace all four batteries with new ones.
No display appears on the unit.
Check the battery installation for proper placement
of the battery polarities.
Blood pressure varies constantly. Many factors
including stress, time of day, how you wrap the
Measurement values appear too
cuff, may affect your blood pressure.
high or too low.
Review the sections "Before Taking a
Measurement" and "Taking a Measurement".
HEM-8722-WM_A_M09_120831.pdf
FCC STATEMENT
Note:
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)
This product has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful interference
in a residential installation. The product generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instructions,
may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If the product does
cause harmful interference to radio or television reception, which can be determined by
turning the product on and off, the user is encouraged to try to correct the interference
by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the product and the receiver.
• Connect the product into an outlet on a circuit different from that to which the
receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only)
This digital apparatus does not exceed the Class B limits for radio noise emissions
from digital apparatus as set out in the interference-causing equipment standard entitled
"Digital Apparatus", ICES-003 of the Canadian Department of Communications.
Cet appareil numérique respecte les limites de bruits radioeléctriques applicables aux
appareils numériques de Clase B prescrites dans la norme sur le materiel brouilleur:
"Appareils Numériques", ICES-003 édictée par le minister des communications.
Changes or modifications not expressly approved by the party responsible for
compliance could void the user's authority to operate the equipment.
LIMITED WARRANTY
Your ReliOn
®
7400REL Automatic Blood Pressure Monitor, excluding the arm cuff, is
warranted to be free from defects in materials and workmanship appearing within 5
years from the date of purchase, when used in accordance with the instructions
provided with the monitor. The arm cuff is warranted to be free from defects in
materials and workmanship appearing within one year from the date of purchase when
the monitor is used in accordance with the instructions provided with the monitor. The
above warranties extend only to the original retail purchaser.
We will, at our option, repair or replace without charge any monitor or arm cuff
covered by the above warranties. Repair or replacement is our only responsibility and
your only remedy under the above warranties.
To obtain warranty service contact Customer Service by calling 1-800-634-4350 for the
address of Inspection center.
Enclose the Proof of Purchase. Include a letter, with your name, address, phone
number, and description of the specific problem. Pack the product carefully to prevent
damage in transit. Because of possible loss in transit, we recommend insuring the
product with return receipt requested.
THE FOREGOING IS THE SOLE WARRANTY PROVIDED BY OMRON IN
CONNECTION WITH THIS PRODUCT, AND OMRON HEREBY DISCLAIMS
ANY OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING IMPLIED
WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE. IMPLIED WARRANTIES AND OTHER TERMS THAT MAY BE
IMPOSED BY LAW, IF ANY, ARE LIMITED IN DURATION TO THE PERIOD
OF THE ABOVE EXPRESS WARRANTY.
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER
SPECIAL, INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS,
EXPENSES OR DAMAGES.
This warranty provides you with specific legal rights, and you may have other rights
that vary by jurisdiction. Because of special local requirements, some of the above
limitations and exclusions may not apply to you.
FOR CUSTOMER SERVICE
Visit our web site at:
www.omronhealthcare.com
Call toll free:
1-800-634-4350
SPECIFICATIONS
Model:
7400REL (HEM-8722-WM)
Display:
LCD Digital Display
Pressure: 0 to 299 mmHg, Pulse: 40 to 180/min.
Measurement Range:
Pressure: ±3mmHg or 2% of reading
Accuracy/Calibration:
Pulse: ±5% of reading
Automatic by electric pump
Inflation:
Automatic pressure release valve
Deflation:
Rapid Air Release:
Automatic exhaust valve
Measurement Method:
Oscillometric method
Power Source:
120V 60Hz AC adapter or 1.5V 4 "AA" batteries
Approx. 300 uses with 4 new alkaline batteries
Battery Life:
50ºF to 104ºF (10ºC to 40ºC) /15 to 90% RH
Operating Temperature
/Humidity:
-4ºF to 140ºF (-20ºC to 60ºC) /10 to 95% RH
Storage Temperature
700-1060 hPa
/Humidity/Air Pressure:
Main Unit Weight:
Approximately 13 3/8 oz. (380g)
not including batteries
Main Unit Dimensions:
Approx. 6"(l) × 4 7/8"(w) × 3 1/2" (h)
(152 mm × 123 mm × 88 mm)
Cuff Size:
Approx. 6"(w) × 22 1/8"(l) (151 mm × 563 mm)
Cuff tube 35 3/8" (900 mm)
Fits arm circumferences 9" to 17"
Cuff Circumference:
(220 mm to 420 mm)
Up to 100 readings per user
Memory:
Main Unit, Wide Range Cuff, AC Adapter, Instruction
Contents:
Manual, and Quick Start Guide
Wide Range Cuff 9'' - 17'' (220mm - 420mm), AC Adapter
Optional Accessories:
NOTE: These specifications are subject to change without notice.
Manufactured for OMRON HEALTHCARE Co., Ltd.
Made in China
Distributed by OMRON HEALTHCARE, INC.
1925 West Field Court
Lake Forest, IL 60045 U.S.A.
www.omronhealthcare.com
© 2012 OMRON HEALTHCARE, INC.
ReliOn
®
is a registered trademark used under license by OMRON HEALTHCARE,
INC.
.................................................................
Distributed by:
Wal-Mart
Stores, Inc.
®
Bentonville, AR 72716

Advertisement

loading

Summary of Contents for ReliOn 7400REL

  • Page 1 Your ReliOn ® 7400REL Automatic Blood Pressure Monitor, excluding the arm cuff, is Category / Blood pressure Systolic blood pressure Diastolic blood pressure The 7400REL comes with the following components: warranted to be free from defects in materials and workmanship appearing within 5 5.