Setting Up The Master Unit - Agilent Technologies N5700 Series User Manual

System dc power supply
Hide thumbs Also See for N5700 Series:
Table of Contents

Advertisement

Setting up the Master Unit

Series N5700 User's Guide
-S
-LS
+LS
+S
+
V
MASTER
POWER SUPPLY
-
V
J1-25
Parallel
Current Program
J1-8 J1-12
J1-10
+
V
SLAVE
POWER SUPPLY
-
V
-S
-LS
+LS
+S
-S
+S
+
V
MASTER
POWER SUPPLY
-
V
J1-25
Parallel
Current Program
J1-8 J1-12
J1-10
+
V
SLAVE
POWER SUPPLY
-
V
-S
+S
One of the units operates as a master and the remaining units are
slaves. The slave units operate as controlled current sources
following the master output current. In remote operation, only the
master unit can be programmed by the computer while the slave
units may be connected to the computer for voltage, current and
status readback only.
It is recommended that each unit supplies only up to 95% of its
current rating because of the imbalance that may be caused by
cabling and connections voltage drops.
Connect the sensing circuit for either local or remote sensing as
shown in the previous figures. Set the master unit output voltage to
the desired voltage. Program the current limit to the desired load
current limit divided by the number of parallel units. During
operation, the master unit operates in constant voltage mode,
regulating the load voltage at the programmed output voltage.
As short as possible
Twisted
pair
Local Sensing
+S
-S
Twisted
pair
As short as possible
+S
Twisted
pair
-S
Twisted
pair
-S
+S
Remote Sensing
Installation 2
LOAD
LOAD
29