Feature
IR Temperature Range
Accuracy
Repeatability
Display Resolution
Spectral Response
Response Time (95 %)
K-Type Thermocouple
Input Temperature Range
K-Type Thermocouple
Input Accuracy
K-Type Thermocouple
Resolution
Distance:Spot
(90 % energy)
Laser sighting
Emissivity
Data storage
Communication
Operating Altitude
Storage Altitude
Relative Humidity
Operating Temperature
Storage Temperature
Vibration
Weight
Dimensions
Power
Battery Life
CE Certification
561
-40 C to 550 C
(-40 F to 1022 F)
<0 °C: ±(1.0 °C + 0.1 °/1 °C)
>0 °C: ±1 % or ± 1.0 °C, whichever is greater
(<32 F±2 °F ±0.1 °/1 °F)
>32 °F: ±1 % or ±2 °F), whichever is greater
0.5 % of reading or
1 °C (2 °F),
whichever is greater.
0.1 C / 0.1 F
8 to 14 m
<500 ms
0 C to 100 C
(32 F to 212 F)
Input accuracy 2.2 C
( 4 F)
0.1 C (0.1 F)
12:1
Single laser, output <1 mW Class II, wavelength 630 to 670 nm
Lo, Med, Hi
-
none
3000 meters above mean sea level
12,000 meters above mean sea level
10 % to 90 % RH non-condensing up to 30 °C (86 °F)
0 C to 50 C
(32 F to 122 F)
-20 C to 65 C
(-4 F to 149 F)
2.5 G, IEC 68-2-6
0.322 kg (0.7099 lb)
17.69 cm (6.965 in) H x 16.36 cm (6.441 in) L x 5.18 cm (2.039 in) W
2 AA /LR6 Batteries (alkaline or NiCD)
12 hours with laser and backlight on; 100 hours with laser and backlight
off, at 100 % duty cycle (thermometer continuously on)
EN/IEC 61326-1:2006, Class B, Criteria A
EN/IEC 61010-1:2001
EN/IEC 60825-1:2007
566
-40 C to 650 C
-40 C to 800 C
(-40 F to 1202 F)
(-40 F to 1472 F)
0.5 % of reading or 0.5 °C (1 °F),
whichever is greater.
-270 C to 1372 C (-454 F to 2501 F)
-270 C to -40 C: (1 °C + 0.2 °/1 °C)
(-454 F to -40 F: (2 °F + 0.2 °/1 °F))
-40 C to 1372 C: 1 % or 1 °C (-40 F to
2501 F: 1 % or 2 °F), whichever is greater
0.1 C/0.1 F
30:1
50:1
Digitally adjustable from 0.10 to 1.00 by 0.01
or via built-in table of common materials
20 points
99 points
USB 2.0
0 °C to 50 °C (32 °F to 122 °F)
-20 °C to 60 °C (-4 °F to 149 °F)
2 AA /LR6 Batteries
or USB connection
when used with a
PC