External Contact Probe; Troubleshooting - Fluke 56x User Manual

Infrared thermometers
Hide thumbs Also See for 56x:
Table of Contents

Advertisement

External Contact Probe

To avoid electrical shock or personal injury, do not connect the
external contact probe to live electrical circuits.
The thermometers come with a bead K-type thermocouple probe. The probes
attach to the thermometers via the probe input located on the top of the
thermometer (see Figure 9).
With the probe installed, the probe symbol () appears on the display. The
probe can be used simultaneously while the thermometer is taking non-contact
measurements. The probe readings are shown below the non-contact
measurements. The thermometer will stay on when there is a probe inserted.
When used with a personal computer and FlukeView Forms (568 only), the auto
off feature is disabled.

Troubleshooting

See Table 6 for remedies to possible problems when using the thermometer.
Symptom
--- (on display)
Blank display
Laser does not work
Failed communication via USB
Inaccuracy
Settings such as emissivity,
date/time, F/C, and saved data
lost
XWWarning
Table 6. Troubleshooting
Cause
Target temperature is over or
under range.
Low batteries
Thermometer is asleep
Possible dead batteries
Low or dead batteries
Ambient temperature is above
40 C (104 F)
Unit is off
FlukeView Forms is not running
Possible incorrect emissivity
setting, field of view, or spot size
Battery dead or not replaced in
<1 minute of removal
Infrared Thermometers
External Contact Probe
Action
Select target within
specifications
Replace Batteries
Pull trigger
Replace batteries
Replace batteries
Use in area with lower ambient
temperature
Pull the trigger
Start FlukeView Forms
See "Emissivity", "Field of View"
and "Distance and Spot Size"
sections.
Reset settings. Replace
batteries as soon as low battery
indicated; Exchange the
batteries within one minute of
removal. Refer to "Changing the
Batteries".
21

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents