5LXO
INDICATORS:
2
Label
Indicator Lights
O
Purity
Symbol
2
SYSTEM FAILURE
RED - Continuous Audible
Below 73%
Alarm Sieve-GARD™
Compressor Shutdown
73% to 85%
YELLOW
SYSTEM OKAY
O
GREEN
2
Over 85%
5LX INDICATORS:
Label
Status
Indicator Lights
Symbol
SYSTEM FAILURE
RED
I/O
SYSTEM OKAY
GREEN