Sensorsoft Thermometer User Manual
ARGUMENT-byte
The number of the Sensorsoft Thermometer's internal register you want to read or write. This byte
is required for the temperature command. This byte is not required for the status command.
CRC-bytes
A two byte (16 bit) Cyclic Redundancy Check that is calculated on the entire packet from the
CMD byte to the last ARGUMENT byte or ADDRESS byte. In the examples below, the last two
bytes in each command string are CRC bytes that were pre-calculated.
Send the following bytes to request a status reading:
<C1h><0Bh><00h><01h><00h><00h><00h><00h><00h><047h><098h>
Send the following bytes to request a temperature reading in Celsius (0.5 C resolution):
<C5h><0Ch><00h><01h><00h><00h><00h><00h><00h><01h><0Eh><49h>
Send the following bytes to request a temperature reading in Celsius (0.1 C resolution):
<C5h><0Ch><00h><01h><00h><00h><00h><00h><00h><02h><6Dh><79h>