6.1.6 Connecting Micropak

It is necessary to perform the Micropak 9 / 7 / 5 setting when connecting the Micropak to the
counter unit.
For information about setting up Micropak 9 / 7 / 5, refer to the Micropak user's manual.
Connection
(1)
Counter unit
EXT.LOAD
SERIAL
Micropak 9/7 setup
(2)
Set Equipment--> Counter-->
Set Equipment--> Counter--> X-axis Counter--> RS-232C Connection
Set Equipment--> Counter--> Counter connected--> MF
Set Conditions--> Measuring Rang--> X Y Z
Micropak 5 setup
(3)
Set the Dip switch DPS1 and DPS2 according to the following.
DPS1 No.1 : ON, No.2 : OFF, No.3 : OFF, No.4 : OFF
DPS2 No.5 : ON, No.6 : OFF
Modify also the parameter setup of the counter simultaneously. (Set "4800bPS" at the
parameter number F7-1 according to "3.4.2 About parameters".)
(Note)
To connect the counter to Micropak 5, unless the default setting are modified, the counter
cannot be connected to Micropak 5. (Because MPK-5 does not accept the Q-axis data.)
Modify also the parameter setup of the counter so that Q-axis output may not be performed.
(Set "2 AS" at the parameter number F8 according to "3.4.2 About parameters".)
No.99MBA043A
Order No.937998
External load cable
Order No.12AAA202
Connecting cable
(Set No . 12AAA500
f o r t wo p i ec es o f a b o v e
m en t i o n ed c a b l e)
--> 9600 bps
Baud Rate
Y-axis Counter--> RS-232C Connection
Z-axis Counter--> RS-232C Connection
6. Specifications
Micropak 9/7/5
EXT.LOAD
INPUT
4800bps
6 - 9