Angle counter
[1] Angle display
Range : ± 360 ° ( ± 370 ° in ABS mode)
Resolution : 1' (or 0.01 ° )
[2] Zero-set key
Sets the angle counter to zero at any position desired.
ABC mode : Hold down the key for about 3 seconds.
INC mode : Press then release the key.
[3] ABS/INC mode indicator
The LED for the active mode is lit.
[4] Angle unit selector/offset switch
Unit of angle display : Switches the smallest unit of the angle counter between 0.01 °
and 1'(Default is 1' at power on). For this switching hold down the switch for about 2
seconds.
Offset function : Enables an easy angle measurement. Each time the switch is
pressed the displayed angle increments by 90 ° .
[5] ABS/INC mode selector switch
Switches between ABS and INC modes. (Default is ABS mode at power ON.)
INC mode
Use the INC mode to perform comparative measurement. Pressing the [ABS/INC] key
when in ABS mode switches to INC mode and the counter is Set to Zero.
Press the [ABS/INC] key again to return to ABS mode where the current screen angle
from the datum is displayed.
Error codes
E-oF : Counter overflow
E-oS : Overspeed error
If one of the above errors are displayed, press the [ZERO] key to clear the error. Then
set the datum edge again in ABS mode.
No.99MBA043A
[1]