Download Print this page

GE DigitalFlow GS868 Product Overview And Specifications page 3

Panametrics ultrasonic steam mass flowmeter
Hide thumbs Also See for DigitalFlow GS868:

Advertisement

GS868 Specifi cations
Operation and Performance
Fluid Types
Saturated and superheated steam
Pipe Sizes
2 in to 48 in NB (50 mm to 1200 mm DN)
Pipe Materials
All metals. Consult GE for other materials.
Flow Accuracy (Velocity)
±1% to 2% of reading typical
Accuracy depends on pipe size and whether measurement
is one-path or two-path. Accuracy to ±0.5% of reading
may be achievable with process calibration.
Repeatability
±0.2% to 0.5% of reading
Range (Bidirectional)
–150 to 150 ft/s (–46 to 46 m/s)
Rangeability (Overall)
1500:1
Specifi cations assume a fully developed fl ow profi le
(typically 20 diameters upstream and 10 diameters
downstream of straight pipe run) and fl ow velocity greater
than 3 ft/s (1 m/s).
Measurement Parameters
Mass fl ow, standard and actual volumetric fl ow, totalized
fl ow, and fl ow velocity
Electronics
Flow Measurement
Patented Correlation Transit-Time mode
Enclosures
• Standard: Epoxy-coated aluminum weatherproof
Type 4X/IP66, FM/CSA Class I, Division 2, Groups
A,B,C&D
• Optional: Stainless steel, fi berglass, explosion-proof,
fl ameproof, ISSep 03ATEX113
Transit-time fl ow measurement technique
Dimensions
Standard: Weight 11 lb (5 kg),
size (h x w x d) 14.24 in x 11.4 in x 5.12 in
(362 mm x 290 mm x 130 mm)
Channels
• Standard: One channel
• Optional: Two channels (for two pipes or two-paths)
Display
Two independent software-confi gurable 64 x 128 pixel
backlit LCD graphic displays
Keypad
39-key tactile-feedback membrane keypad
Power Supplies
• Standard: 100 to 130 VAC, 50/60 Hz
or 200 to 265 VAC, 50/60 Hz
• Optional: 12 to 28 VDC, ±5%
Power Consumption
20 W maximum
Operating Temperature
–4°F to 131°F (–20°C to 55°C)
Storage Temperature
–67°F to 167°F (–55°C to 75°C)
Standard Inputs/Outputs
Two 0/4 to 20 mA isolated outputs, 550Ω maximum load

Advertisement