Transporting Pure Water (Permeate) To Point-Of-Use; E-Series Feed Water Requirements - GE OSMONICS E4 Series Installation, Operation And Maintenance Manual

Ge water purification machines installation, operation, and maintenance manual e4, e4le, ez4
Table of Contents

Advertisement

Temperature
Inlet Pressure
Chlorine
(continuous feed)
Operating pH
Silt Density
Index (SDI)
*
American Standard for Testing Materials
2.6

Transporting Pure Water (Permeate) to Point-of-Use

The pure water, or permeate, is in an aggressive state and should only be transport-
ed from the machine to the point-of-use in food grade flexible nylon, stainless steel
(SS) tubing, or polyvinyl chloride (PVC) material for the inlet, permeate, and concen-
trate piping sizes. Refer to Connections (Table 2.1) for inlet, permeate, and concen-
trate piping sizes.
WARNING:
Table 2.2
E-Series
Feed Water Requirements
Typical:
Limits:
Minimum:
Maximum:
0 parts per million (ppm)
5.5 - 8.5
Less than or equal to 5 to minimize
membrane element fouling and
extend cleaning intervals.
Refer to ASTM* standard D4189.
MACHINE DAMAGE MAY OCCUR IF PERMEATE BACK PRESSURE
EXCEEDS 60 PSIG (4.1 BARG) DURING OPERATION.
3
50° - 85°F (10° - 29°C)
33° - 104°F (0.16° - 40°C)
30 psig (2.1 barg)
60 psig (4.1 barg)