Product Water Analysis Procedure - GE MILLENIUM OSMONICS MILLENIUMTM REVERSE OSMOSIS UNIT Operation And Maintenance Manual

Ge osmonics milleniumtm reverse osmosis unit operation and maintenance manual
Table of Contents

Advertisement

2.3

Product Water Analysis Procedure:

In order to ensure that the RO is performing at or above AAMI (Association for the
Advancement of Medical Instrumentation) Standards, it is necessary to perform an
AAMI Water Quality Analysis test. For new units, the RO must be operated (rinsed to
drain) for at least two hours before taking a sample, to ensure that the membranes are free
of preservative or other contaminants which would invalidate the test.
WARNING:
There are many documented clinical risks in using less than AAMI Standard
product water (refer to Technote 103, "AAMI Hemodialysis Water Quality
Standards").
NOTE:
Follow the procedures recommended by your water quality-testing laboratory if
they vary from the following steps.
Materials Required: Produce water sampling kit, sterile gauze, tape, and container (such
as a small bucket).
1.
Ensure that all pre-RO devices are operating correctly and are performance
verified by the appropriate testing.
2.
Connect the RO unit to the required utilities (feed water, drain, power).
3.
Operate the RO unit with product and waste flows to drain for at least 15 minutes
after the product water quality has stabilized before drawing the water sample.
4.
Hold the product line over the container and run product water for 60 seconds.
5.
Aseptically remove the lid from the sample bottle and rinse the bottle twice with
RO product water.
6.
Fill the bottle completely with RO product water and cap it immediately.
7.
Fill out the laboratory requisition form. Ensure the requisition form has, at a
minimum, information to identify the facility or location of the RO, the serial
number or facility equipment number to ensure which RO was tested and name,
phone number and address of whom to contact.
8.
Place the bottle and the completed requisition in the pre-addressed mailer and
mail it directly to the laboratory.
NOTE:
Total chlorine must be measured on-site. Tests for Chlorine/Chloramine (total
chlorine) are not typically reported on the AAMI Analysis. Chlorine and
chloramine dissipate from the water sample over time and reported results will be
inaccurate.
NOTES:
14693e - 31 Mar 03
M illenium™ RO System by GE Osmonics
M
2 - 6
Installation Guidelines and Instructions

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Osmonics millenium

Table of Contents