Ac Voltage (Sinewave) - Fluke 5500A/EP User Manual

Enhancement package
Table of Contents

Advertisement

AC Voltage (Sinewave)

Range
1 mV to 32.999 mV
33 mV to 329.999 mV
0.33V to 3.29999V
3.3V to 32.9999V
33V to 329.999V
330V to 1000V [1]
1.
The standard Hewlett Packard 3458A can measure up to 700V. A high voltage measuring option is
available from Hewlett Packard for making measurements up to 1000V.
Part II Specifications: 5500A and 5500A/EP System Repeatability
Frequency
10 Hz to 45 Hz
45 Hz to 10 kHz
10 kHz to 20 kHz
20 kHz to 50 kHz
50 kHz to 100 kHz
100 kHz to 500 kHz
10 Hz to 45 Hz
45 Hz to 10 kHz
10 kHz to 50 kHz
50 kHz to 100 kHz
100 kHz to 500 kHz
10 Hz to 45 Hz
45 Hz to 10 kHz
10 kHz to 50 kHz
50 kHz to 100 kHz
100 kHz to 500 kHz
10 Hz to 45 Hz
45 Hz to 10 kHz
10 kHz to 50 kHz
50 kHz to 100 kHz
45 kHz to 1 kHz
1 kHz to 20 kHz
45 Hz to 1 kHz
1 kHz to 10 kHz
24 hour Repeatability, 2 C, 10% rh
(ppm output + V)
700 + 3 V
300 + 3 V
300 + 3 V
300 + 3 V
400 + 5 V
1200 + 10 V
400 + 10 V
75 + 4 V
125 + 8 V
200 + 32 V
650 + 70 V
275 + 60 V
50 + 25 V
100 + 50 V
200 + 125 V
450 + 250 V
275 + 650 V
75 + 200 V
150 + 600 V
450 + 1.8 mV
75 + 2 mV
150 + 6 mV
75 + 20 mV
50 + 16 mV
25