Page 2
Fluke authorized resellers shall extend this warranty on new and unused products to end-user customers only but have no authority to extend a greater or different warranty on behalf of Fluke. Warranty support is available if product is purchased through a Fluke authorized sales outlet or Buyer has paid the applicable international price.
DC Voltage...................... 18 DC Current ...................... 19 Resistance......................19 AC Voltage (Sinewave)................... 20 AC Current (Sinewave) ................... 22 Part II Specifications: 5500A and 5500A/EP System Repeatability....23 DC Voltage...................... 23 DC Current ...................... 23 Resistance......................24 AC Voltage (Sinewave)................... 25 AC Current (Sinewave) ...................
List of Tables Table Title Page Switch Accessory Lead Functions ................6 5500A/EP 9001 MET/CAL Procedures..............11 Switch Accessory Remote Commands ..............15...
List of Figures Figure Title Page 5500A/EP Enhancement Package Contents............. 1 Battery Eliminators for the 5500/EP Switch Accessory .......... 2 Switch Accessory Leads ..................7 Connection for 5500A Characterization ..............12 Connections for DMM Calibration ................13...
Fluke 8842A Hewlett Packard 34401A Hewlett Packard 3478A Keithley 2000 To use the 5500A/EP, you will need a Fluke 5500A calibrator, a Hewlett Packard 3458A DMM, two IEEE controller cards, and a personal computer running Microsoft Windows ® and Fluke MET/CAL calibration software, Version 4.3 or higher.
Inspect the enclosed equipment carefully for damage and immediately report any damage to the shipping company. When you unpack the 5500A/EP, check for all equipment shown in Figure 1. Report any shortage to the place of purchase or to the nearest Fluke Service Center.
Installing the Software Installation consists of four parts: modify your MET/CAL, establishing traceability for the 5500A, the 5500A/EP (i.e., the switch accessory), and the Hewlett Packard 3458A, entering the configuration information for each, and then installing the verification and adjustment procedures.
3. To enter traceability information for the switch accessory, select Add to open the Add New Trace Entry dialog box. Type “Fluke 5500EP” into the top field. Fill in the remainder of the dialog box and click OK when you are finished.
Enter the alias as 5500EP. Installing the Verification and Adjustment Procedures Place the 5500A/EP disk into disk drive A or B, and select File/Run from the Program Manager. Type “a:setup” or “b:setup” and click OK. Follow the on screen prompts.
5500A/EP Enhancement Package Users Manual Switch Accessory Leads The switch accessory leads are shown in Figure 3 and their functions are provided in Table 1. The functions refer to the terms provided on the panels for the 5500A, Hewlett Packard 3458A, and the specified DMMs.
Page 14
The Switch Accessory Black Black 1000V MAX White 11A MAX 2A MAX White Yellow FLUKE CORPORATION MADE IN USA Yellow w3f.eps Figure 3. Switch Accessory Leads...
It also automatically performs a zeroing operation on the 5500A. Setting up the Equipment The 5500A/EP is designed to work with the same hardware configuration for both the characterization of the 5500A, and adjustment and verification of the DMM. The only difference between the two setups is the use of a Hewlett Packard 3458A for characterization.
Calibrating an Instrument Connecting the Instruments This procedure sets up the instruments in their permanent configuration for both the characterization of the 5500A, and adjustment and verification of the DMM. Refer to Figures 4 and 5 during setup. Warning The binding post adapters have retractable safety shrouds that may expose the metal plug.
1. Open the MET/CAL program group and select the Runtime module. 2. From the procedure list, select the program: Fluke 5500A: (24 hr.) CAL CHAR /3458 /5500A/EP 3. Follow the prompts to complete the program. The MET/CAL program allows for settling times after any additional connections are made during the procedure.
Page 18
Name that appears in procedure directory Function of procedure Fluke 8840A: CAL ADJ IEEE /5500A/EP Adjusts the Fluke 8840A Fluke 8840A: (1 yr.) CAL VER IEEE /5500A/EP Verifies the Fluke 8840A Fluke 8842A: CAL ADJ IEEE /5500A/EP Adjusts the Fluke 8842A Fluke 8842A: (1 yr.) CAL VER IEEE /5500A/EP...
Page 19
5500A/EP Enhancement Package Users Manual Wall Outlet Fluke 5500A (Back) Battery Serial 2 Eliminator To UUT Serial 1 From Host HP3458 (Back) IEEE-488 IEEE-488 To PC IEEE Port 0 HP3458 (Front) Fluke 5500A (Front) INPUT SENSE NORMAL SCOPE TRIG 200V PK...
Page 20
Calibrating an Instrument Wall Outlet Fluke 5500A (Back) Battery Serial 2 Eliminator To UUT Serial 1 From Host Fluke 8842 (Back) IEEE-488 IEEE-488 To PC To PC IEEE Port 0 IEEE Port 1 Fluke 8842A (Front) INPUT SENSE Fluke 5500A (Front)
5500A/EP Enhancement Package Users Manual Remote Commands for Operating the Switch Accessory The switch accessory has 11 states, all of which can only be activated when the switch accessory receives a remote command. Remote commands are sent from a PC and transmitted to the 5500A from the IEEE 4888 or Serial 1 port.
Page 22
Remote Commands for Operating the Switch Accessory Table 3. Switch Accessory Remote Commands Mode Remote Description Command Voltage, Ohms Power- on default. All relays are OFF. Used for Voltage and OHMS COMP off. (no compensation) 4-Wire Short Circuit A> Relay cluster 5 is ON. Provides a 4-Wire short to the UUT.
5500A/EP Enhancement Package Users Manual Operating Guidelines Follow these guidelines carefully when operating the switch accessory. Caution Using the 4-Wire Ohms mode for carrying more than 2A can damage the switch accessory. For currents greater than 2A, use Current mode, and the A output lead.
Packard 3458A DMM. Specifications There are two parts to the specifications for the 5500A/EP, referred to as Parts I and II. Part I specifications provide the 24-hour uncertainty of cardinal points. Part II refers to the repeatability of the 5500A when it is used with the 5500A/EP switch accessory.
Part I Specifications: Absolute System Uncertainty DC Current Cardinal Point Output 24 hour Uncertainty, 2 C, 10% rh (ppm output) 1 mA 60 ppm 10 mA 60 ppm 100 mA 110 ppm 190 mA 97 ppm [1] 0.5A 235 ppm 190 ppm 200 ppm [2] 2.2A...
Part II Specifications: 5500A and 5500A/EP System Repeatability These specifications refer to the repeatability of the 5500A when used with the 5500A/EP switch accessory. It includes error terms for the 5500A calibrator and switch accessory, These specifications includes error terms for both the 5500A and 5500A/EP, assuming the same loading and application of the environmental conditions that are specified.
5500A/EP Enhancement Package Users Manual Resistance Range 24 hour Repeatability, 2 C, 10% rh (ppm output + to 10.99 12 + 1.1 m to 32.999 12 + 2 m to 109.999 9 + 2 m to 329.999 9 + 2 m to 1.09999 k...
Part II Specifications: 5500A and 5500A/EP System Repeatability AC Voltage (Sinewave) Range Frequency 24 hour Repeatability, 2 C, 10% rh (ppm output + V) 10 Hz to 45 Hz 700 + 3 V 45 Hz to 10 kHz 300 + 3 V 1 mV to 32.999 mV...
5500A/EP Enhancement Package Users Manual AC Current (Sinewave) Range Frequency 24 hour Repeatability, 2 C, 10% rh (ppm output + A) 10 to 20 Hz 650 + 55 nA 20 Hz to 45 Hz 325 + 55 nA 0.029 mA to 0.32999 mA...
Other Specifications Other Specifications Switch Accessory Internal Capacitor Value Maximum Voltage 0.01 F 50V dc or ac; protected with a 90V protection device. General Specifications Warm-up Time Four minutes, after connections to the 5500A and DMM have been made. Standard Interfaces RS-232 Temperature Performance Operating: 15 C to 30 C...