Lakeshore 642 User Manual page 87

Electromagnet power supply
Table of Contents

Advertisement

OPST?
Operational Status Query
Input:
OPST? [term]
Returned: <bit weighting> [term]
nnn
Format:
The integer returned represents the sum of the bit weighting of the operational status bits. Refer to
Remarks:
Paragraph 5.1.4.2.2 for a list of operational status bits.
OPSTE
Operational Status Enable Command
Input:
OPSTE <bit weighting> [term]
nnn
Format:
Each bit has a bit weighting and represents the enable/disable mask of the corresponding operational
Remarks:
status bit in the Operational Status Register. This determines which status bits can set the
corresponding summary bit in the Status Byte Register. To enable a status bit, send the command
OPSTE with the sum of the bit weighting for each desired bit. Refer to Paragraph 5.1.4.2.2 for a list
of operational status bits.
OPSTE?
Operational Status Enable Query
Input:
OPSTE?[term]
Returned: <bit weighting> [term]
nnn
Format:
OPSTR?
Operational Status Register Query
Input:
OPSTR? [term]
Returned: <bit weighting> [term]
nnn
Format:
The integers returned represent the sum of the bit weighting of the operational status bits. These status
Remarks:
bits are latched when the condition is detected. This register is cleared when it is read. Refer to
Paragraph 5.1.4.2.2 for a list of operational status bits.
RATE
Output Current Ramp Rate Setting Command
Input:
RATE <rate> [term]
+n.nnnn
Format:
<rate>
Sets the output current ramp rate. This value will be used in both the positive and negative directions.
Remarks:
Setting value is limited by LIMIT.
RATE?
Output Current Ramp Rate Setting Query
Input:
RATE? [term]
Returned: <rate> [term]
+n.nnnn (Refer to command for description)
Format:
RDGI?
Current Output Reading Query
Input:
RDGI? [term]
Returned: <current> [term]
Format:
±nn.nnnn
<current>
Computer Interface Operation
Lake Shore Model 642 Electromagnet Power Supply User's Manual
Refer to Paragraph 5.1.4.2.2 for a list of operational status bits.
Specifies the rate at which the current will ramp at when a new output current setting is
entered: 0.0001 – 99.999 A/s.
Actual measured output current.
5-35

Advertisement

Table of Contents
loading

Table of Contents