4.2

DISPLAY DEFINITION

The Model 642 has an 8 line by 40 character vacuum fluorescent (VF) display capable of showing both text and graphic
images. The features displayed during normal operation include current measurement, voltage measurement, current
programming, ramp rate, magnet water status, internal water status, program mode, and internal temperature. Other
display configurations appear during parameter setting and data entry operations. These displays are illustrated in their
individual operation paragraphs. A typical display is shown below.
4.3

LED ANNUNCIATORS

There are five LED annunciators on the front panel that are used to indicate the status of the instrument. These provide
easy verification of the operation of the instrument. See Figure 4-2 for LED locations.
Fault
Compliance
Power Limit
Ramping
Remote

4.3.1 Fault LED

The Fault LED will light whenever an error condition is encountered. It may also be accompanied by an alarm depending
on the fault. (Refer to Table 7-2 Instrument Hardware Errors, and Table 7-3 Operational Errors.)

4.3.2 Compliance LED

The Compliance LED lights whenever the maximum output voltage is reached. This can happen when attempting to
rapidly ramp a magnet with higher than usual voltage required to overcome the magnet's inductance. The led will go out
when the condition clears.

4.3.3 Power Limit LED

The Model 642 has a hardware power limit to protect the internal power MOSFETs. If the power supply is driving a load
which has a resistance lower than the supply's rated minimum, the power required may be higher than the devices can
safely handle. If this happens, the power in the devices is prevented from exceeding the safe limit and the Power Limit
LED will light to alert the operator to the condition. The led will go out when the condition clears.

4.3.4 Ramping LED

The Ramping LED lights whenever the internal control circuitry is changing the output current. When the ramp is
completed and the current is at the desired point, the LED goes out. The LED does not light when the output current is
being controlled by an external source.

4.3.5 Remote LED

The Remote LED is lit when the Remote key has been pressed to accept remote computer programming input, or upon
receiving the first command over the IEEE bus. When the LED is lit, the main keypad is locked out. Pressing the Local
key will return the unit to the local mode and reestablish keypad functions.
4-2
Lake Shore Model 642 Electromagnet Power Supply User's Manual
Output I
+70.0000
Set: +70.0000 A
Magnet Water: Off
Int Water: Off
Table 4-1. Model 642 LED Descriptions
On when a hardware fault condition exists, blinking when a soft fault condition exists.
On when maximum compliance voltage is reached.
On when power in internal devices reaches the maximum limit.
On when output current is ramping, blinking when ramp is paused.
On when instrument is in remote computer interface mode.
Output V
+35.000
A
Rate:0.0001 A/s
Pgm Mode: Internal
Int Temp: 27.5 ±C
V
Operation