Digital Blood Pressure Monitor
Model UA-854
Instruction Manual
Manuel d
'
instructions
Manual de Instrucciones
Manuale di Istruzioni
Bedienungsanleitung
使用手册
1WMPD4001741B

Advertisement

Table of Contents
loading

Summary of Contents for A&D UA-854

  • Page 1 Digital Blood Pressure Monitor Model UA-854 Instruction Manual Manuel d ’ instructions Manual de Instrucciones Manuale di Istruzioni Bedienungsanleitung 使用手册 1WMPD4001741B...
  • Page 2: Table Of Contents

    Contents Dear Customers..............2 Preliminary Remarks ............. 2 Precautions................2 Parts Identification..............3 Symbols................4 Operation Mode..............5 Using The Monitor ..............6 Installing / Changing The Batteries ........6 Connecting The Air Hose ..........6 Connecting The AC Adapter ..........6 Adjusting The Built-in Clock ..........
  • Page 3: Dear Customers

    Dear Customers Congratulations on purchasing a state-of-the-art A&D blood pressure monitor, one of the most advanced monitors available today. Designed for ease of use and accuracy, this monitor will facilitate your daily blood pressure regimen. We recommend that you read through this manual carefully before using the device for the first time.
  • Page 4: Parts Identification

    Parts Identification AM Button MEMORY Button Display PM Button Air Connector Plug Arm Cuff START Button Air Socket Air Hose (Orange illumination during the AM) (Blue illumination during the PM) Battery Cover Battery Compartment 1.5V Batteries DC Jack (R6P, LR6 or AA) Part of Display MEMORY Clock display...
  • Page 5: Symbols

    Symbols Symbols Function / Meaning Recommended Action Standby and Turn the device on. Battery installation guide Direct current Serial number Date of manufacture 2007 Type BF: Device, cuff and tubing are designed to provide special protection against electrical shocks. The indicator that appears while Measurement is in measurement is in progress.
  • Page 6: Operation Mode

    Operation Mode 1. Normal Measurement With Storing Data Press the START button. Blood pressure is measured and data is stored within the device. The device can store the last 90 sets of data in memory. 2. Recalling The Data Press the MEMORY button to recall the data. Averaged systolic The average data of all the measurements Averaged diastolic...
  • Page 7: Using The Monitor

    Using The Monitor Installing / Changing The Batteries 1. Slide the battery cover to open it. Step 2 2. Remove the used batteries and insert batteries into battery compartment as shown, taking care that the polarities (+) and (-) are correct. 3.
  • Page 8: Adjusting The Built-In Clock

    Using The Monitor Adjusting The Built-in Clock Press and hold Adjust the clock prior to use. Year 1. Press and hold the AM button until the year starts blinking. 2. Select the year using the PM button. Month Press the AM button to set the current year and move to month/day selection.
  • Page 9: Alarm On/Off Mark On The Display

    Using The Monitor Alarm ON/OFF Mark On The Display When the alarm is on, a “ ” appears on the LCD. To turn the alarm off, simply press the AM button once. The alarm will sound for 1 minute. To stop the alarm sound, press any button once. Rolled up shirtsleeve Attaching The Arm Cuff 1.
  • Page 10: Measurements

    Measurements Model UA-854 is designed to detect the pulse and to inflate the cuff to a systolic pressure level automatically. If your systolic pressure is expected to exceed 230 mmHg, read "Measurement With The Desired Systolic Pressure" on the next page.
  • Page 11: Measurement With The Desired Systolic Pressure

    Measurements Measurement With The Desired Systolic Pressure If your systolic pressure is expected to exceed 230 mmHg, use this procedure. Place the cuff on the arm (preferably the left arm). At heart level Press and hold the button until a START number about 30 to 40 mmHg higher Press and hold the button...
  • Page 12: Recalling Memory Data

    Recalling Memory Data Note: This device stores the last 90 measurements in memory. Clock display 1. Press the MEMORY button. The average data of all the measurements Averaged systolic and the number of data are displayed. Averaged diastolic Number of data Average of all the data Press...
  • Page 13: Recalling Memory Data

    Recalling Memory Data 4. Press the following key to recall various data when the average data of all the measurements is displayed. AM button To recall the average data during the AM. The START button illuminates in orange. Then, each time the AM button is pressed, the memory data taken during the AM is displayed from the most recent.
  • Page 14: What Is An Irregular Heartbeat

    What Is An Irregular Heartbeat Model UA-854 blood pressure monitor provides a blood pressure and pulse rate measurement even when an irregular heartbeat occurs. An irregular heartbeat is defined as a heartbeat that varies by 25% from the average of all heartbeats during the blood pressure measurement. It is important that you be relaxed, remain still and do not talk during measurements.
  • Page 15: About Blood Pressure

    About Blood Pressure What Is Blood Pressure? Blood pressure is the force exerted by blood against the walls of the arteries. Systolic pressure occurs when the heart contracts. Diastolic pressure occurs when the heart expands. Blood pressure is measured in millimeters of mercury (mmHg).
  • Page 16: Troubleshooting

    described in this manual to get to know your normal blood pressure. Regular readings give a more comprehensive blood pressure history. Be sure to note the date and time when recording your blood pressure. Consult your doctor to interpret your blood pressure data. Troubleshooting Problem Possible Reason...
  • Page 17: Maintenance

    However it is generally recommended to have the device inspected every 2 years, to ensure proper functioning and accuracy. Please contact either your authorized dealer or A&D for maintenance. Technical Data Type UA-854 Measurement method Oscillometric measurement Measurement range Pressure: 20 - 280 mmHg...
  • Page 18: Emc Table

    Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the following. Portable and mobile RF communication equipment (e.g. cell phones) can affect Medical Electrical Equipment. The use of accessories and cables other than those specified (other than boso original parts) may result in increased emissions or decreased immunity of the unit.
  • Page 19 Guidance and manufacturer’s declaration – electromagnetic immunity The A&D unit is intended for use in the electromagnetic environment specified below. The customer or the user of the A&D unit should assure that it is used in such an environment. Immunity IEC 60601 Compliance Electromagnetic environment –...
  • Page 20 Guidance and manufacturer’s declaration – electromagnetic immunity The A&D unit is intended for use in the electromagnetic environment specified below. The customer or the user of the A&D unit should assure that it is used in such an environment. Immunity IEC 60601 Compliance level Electromagnetic test...
  • Page 21 A&D Company, Limited 1-243 Asahi, Kitamoto-shi, Saitama 364-8585 JAPAN Telephone: [81] (48) 593-1111 Fax: [81] (48) 593-1119 A&D INSTRUMENTS LTD. Unit 24/26 Blacklands Way, Abingdon Business Park, Abingdon, Oxfordshire OX14 1DY United Kingdom Telephone: [44] (1235) 550420 Fax: [44] (1235) 550485 A&D Australasia Pty Ltd.

Table of Contents