D. AC Current
Range
600µA
6000µA
60mA
600mA
10A
Remarks:
l
At
5A range: Continuous measurement is allowed.
l
At >5A range : For continuous measurement 10 seconds and interval not less than 15 minutes.
l
Displays:
True RMS (applicable to the range of 10%~95%)
AC peak factor 3.0.
Input short circuit has around less than 30 remaining digits, it will not affect accuracy.
AC+DC measurement accuracy = range's accuracy + 1%
42
Resolution
0.1µA
40Hz~10kHz: (1.0%+5)
1µA
>10kHz~15kHz: (2%+5)
0.01mA
40Hz~10kHz: (1%+5)
0.1mA
>10kHz~15kHz: (3%+5)
40Hz~5kHz: (2.0%+6)
10mA
Model UT803: OPERATING MANUAL
Accuracy
Overload Protection
Fuse 500mA, 250V, fast
type, 5x20mm.
Fuse 10A,250V,
fast type, 5x20mm.