Getting Started

Some helpful Tips

Running Internal Alignments

N O T E

Enable the Options

20
Please refer to the following descriptions for your further
use, including alignment, external reference, firmware update
and option activation.
To meet specifications, the analyzer must be manually
aligned.
When the alignment runs, you will hear clicking as the attenuator
settings changing. This is not an indication of trouble.
Manually Performing an Alignment
1
Connect a BNC cable with N- BNC adapter between the CAL
OUT and RF IN connector.
2
After warm- up, press
Please refer to
You are required to enter the option license key to enable
the option. Contact your nearest Agilent Office for
purchasing a license. Please refer to the procedures below to
activate the options you have purchased. (option TG3 and
option PA3)
1
> More > More > Licensing > Option
Preset/
Press
System
2
Enter the option number which you want to enable, Press
[Enter] to confirm your input.
3
> More > More > Licensing > License key
Preset/
Press
System
4
Enter
and terminate
enabled immediately.
> Alignment > Align > All.
Preset/
System
"Alignment"
on page 93 for details.
the license input. The option will be
N9320A Spectrum Analyzer