Chapter 3
This query returns one of the following values:
Return
Value
0
New constants are in RAM.
1
More measurements are needed.
5
A return value of 5 (calibration aborted) is given if either:
a. The command is received when the module is in current calibration or
resistance calibration mode, or
b. The calibration in process has entered more than 50 values without a
successful calibration.
When the channel voltage calibration is completed, the new calibration constants
are used by the module, but not stored in non-volatile memory.
If a voltage calibration is interrupted before completion (that is, by another
command being received), no constants are changed.
Refer to Appendix D, "Voltage/Current Output Adjustment", starting on page 155,
for additional information about this command.
Related Commands:
CALibration:STORe
Agilent E1418A SCPI Command Reference
Meaning
Calibration is complete.
Calibration aborted.
,
CALibration
CALibration
n
:CONFigure:VOLTage?
75