Agilent Technologies E1418A User Manual page 111

8/16-channel d/a converter module scpi programming guide
Table of Contents

Advertisement

Command
STATus
:OPERation:CONDition?
:OPERation:ENABle <
:OPERation:ENABle?
:OPERation[:EVENt]?
:PRESet
:QUEStionable:CONDition?
:QUEStionable:ENABle <
:QUEStionable:ENABle?
:QUEStionable[:EVENt]?
SYSTem
:ERRor?
:VERSion?
TEST
:NUMBer? <
:TST[:RESults]?
*TST?
TRIGger
[:IMMediate]
:SOURce <
:SOURce?
ABORt
INITiate[:IMMediate]
Chapter 3
Agilent E1418A Command Quick Reference
mask
>
mask
>
number
>
source
>
Returns the contents of the Operation Status Condition
Register. Reading the register does not affect its contents.
Specifies which bits of the associated Event Register are
included in bit 7 of the Status Byte Register. The Summary
bit is the bit-for-bit logical AND of the Event Register and the
unmasked bit(s).
Returns the value of the mask set for the Operation Status
Register.
Returns the value of the contents of the Operation Status Event
Register. Reading the register clears all bits in the register.
Presets all Enable Registers to 0. The command has no
effect on Condition or Event Registers.
Returns the value of the Questionable Status Condition
Register. Reading the register does not affect its contents.
Specifies which bits of the associated Event Register are
included in bit 3 of the Status Byte Register. The Summary
bit is the bit-for-bit logical AND of the Event Register and the
unmasked bits.
Returns the mask set for the Questionable Status Enable
Register.
Returns the value of the Questionable Status Event
Register. Reading the register clears all bits in the register.
Returns an error number and message from the
instrument's error queue.
Returns the SCPI version with which the driver complies.
Performs a single internal self-test on the module and
returns a 16-bit integer indicating the results of the self-test.
Returns an array showing the result of all self-tests
performed by the *TST? query.
Performs all internal self-tests on the module and returns a
16-bit integer indicating the results of the self-test.
Triggers any channels waiting for trigger when the trigger
source is set to either BUS or HOLD.
Sets the trigger source for the module.
Queries the trigger source for the module.
Disables the trigger system and places the module in the
idle state.
Changes the trigger system from the idle state to the
wait-for-trigger state.
Agilent E1418A SCPI Command Reference
Description
109

Advertisement

Table of Contents
loading

Table of Contents