Agilent Technologies E1418A User Manual page 134

8/16-channel d/a converter module scpi programming guide
Table of Contents

Advertisement

Using the Trigger
Control Register
132
Agilent E1418A Register-Based Programming
Trigger Source: This field selects the trigger source for trigger-enabled
channels.
Trigger Source Selected
TTLTRIG0*
TTLTRIG1*
TTLTRIG2*
TTLTRIG3*
TTLTRIG4*
TTLTRIG5*
TTLTRIG6*
TTLTRIG7*
SOFTWARE TRIGGER
EXTERNAL TRIGGER
To set up the Trigger Control Register for the VXIbus trigger types:
SYNC mode: This mode allows the card to receive a trigger from one of
the VXI backplane TTL trigger lines. To set the VXIbus SYNC trigger
mode, the Trigger Mode field is set to SYNC (00) and the Trigger Source
field is set to one of the *TTLTRIG lines. The Trigger Out field is not used.
OUTPUT mode: This mode allows the card to receive a trigger from one
TTL trigger line and also source a trigger on another TTL trigger line. The
output mode is not a VXIbus defined trigger mode. The output mode is
used to synchronize additional modules. The Trigger Mode field is set to
OUTPUT (11), the Trigger Source field is set one of the *TTLTRIG lines,
and the Trigger Out field is set to a different *TTLTRIG line.
Bits
03
02
01
00
0
0
0
0
0
0
0
1
0
0
1
0
0
0
1
1
0
1
0
0
0
1
0
1
0
1
1
0
0
1
1
1
1
0
0
0
1
0
0
1
Appendix B

Advertisement

Table of Contents
loading

Table of Contents