311Xl Series; Introduction; Description; Method Of Analysis - Teledyne 311XL Series Instruction Manual

0-2 ppm full scale trace oxygen analyzer
Table of Contents

Advertisement

1.

INTRODUCTION

1.1
Description: The Teledyne Analytical Instruments (TAI) Model 311XL and
311TCXL are portable, intrinsically safe trace oxygen analyzer which can be operated without
an external power source. These analyzers will be referred to as Model 311XL or 311XL Series
analyzers in this manual.
The instrument provides for trace oxygen analysis in decade steps ranging from 0-2 to 0-10,000
ppm (full scale) plus a special calibration range that encompasses the known oxygen
concentration of atmospheric air (209,000 ppm).
Sample oxygen is read from an extremely accurate integral meter whose range of measurement is
determined by the position of the range selector switch. The linear 5 inch scale (mirror equipped
to eliminate parallax) provides excellent resolution and accuracy.
Sample gas is introduced and vented via a pair of quick disconnect fittings that feature integral
shutoff valves which automatically close when the mating male fitting is withdrawn. The fittings
are an integral part of the measuring cell manifold so that internal sample passage volume is at
an absolute minimum. Sample flow control of 0.5 to 5 liters/min. must be accomplished with
accessory equipment.
1.2

Method of Analysis.

electrochemical transducer which functions as a fuel cell. Oxygen diffusing into the cell reacts
electro-chemically to produce an electrical current that is proportional to the oxygen
concentration in the gas phase immediately adjacent to the transducer's sensing surface. The
linear, but minute, signal produced by the transducer from trace oxygen is amplified by a two
stage amplifier. The dual stages of amplification provide enough gain to drive meter and
thermistor controlled network utilized to compensate for the positive temperature co-efficient of
the transducer.
1.3

Outstanding Features.

the Model 311XL:
1.3.1 Micro-Fuel Cell: The Micro-Fuel Cell* is a sealed electrochemical trans-
ducer with no electrolyte to change or electrodes to clean. When the cell reaches the end of its
useful life, it is merely thrown away and replaced, as one would replace a worn out battery in a
flashlight.

311XL SERIES

TRACE OXYGEN ANALYZER
The sample oxygen is measured by a unique
The following unique features are incorporated into
-1-

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents