Fluke 8508A Service Manual page 62

Reference multimeter
Hide thumbs Also See for 8508A:
Table of Contents

Advertisement

8508A
Service Manual
Zero Point
1. On the Multimeter, select the required range.
2. On the Calibrator, select zero
3. On the Multimeter, press
4. Press
Range Gain Point
1. On the Calibrator, select Range Gain point Output value. Alternatively if using a
2. On the Multimeter, Press
3. Select
4. Use the numeric keys with the
5. Press
6. On the Calibrator, set
7. Press the
Tru
LoI Mode Calibration
1. Press the
2. Press the
Range Zero and Gain Calibration Procedure
After the initial setup and connecting up, use the following procedure to calibrate range
zero and range gain for the ranges and points detailed in Table 3-5 for the Tru LoI
mode. Just one range can be calibrated if required, but for a full calibration start with the
2
On each range, the Multimeter automatically recognizes the appropriate value as range
zero or range gain from the applied resistance value.
It is unlikely that the resistance standard will be at exactly the nominal
value required. The Set feature allows a user to enter the true output value
of the calibration standard where it differs from nominal range gain point.
If the resistance value is exactly the nominal value required, omit steps 2-4
for the Range Gain Point.
3-22
resistor connect the standard resistor for a 4-wire resistance zero.
to initiate the calibration operation.
SAMPLE
Calibration is complete when the Busy legend goes out and the calibrated
measurement is displayed.
standard resistor re-connect the standard resistor to measure its resistance.
on the
Set
CALIBRATION menu
the calibrator (or the value of the resistance standard) at the range gain value, then
press Enter.
to initiate the calibration operation.
SAMPLE
Calibration is complete when the Busy legend goes out and the calibrated
measurement is displayed.
STANDBY
key to revert to the ranges menu.
key and select
Plus
key and select
CONFIG
range and work up to the 20 k range.
and Operate. Alternatively if using a standard
to enter the
CAL
CALIBRATION
to revert to
CAL
CALIBRATION
.
menu to key in the true output value of
SET VALUE
.
From the
Tru .
Tru
,
and
Filt
LoI
4w
Note
menu.
menu.
menu select the
range.
2
.