Fluke 8508A Service Manual page 24

Reference multimeter
Hide thumbs Also See for 8508A:
Table of Contents

Advertisement

8508A
Service Manual
Range
200 A, 2 mA &
20 mA
200 mA
2 A
20 A
Type........................................................................ True rms, AC coupled. DC coupled gives
Crest Factor........................................................... 3:1 at 50 % range, 1.5:1 at full range
Protection
Front Input............................................................... 20 A rms
Rear Input ............................................................... 2 A rms, Rear Panel Fuse
Settling Time
100 Hz filter............................................................. < 0.5 s
40 Hz filter............................................................... < 1.25 s
10 Hz filter............................................................... < 5 s
1 Hz filter................................................................. < 50 s
1-14
AC Current (Secondary Specifications)
Temperature Coefficient
Frequency (Hz)
15 °C - 30 °C
± ppm Reading/°C
1 - 10
10
10 - 10 k
10
10 k - 30 k
12
30 k - 100 k
40
1 - 10
10
10 - 10 k
15
10 k - 30 k
15
10 - 2 k
10
2 k - 10 k
15
10 k - 30 k
20
10 - 2 k
20
2 k - 10 k
25
5 °C - 15 °C
30 °C - 40 °C
Range
15
200 A
15
2 mA
20
20 mA
60
15
20
20
15
20
30
25
30
200 A to 200 mA Ranges 2 A Range
to 100 ppm of step
to 100 ppm of step to 300 ppm of step
< 1 s
< 2 s
< 10 s
< 50 s
[1] [2] [6] [9]
Input Impedance ( )
Front
150
150
15.2
15.2
1.8
1.9
1.2
1.3
0.3
0.4
0.04
-
2
2
(ac
dc
)
20 A Range (at 10 A)
< 1 s
< 2 s
< 10 s
< 50 s
Rear