Auto-Reference Function; Main Specifications; Remote Zero-Adjustment Function; Cautions - Panasonic DP-100 series Instruction Manual

High-performance digital display
Hide thumbs Also See for DP-100 series:

Advertisement

12

AUTO-REFERENCE FUNCTION

(ONLY HIGH-FUNCTION TYPE)
The auto-reference function corrects
the setting value using the detected
pressure value during auto-reference
input as the reference pressure.
Using the detected pressure value at
auto-reference input P(a) as a refer-
ence, the set value ' is automatical-
ly corrected to 'set value
+ P(a)'.
Settable range and set pressure range after correction
The setting pressure range is wider than the rating pressure range so
that the auto-reference function can be handled.
If the corrected setting value exceeds the set pressure range when auto-reference input is
carried out, the setting value will be automatically corrected to within the set pressure range.
Thus, take care not to exceed the set pressure range.
Operation chart
During normal operation (each
comparative output set to N.O.)
ON
Output
OFF
Applied pressure (kPa
0
10 20 30 40 50 60
)
Displayed value (kPa)
0
10 20 30 40 50 60
Set value (kPa)
10 20
Note: The setting values shift in the same manner during the EASY mode or the win-
dow comparator mode.
The detected pressure value at auto-reference input becomes 'zero'
when the setting of the analog voltage output / external input selection
function is changed or the power is turned ON again.
The auto-reference input value can be checked when setting the threshold value
in RUN mode. Refer to the threshold value setting in '
13

REMOTE ZERO-ADJUSTMENT FUNCTION

(HIGH-FUNCTION TYPE)
The remote zero-adjustment function forcibly sets the pressure value to
'zero' when the external signal is input.
The setting value is not corrected when remote zero-adjustment is input.
Make sure that the pressure and setting value during remote zero-adjust-
ment do not exceed the settable pressure range.
Operation chart
During normal operation (each
comparative output set to N.O.)
ON
Output
OFF
Applied pressure (kPa
)
0
10 20 30 40 50 60
Displayed value (kPa)
0
10 20 30 40 50 60
Set value (kPa)
10 20
Note: The setting values shift in the same manner during the EASY mode or the win-
dow comparator mode.
The remote zero-adjustment value is cleared when the setting of the analog
voltage output / external input selection is changed or the power is turned ON
again, and normal operation based on the atmospheric pressure is resumed.
The remote zero-adjustment value can be confirmed when setting the
threshold value in RUN mode. Refer to the threshold value setting in
'
8
RUN MODE'.
14
ERROR INDICATION
Cause
Error message
The load is short-circuited caus-
ing an overcurrent to flow.
Pressure is applied during
zero-point adjustment.
External input is carried out out-
side the rated pressure range.
Communication error
(Disconnection, faulty connection, etc.)
Communication error
(Incorrect model.)
The applied pressure exceeds the up-
per limit of the display pressure range.
The applied pressure exceeds the lower limit (re-
verse pressure) of the display pressure range.
Before auto-
reference
A f t e r a u t o -
reference
Atmospheric pressure
P(a)
Set value
Set value '
Pressure
Set value
Detected pressure value
reference is:
at auto-reference input
'
During remote zero-adjustment input
(each comparative output set to N.O.)
Detected pressure at auto-reference input: 10kPa
Output mode: Hysteresis mode
ON
Output
OFF
Applied pressure (kPa
0
10 20 30 40 50 60
)
Displayed value (kPa)
0
10 20 30 40 50 60
P(a)
Set value (kPa)
10 20 30
RUN MODE' for details.
8
During remote zero-adjustment input
(each comparative output set to N.O.)
Detected pressure at remote zero-adjustment input: 10kPa
Output mode: Hysteresis mode
ON
Output
OFF
Applied pressure (kPa
)
0
10 20 30 40 50 60
Displayed value (kPa)
-10
0
10 20 30 40 50
Set value (kPa)
10 20
Corrective action
Turn the power OFF and check the load.
Applied pressure at the pressure port should be brought to atmos-
pheric pressure and zero-point adjustment should be done again.
Applied pressure range should be brought within
the rated pressure range.
Check the wiring when using the copy function.
Make sure that the system is configured of the
same models when using the copy function.
Applied pressure range should be brought within
the rated pressure range.
14

MAIN SPECIFICATIONS

Model
ON
DP-10
- - -
OFF
ON
OFF
' after auto-
P(a)
Item
Pressure type
Rated pressure range
Set pressure range
Withstand pressure
Applicable fluid
Supply voltage
Power consumption
Comparative output
Output operation
Hysteresis
Repeatability
Response time
Analog voltage output
External input
Ambient temperature
Ambient humidity
Temperature characteristics
Material
Weight
Accessories
15

CAUTIONS

DP-100 series is designed for use with non-corrosive gas. It cannot be
used for liquid or corrosive gas.
Use within the rated pressure range.
Do not apply pressure exceeding the pressure withstandability value. The
diaphragm will get damaged and correct operation shall not be maintained.
Make sure that the power supply is off while wiring.
Take care that wrong wiring will damage the sensor.
Verify that the supply voltage variation is within the rating.
If power is supplied from a commercial switching regulator, ensure that the frame
ground (F.G.) terminal of the power supply is connected to an actual ground.
In case noise generating equipment (switching regulator, inverter motor, etc.) is used in the vicini-
ty of this sensor, connect the frame ground (F.G.) terminal of the equipment to an actual ground.
Do not use during the initial transient time (0.5 sec.) after the power supply is switched on.
Do not run the wires together with high-voltage lines or power lines or put
them in the same raceway. This can cause malfunction due to induction.
The specification may not be satisfied in a strong magnetic field.
Avoid dust, dirt, and steam.
Take care that the sensor does not come in direct contact with water,
oil, grease, or organic solvents, such as, thinner, etc.
Do not insert wires, etc, into the pressure port. The diaphragm will get
damaged and correct operation shall not be maintained.
Do not operate the keys with pointed or sharp objects.
Make sure that stress by forcible bend or pulling is not applied directly to the sensor cable joint.
This is a CE conformity product complying with EMC Directive. The standard with regard to im-
munity that applies to this product is EN 61000-6-2, and in order to meet the standard, every ca-
ble connected to this product must be within 10m with 0.3mm
CE conformity is not required, the cable length can be up to 100m with 0.3mm
SUNX Limited
Head Office
2431-1 Ushiyama-cho, Kasugai-shi, Aichi, 486-0901, Japan
Phone: +81-(0)568-33-7211 FAX: +81-(0)568-33-2631
Overseas Sales Dept.
Phone: +81-(0)568-33-7861 FAX: +81-(0)568-33-8591
None: Cable with connector enclosed, J: No cable with connector
None: NPN output type, P: PNP output type
None: R
1 8
+M5 female screw, E: G
1 8
+M5 female screw, N: NPT
None: Standard type, A: High-function type
1: Low-pressure type, 2: High-pressure type
Type
Standard type
Low-pressure type High-pressure type
Gauge pressure
-100 to +100 kPa
-0.1 to +1.0 MPa
-100 to +100 kPa
-0.1 to +1.0 MPa
500 kPa
1.5 MPa
Non-corrosive gas
12 to 24 V DC
10% Ripple P-P 10 % or less
Normal operation: 840mW or less (current consumption 35mA or less at 24V supply voltage)
ECO mode (STD): 600mW or less (current consumption 25mA or less at 24V supply voltage)
ECO mode (FULL): 480mW or less (current consumption 20mA or less at 24V supply voltage)
<NPN output type>
• NPN open-collector transistor
• Maximum sink current: 100mA
• Applied voltage: 30V DC or less
(between comparative output and 0V)
• Residual voltage: 2V or less
(at 100mA sink current )
Selectable either N.O. or N.C., with key operation
Min. 1 digit (variable) (however, 2 digits when using psi units)
0.1% F.S.
0.2% F.S.
within 2 digits
within 2 digits
2.5ms, 5ms, 10ms, 25ms, 50ms, 100ms, 250ms, 500ms, 1.000ms
or 5,000ms selectable with key operations
<High-function, low-pressure type>
• Output voltage: 1 to 5V
• Zero point: Within 3V
5% F.S.
• Span: Within 4V
5% F.S.
• Linearity: Within
1% F.S.
• Output impedance: 1kΩ approx.
<High-function NPN output type>
• ON voltage: 0.4V DC or less
• OFF voltage: 5 to 30V DC or open
• Input impedance: 10kΩ approx.
• Input time: 1ms or more
-10 to +50°C (No dew condensation or icing allowed), Storage: -10 to +60°C
35 to 85% RH, Storage: 35 to 85% RH
±1% F.S.
±0.5% F.S.
(20°C reference)
(20°C reference)
Enclosure: PTB (with glass fiber), LCD display: Acrylic, Pressure port: Brass
(nickel-plated), Mounting screw section: Brass (nickel-plated), Switch: Silicon rubber
40g approx.(DP-100-E type: 45g approx.) (Main body only)
CN-14A-C2 (Cable with a connector, 2m long) optional for J type), Unit switching label: 1 pc.
WARNING
1 8
+M5 female screw
High-function type
Low-pressure type
High-pressure type
-100 to +100 kPa
-0.1 to +1.0 MPa
-100 to +100 kPa
-0.1 to +1.0 MPa
500 kPa
1.5 MPa
<PNP output type>
• PNP open-collector transistor
• Maximum source current: 100mA
• Applied voltage: 30V DC or less
(between comparative output and +V)
• Residual voltage: 2V or less
(at 100mA source current )
0.1% F.S.
0.2% F.S.
within 2 digits
within 2 digits
<High-function, high-pressure type>
• Output voltage: 0.6 to 5V
• Zero point: Within 1V
5% F.S.
• Span: Within 4.4V
5% F.S.
• Linearity: Within
1% F.S.
• Output impedance: 1kΩ approx.
<High-function PNP output type>
• ON voltage: 5V to +V DC
• OFF voltage: 0.6V DC or less or open
• Input impedance: 10kΩ approx.
• Input time: 1ms or more
±0.5% F.S.
±1% F.S.
(20°C reference)
(20°C reference)
2
, or more, cable. However, in case
2
, or more, cable.
http://www.sunx.co.jp/
PRINTED IN JAPAN