Single channel current transducer system (14 pages)
Summary of Contents for Signaltec CT 1000
Page 1
TPS II Single Channel Current Transducer System Installation Manual MA TPS II CT, Ver. 1.0 11/2021 SIGNALTEC GmbH, Reutersbrunnenstrasse 27, D-90429 Nürnberg, GERMANY, T. +49 911 2550993-0, F. +49 911 2550993-13, Email: info@signaltec.com, Web: www.signaltec.com...
Page 2
7.4.1. Passive plug-on burden resistors 7.4.2. Active plug-on voltage amplifiers Switching off and deinstallation Accessories MA TPS II CT, Ver. 1.0 11/2021 SIGNALTEC GmbH, Reutersbrunnenstrasse 27, D-90429 Nürnberg, GERMANY, T. +49 911 2550993-0, F. +49 911 2550993-13, Email: info@signaltec.com, Web: www.signaltec.com...
Page 3
Please compare your order papers with our packing list and the received goods. MA TPS II CT, Ver. 1.0 11/2021 SIGNALTEC GmbH, Reutersbrunnenstrasse 27, D-90429 Nürnberg, GERMANY, T. +49 911 2550993-0, F. +49 911 2550993-13, Email: info@signaltec.com, Web: www.signaltec.com...
Page 4
For this purpose, passive plug-on burden resistors and active plug-on output voltage amplifiers are available. MA TPS II CT, Ver. 1.0 11/2021 SIGNALTEC GmbH, Reutersbrunnenstrasse 27, D-90429 Nürnberg, GERMANY, T. +49 911 2550993-0, F. +49 911 2550993-13, Email: info@signaltec.com, Web: www.signaltec.com...
Page 5
Status-LED red: Transducer overloaded or interruption in the output loop, switch primary current off! MA TPS II CT, Ver. 1.0 11/2021 SIGNALTEC GmbH, Reutersbrunnenstrasse 27, D-90429 Nürnberg, GERMANY, T. +49 911 2550993-0, F. +49 911 2550993-13, Email: info@signaltec.com, Web: www.signaltec.com...
Page 6
CT 2000 2000 A MCTS/VM1/0.02 2000 A / 7000 mV 3.5 mV / A MA TPS II CT, Ver. 1.0 11/2021 SIGNALTEC GmbH, Reutersbrunnenstrasse 27, D-90429 Nürnberg, GERMANY, T. +49 911 2550993-0, F. +49 911 2550993-13, Email: info@signaltec.com, Web: www.signaltec.com...
Page 7
14 - 1 green - white * A Status-Readout-Interface cable TPS/ROC is available as an option MA TPS II CT, Ver. 1.0 11/2021 SIGNALTEC GmbH, Reutersbrunnenstrasse 27, D-90429 Nürnberg, GERMANY, T. +49 911 2550993-0, F. +49 911 2550993-13, Email: info@signaltec.com, Web: www.signaltec.com...
Page 8
The maximum power consumption of a TPS with transducer IN 2000-S at 2000 A primary current is around 30 W. MA TPS II CT, Ver. 1.0 11/2021 SIGNALTEC GmbH, Reutersbrunnenstrasse 27, D-90429 Nürnberg, GERMANY, T. +49 911 2550993-0, F. +49 911 2550993-13, Email: info@signaltec.com, Web: www.signaltec.com...
Page 9
Maximum measuring resistance (cable- + burden-resistance) depending on primary rms current and temperature CT 100 CT 200 CT 400 CT 500 CT 1000 CT 2000 MA TPS II CT, Ver. 1.0 11/2021 SIGNALTEC GmbH, Reutersbrunnenstrasse 27, D-90429 Nürnberg, GERMANY, T. +49 911 2550993-0, F. +49 911 2550993-13, Email: info@signaltec.com, Web: www.signaltec.com...
Page 11
MCTS/TPS/15/0.75 loop resistance: 1.42 Ω → Maximum allowed burden resistor = 4 Ω - 1.42 Ω = 2.58 Ω MA TPS II CT, Ver. 1.0 11/2021 SIGNALTEC GmbH, Reutersbrunnenstrasse 27, D-90429 Nürnberg, GERMANY, T. +49 911 2550993-0, F. +49 911 2550993-13, Email: info@signaltec.com, Web: www.signaltec.com...
Page 12
MCTS/VM0.66/0.02 1 mV/A @ 1000 A CT 2000 MCTS/VM0.66/0.02 0.5 mV/A @ 2000 A MA TPS II CT, Ver. 1.0 11/2021 SIGNALTEC GmbH, Reutersbrunnenstrasse 27, D-90429 Nürnberg, GERMANY, T. +49 911 2550993-0, F. +49 911 2550993-13, Email: info@signaltec.com, Web: www.signaltec.com...
Page 13
Before you switch off the TPS rack, open the output current loop or remove any cable between rack and sensor, make sure that the primary current is switched off. An unpowered transducer can be damaged. MA TPS II CT, Ver. 1.0 11/2021 SIGNALTEC GmbH, Reutersbrunnenstrasse 27, D-90429 Nürnberg, GERMANY, T. +49 911 2550993-0, F. +49 911 2550993-13, Email: info@signaltec.com, Web: www.signaltec.com...
Page 14
BNC to BNC test lead, length 1 m BNC4L1 BNC to 4 mm banana-plug test lead, length 1 m MA TPS II CT, Ver. 1.0 11/2021 SIGNALTEC GmbH, Reutersbrunnenstrasse 27, D-90429 Nürnberg, GERMANY, T. +49 911 2550993-0, F. +49 911 2550993-13, Email: info@signaltec.com, Web: www.signaltec.com...