INSTALLATION OF TEMPERATURE RECORDER
4. Remove the caps (both outside and inside) covering the access port for freezer and take out the
thermal insulation in the port. Pass the recorder sensor to the chamber through the port. Then replace
the thermal insulation and caps (both outside and inside). (Fig.13)
Note: Make a cut on the caps for access port to
pass the capillary tube as shown in the figure
below.
5. Fix the recorder sensor to the attachment hole on the left side of the freezer by using the middle size
clips enclosed with the recorder fitting (MPR-S30: optional component). (Fig. 14)
6. Arrange the capillary tube so that it cannot contact the shelf. Care should be taken for the following
points to avoid the crush of the tube when bending the capillary tube.
Do not bend the tube of 30 mm from the edge of
the sensor.
Do not bend and extend the tube repeatedly at
the same point.
The bending should be R10 R15.
27

Fig. 13

Fig. 14