ROUTINE MAINTENANCE

Replacement of fluorescent lamp

1. The fluorescent lamp is placed vertically at the
center of the frame (MPR1013 and MPR-1013R) and
horizontally at the top front of the cabinet (MPR-513
and MPR-513R)
2. Turn off the fluorescent lamp switch and disconnect
the power supply plug .
3. For MPR-513 and MPR-513R, remove the top shelf,
for MPR-1013 and MPR-1013R, remove all shelves in
the left side of the cabinet.
4. As shown in Fig. 1 and Fig. 2, pull the fluorescent
lamp towards the rear together with the cover and
wiring.
5. Take out the fluorescent lamp and remove the
connector water-proofing rubber at both ends.
shown in Fig. 3, shift the rubber by rolling it.
6. Pull out the fluorescent lamp. Pull gently from the
cover.
7. After replacing a new fluorescent lamp, return it to
the original position.
Recommended fluorescent lamp:
MPR-513, MPR-513R:
MPR-1013, MPR-1013R: FL40SD

Replacement of glow starter

The glow starter is located on the right side of top front for MPR-513 and MPR-513R, and center of top
front for MPR-1013 and MPR-1013R in the chamber.
1. Turn off the fluorescent lamp and then disconnect the power
supply cord of the refrigerator.
2. Remove the water-proof cover and take out the glow starter.
3. Attach a new lamp surely. The starter designation is as follows:
MPR-513, MPR-513R:
MPR-1013, MPR-1013R:
4. Replace the water-proof cover on the glow starter.
As
FL20SD
FG-1P
FG-4P
22
MPR-1013
MPR-1013R
Fig. 1
MPR-513
MPR-513R
Fig. 2
Fig. 3
glow starter
cover