ROUTINE MAINTENANCE

WARNING
Always disconnect the power supply to the unit prior to any repair or maintenance of the unit in
order to prevent electric shock or injury.
Ensure you do not inhale or consume medication or aerosols from around the unit at the time of
maintenance. These may be harmful to your health.
CAUTION
Always wear dry gloves to protect hands at the time of maintenance. Failure to wear gloves may
result in injury from edges and corners.
Cleaning of cabinet
1. Clean the unit once a month. Regular cleaning keeps the unit looking new.
2. Use a dry cloth to wipe off small amounts of dirt on the outside and inside of the unit and all
accessories. If the unit is very dirty, use a neutral detergent.
3. After cleaning, wipe away the cleaner completely with a cloth washed in clean water.
4. Never pour water onto or into the unit. Doing so can damage the electric insulation and cause failure.
The compressor and other mechanical parts are completely sealed. This unit requires absolutely no
lubrication.
Remove dust from the power supply plug periodically.
Cleaning of condenser filter (MPR-1013, 1013R)
1. Remove the screws on the bottom of the unit cover
and remove the cover as shown in the figure.
2. Slide the filter to the left with holding on the black
marking on the filter as shown in the figure.
3. Clean the filter with clean water and dry it
completely .
4. After returning the filter to the original position and
attach the unit cover.
20