Summary of Contents for Spectron BE55-2U

  • Page 1 Operating Manual for the Pressure Control Panels of the BM55-2U/BM56-2U BE55-2U/BE56-2U series BM55-2U BM56-2U with outlet adapter with outlet adapter and valve...
  • Page 2: Table Of Contents

    Operating Manual of the BM and BE series Contents Contents ........................2   Introduction ......................3     General ......................3     Description of the pressure control panels of the BM and BE series ....3     Intended use ....................4  ...
  • Page 3: Introduction

    All rights reserved. Updates No update service is provided for this operating manual by Spectron Gas Control Systems GmbH. Changes can be made to this operating manual without the need to notify anyone about them. Description of the pressure control panels of the BM and BE series Spectron pressure control panels are available in BM and BE versions.
  • Page 4: Intended Use

    Intended use Intended use The BM55-2U and BM56-2U pressure control panels described are intended for use with non-corrosive gases up to a purity grade of 6.0. The BE55-2U and BE56-2U pressure control panels are suitable, in addition, for corrosive gases up to a purity grade of 6.0. The permissible gases and pressure ranges are specified on the type plate.
  • Page 5: Personnel Requirements

    Operating Manual of the BM and BE series Introduction Personnel requirements Definition of an authorised person An authorised person is a person with a technical training who has received technical instruction about the system as a whole and the associated hazards – gas cylinder – gas type – gas cylinder valve – pressure regulator –...
  • Page 6: Safety Features

    Operating Manual of the BM and BE series For your safety  BGR 500 2.33 German trade association rules for the operation of gas handling systems  German technical rules for pressurised gases (TRG)  German technical rules for pipelines (TRR) ...
  • Page 7 Operating Manual of the BM and BE series For your safety Danger of death The pressure control panel must be suitable for the relevant gas and the pressure ranges involved. If pressure control panels are used that are not Only use for the gases indicated on the device. suitable for the relevant gas and pressure range, there is a risk of a fire or explosion occurring as If there are no gas types specified on the pressure...
  • Page 8: Description

    Operating Manual of the BM and BE series Description Overview of the pressure control panel Drawing of the pressure control panel BM/BE56-2U Elements of the pressure control panel Item Designation Correct functioning Pressure regulator Regulates the variable inlet pressure to achieve a set outlet pressure level.
  • Page 9 Operating Manual of the BM and BE series Description Flow chart of gas supply with a pressure control panel Legend Process gas valve with inlet filter Waste gas or purge gas valve Pressure regulator - single-stage Pressure regulator - two-stage Inlet pressure gauge Outlet pressure gauge Relief valve...
  • Page 10: Functional Description

    95/9/EEC (ATEX 95). You will find more detailed information at www.spectron.de . Another option is to equip the pressure control panel with pneumatically operated valves.
  • Page 11: Technical Data

     Note The technical data can be taken from the Spectron data sheet for the relevant product. If this is not available, you can view and download it at www.spectron.de. The maximum inlet and outlet pressures and the gas type are indicated on the type plate.
  • Page 12 Operating Manual of the BM and BE series Description How the operating pressure is adjusted The pressure control panel is set at the factory to a specific nominal operating pressure or average pressure (with the switching lever positioned in the middle). If necessary, this nominal operating pressure can be changed.
  • Page 13: Operation

    Note You can find instructions on how to install the pressure control panel in the installation manual MA_BM+BE. You can view and download this at www.spectron.de. Putting the pressure control panel into operation Important Before putting the equipment into operation for the first time, the entire pressure control panel must be purged via the process gas valve right up to the consumer.
  • Page 14: Changing The Cylinder

    Operating Manual of the BM and BE series Operation Step Activity Position the switching lever to specify which is the operational side and which is the reserve side. Pressure purge the system:  Slowly open the gas cylinder valves on the left and right. ...
  • Page 15: Taking The Equipment Out Of Operation

    Operating Manual of the BM and BE series Operation The procedure for changing a cylinder is described below, taking the right-hand gas cylinder by way of example: Step Activity Acknowledge the change of cylinder. Position the switching lever with the arrow marking pointing up. "In operation"...
  • Page 16: Problems

    Operating Manual of the BM and BE series Problems Problem/cause Remedy Gas is released from the relief valve. Close all valves immediately. Impermissible increase in the outlet Have the pressure control panel checked immediately by the pressure. manufacturer or an authorised specialist company. The pressure regulator is making noises.
  • Page 17: Maintenance, Cleaning And Repairs

    Repairs may only be carried out by specialist personnel in authorised repair workshops. After repairs, the entire pressure control panel must be checked in accordance with the original Spectron inspection instructions. Safe and reliable operation can only be guaranteed if original spare parts are used.
  • Page 18 Spectron Gas Control Systems GmbH Fritz-Klatte-Straße 8 D-65933 Frankfurt Germany Phone: +49 69 38016-0 Fax: +49 69 38016-200 Email: info@spectron.de Internet: www.spectron.de...