Clamp-On Ultrasonic Flow Meter Sensor
Operation & Maintenance Manual
Read the user's manual carefully before starting to use the unit.
Producer reserves the right to implement changes without prior notice.
1
25-0790 © Icon Process Controls Ltd.

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Truflo UltraFlo UF-500 and is the answer not in the manual?

Questions and answers

Summary of Contents for Icon Process Controls Truflo UltraFlo UF-500

  • Page 1 — UltraFlo UF-500 ® ® Clamp-On Ultrasonic Flow Meter Sensor Operation & Maintenance Manual Read the user's manual carefully before starting to use the unit. Producer reserves the right to implement changes without prior notice. 25-0790 © Icon Process Controls Ltd.
  • Page 2: Table Of Contents

    11. 4-20mA Calibration ......................................................10 12. Outside Dimensions .....................................................11 13. Minimum Flow Range ....................................................11 14. Application Example ....................................................11 15. Dimensions...........................................................12 16. Installation Positions .....................................................13 17. Maintenance ........................................................14 18. Troubleshooting .......................................................14 19. Warranty, Returns and Limitations ..............................................15 25-0790 © Icon Process Controls Ltd.
  • Page 3: Introduction

    24 VDC Connection Display OLED 128 * 64 Dot Matrix Teflon Epoxy ® Coated Aluminum Totalizer Units Gal | L | m �� Standards & Approvals PTFE Magnetic Connection RoHS Compliant (for easier installation) 25-0790 © Icon Process Controls Ltd.
  • Page 4: Installation

    The Coupling Gel DC-/4-20mA- Black coupling gel is provided with the package. Gray 4-20mA+ Option OCT+ RS485 A Blue OCT- RS485 B White Shield Coupling Pads Install Polycarbonate Cover. Connect M12 Cable. 25-0790 © Icon Process Controls Ltd.
  • Page 5: Operation

    Signal Quality (SQ) should ideally reach 99; minimum acceptable is 50. 5.2 Measurement Conditions The fluid must be free from bubbles and particles. Avoid measuring under turbulent or unstable flow. Maintain a back pressure of at least 0.3 MPa to prevent cavitation. 25-0790 © Icon Process Controls Ltd.
  • Page 6: Display Unit Features

    "00" is the minimum signal that could be detected and "99" represents the maximum. Normally, the transducer position should be adjusted repeatedly and coupling compound should be checked frequently until the signal quality detected is as strong as possible. 25-0790 © Icon Process Controls Ltd.
  • Page 7: Main Display Layout

    SQ99 12:30:18 Time 3.8 f/s Flow Velocity Flow 895.2 GPM Flow Rate 203119 gal Net Totalizer Velocity | | Totalizer 20-03-23 12:30 Date Time 27.7 Flow Rate Flow Velocity Unit 6311.1 gal Net Totalizer 25-0790 © Icon Process Controls Ltd.
  • Page 8: Quick Setup

    Step 3: Select 1. Wall Thickness and press to begin editing. Use the Up and Down keys to adjust the values as needed. Press to confirm the modification and return to the previous menu. 25-0790 © Icon Process Controls Ltd.
  • Page 9 Select 2. Set Zero. Press twice to begin the zeroing process. Do not interact with the device during this step. Once complete, it will automatically return to the previous menu. 25-0790 © Icon Process Controls Ltd.
  • Page 10: 20Ma Calibration

    During calibration, use the up and down keys to fine-tune the current output while observing the value on a connected ammeter (or multimeter). Refer to Figure for the wiring diagram. 25-0790 © Icon Process Controls Ltd.
  • Page 11: Outside Dimensions

    To achieve highest accuracy, position sensor at least NC-25 Flow Sensor 10D pipe prior, and Flow/No Flow 5D pipe after NC-30 Sensor Liquid Level Sensing through Plastic Wall Chemical Tank 25-0790 © Icon Process Controls Ltd.
  • Page 12: Dimensions

    58 - 64 126 (Max.) UF500-80 3" 86 - 92 154 (Max.) UF500-100 4" 108 - 116 177 (Max.) UF500-150 6" 158 - 169 UF500-200 8" 196 - 223 UF500-250 10" 250 - 277 25-0790 © Icon Process Controls Ltd.
  • Page 13: Installation Positions

    L > 10D Source : Reducer L > 5D L > 10D Flow Flow Source : Diffuser L > 5D L > 30D Flow Flow Source : Valve L > 10D L > 30D Flow Flow 25-0790 © Icon Process Controls Ltd.
  • Page 14: Maintenance

    Relocate sensor, add support No Flow Detected Empty pipe / Air / Incorrect install Confirm pipe full, check direction Flow = 0 at startup SQ < 50 or Set Zero failed Improve coupling, repeat Set Zero 25-0790 © Icon Process Controls Ltd.
  • Page 15: Warranty, Returns And Limitations

    Icon Process Controls Ltd for a period of one year from the date of sale of such products. Icon Process Controls Ltd obligation under this warranty is solely...