Page 1
Revision 2.0.0 Acoustic Control Systems – ACS Group Saarbrücken, Germany 2022 This instruction manual contains essential information on how to use this ACS product safely and effectively. Before using this product, thoroughly review this instruction manual. Use the product as instructed.
Page 7
Housing for the control unit. If the control Unit is damaged or malfunctioning, please contact your ACS support. Do not try to control the A1040 Mira 3D with other mobile devices that are not specified and provided by ACS, even if the connectivity and capability of executing the control software seems given.
Important basic information Important basic information This A1040 Mira 3D is CE-certified according to 2014/30/EU (EMC). Liquid ingress protection and solid particle protection of the casing is in conformity with level IP53 (standard EN 60529), so the equipment is splash proof and protected against dust.
Scope of Delivery Scope of Delivery Scope of Delivery The following two sections describe the scope of delivery and assembly instructions for A1040 MIRA 3D and A1040 MIRA 3D PRO. A1040 MIRA 3D A1040 MIRA 3D Scope of delivery Part...
Page 13
Operating Manual inside case Hard case with custom foam inlay A1040 MIRA 3D PRO Figure 1: Scope of delivery A1040 MIRA 3D PRO Scope of delivery Part Amount Location Electronic unit A1040 MIRA 3D with 32 x A-DPC®...
Page 14
The A1040 MIRA 3D PRO tomograph leaves the factory fully assembled. Follow up the given procedure to prepare a master to work as a stand-alone tomograph. · Put the A1040 MIRA 3D PRO on the flat stable surface. The transducers lay on the surface. ·...
Page 20
Keys and system LEDs Keys and system LEDs The A1040 Mira 3D Has two key -Types Type One is the key for the operational preparation of the system and to run status checks. This KEy is found on the Main Keypad Figure 3: Keypad position The Keypad has three basic functionalities.
Page 23
Rechagreable Battery of the Main Device The rechargeable batteries of the Main Device shall not to be replaced by the operator, only by full trained personnel that has been authorized by ACS. For safety reasons, do not attempt to remove or replace the battery.
Page 26
Rechargeable Battery of the Control Unit Rechargeable Battery of the Control Unit A1040 Mira 3D is normally equipped with a tablet device as Control Unit that is seated in a specially developed housing. The Control Units provided by ACS are selected to guarantee a save and reliable operation of the all over system.
Page 27
Rechargeable Battery of the Control Unit Rechargeable Battery of the Control Unit Do not try to control the A1040 Mira 3D with other mobile devices that are not specified and provided by ACS, even if the connectivity and capability of executing the control software is given.
Page 30
Never use the device on Surfaces with high dust, sand or other particle load. ACS is not liable for any damage or loss (whether contractually or unauthorized, including negligence) caused by not following this warning.
Page 31
Never use the device on Surfaces with high dust, sand or other particle load. ACS is not liable for any damage or loss (whether contractually or unauthorized, including negligence) caused by not following this warning.
For this measuring device, calibration is imperative. It should be included in the measuring equipment monitoring of the enterprise. Always keep the A1040 Mira 3D clean and maintained state. Avoid unnecessary force when using the device and always dry and clean the device before storage.
Technical Specification Technical Specification Technical Specification The technical specification is given in the A1040 Mira 3d warranty certificate, chapter 3 "Main Technical Specifications", that was given to you as part of the delivery set. The technical specification can be altered by ACS anytime by means of optimizing the device or for safety reasons.
Page 41
Remove the Control Unit from the Main Unit. Switching on the A1040 Mira 3D system Before switching on the A1040 Mira 3D system, check the status of the batteries according the chapters "Rechagreable Battery of the Main Device" and "Rechargeable Battery of the Control Unit"...
Page 61
C-Det C-Det Figure 8: Crack depth estimation with A1040 MIRA 3D Measurements with A1040 MIRA 3D Please, follow these steps: · Keep attention to the device being placed so that the crack is between the eighth and ninth row of transducers, Figure 9 ·...
The operator can use the farm view in both linear or matrix array mode. Transducer assignment The A1040 MIRA 3D addresses the transducer with a column and a column number separated by a dot, for example, column.row. The traducer with a minor address 1.1 has its position at the upper left corner of the instrument.
A-Scan 21.2 Farm View: A1048 MIRA 3D PRO The farm view of A1040 MIRA 3D PRO works equally to the farm view of A1040 MIRA 3D. The A1040 MIRA 3D PRO addresses the transducers of the master and the slave throughout (see Figure 14 ).
Page 73
Instrument configuration section. The operator can specify one of the following configurations: A1040 MIRA 3D or A1040 MIRA 3D PRO . The configuration A1040 MIRA 3D PRO is currently not available. Sub-panel informs the operator about current settings for the master Sub-panel informs the operator about current settings for the slave Indicator shows whether the master/slave is currently connected.
23.1 Instrument configuration After the app installation or re-installation, the operator must configure the instrument. The configuration is also necessary if the operator plans to change the configuration type between A1040 MIRA 3D and A1040 MIRA 3D PRO. 23.1.1 A1040 MIRA 3D To configure A1040 MIRA 3D follow the following steps.
Page 78
Setup Setup Before starting the configuration of the A1040 MIRA 3D PRO refer to the configuration details of A1040 MIRA 3D given on the page A1040 MIRA 3D PRO configuration Switch on the tomograph, establish the connection to the hotspot and start the app.
Setup Setup Figure 25: A1040 MIRA 3D PRO connection 23.2 Automatic transducer testing The delivery kit includes a test block. The block is a plexiglass plate with a slightly rough surface. It is used to perform the functionality test of each array transducer.
Page 84
· Charging (Usb) - charging over USB interface Instrument section. This section shows the charge level state of the master and the slave. In the case of A1040 MIRA 3D, section shows the same information as section (only the master is present). In the case of A1040 MIRA 3D PRO, this section lists the charge levels of both the master and the salve.
For carrying out the updates you need a Windows PC If required, the operator can do the WiFi configuration of the A1040 MIRA 3D or the A1040 MIRA 3D PRO. There are three cases when the WiFi configuration is needed: ·...
Connectivity Connectivity · Disconnect the A1040 MIRA 3D from the PC · Power off the instrument 25.4 WiFi module firmware update Update the firmware only when required by the manufacturer. · Download, unpack and install (if the program has not yet been installed) the...
A1040 MIRA 3D PRO utilize the classical WiFi interface. The Hotspot technology will not be further supported for the A1040 MIRA 3D and the A1040 MIRA 3D PRO. To ensure error-free functioning of the instrument, ACS requests to migrate all the A1040 MIRA 3D and the A1040 MIRA 3D PRO dispatched from ACS head office in Germany before 1.
Press and hold the button power up/down for at least 3 seconds. LED "ON" starts blinking. During this time, the tomograph is booting the operating system, please wait. If you are using A1040 MIRA 3D PRO power on both instruments ·...
. The UPDATE indicator is on now. It confirms the device's readiness for the update. · Windows PC detects the A1040 MIRA 3D storage device. Please, open it in the file explorer. · Copy the new firmware file to the storage. The update indicator is blinking while the firmware file is being copied.
Page 106
Licensing Licensing Combinations The operator can link or unlink electronic units of different A1040 MIRA 3D and A1040 MIRA 3D PRO tomographs to achieve different measurement goals. The A1040 MIRA 3D can operate in the following combinations with any A1040...
File format File format File format The file format of A1040 MIRA 3D and A1040 MIRA 3D PRO is open for any user. This chapter describes user parameters, reading procedures and saving principles. 28.1 File list The application saves the data relevant to the measurement in several files placed in the folder named by the user during the save procedure.
Page 110
MIRA 3D / A1040 MIRA 3D PRO reshape(bscan,param.frame_width,param.frame_height); read b-scan else %array mode = matrix, instrument configuration = A1040 MIRA 3D / A1040 MIRA 3D PRO n = filesize/ (param.frame_width*param.frame_height)/4; %read b-scan reshape(bscan,param.frame_width,param.frame_height,n); Read frame_width and frame_height parameters values from param.txt File vel_data.bin...
File format File format elseif(filesize == 4096 * 120 * 2) %array mode = linear/martix, instrument configuration = A1040 MIRA 3D a1040_mira3d_pro_bscan_liner = reshape(bscan,4096,120)'; %read and reshape 120 vectors 28.2 Modes B-Scan mode For the data measured in b-scan mode the saved data folder contains the files...
Page 115
A1040 MIRA 3D PRO works. Read data saved in raw_data.bin file using code above. The array a1040_mira3d_pro_bscan_linear has 2016 vectors. For the indexing scheme of the vectors refer to the next table:...
Page 124
Method 2: Automation velocity calibration The A1040 MIRA 3D and A1040 MIRA 3D PRO may estimate the sample velocity using ultrasonic wave propagating along the surface, from one DPC sensor to another.