Control low-frequency ultrasonic tomograph systems (44 pages)
Summary of Contents for ACS A1040 MIRA 3D
Page 1
Revision 1.1.0 Acoustic Control Systems – ACS Group Saarbrücken, Germany 2021 This instruction manual contains essential information on how to use this ACS product safely and effectively. Before using this product, thoroughly review this instruction manual. Use the product as instructed.
Housing for the control unit. If the control Unit is damaged or malfunctioning, please contact your ACS support. Do not try to control the A1040 Mira 3D with other mobile devices that are not specified and provided by ACS, even if the connectivity and capability of executing the control software seems given.
Page 7
Important basic information Important basic information This A1040 Mira 3D is CE-certified according to 2014/30/EU (EMC). Liquid ingress protection and solid particle protection of the casing is in conformity with level IP53 (standard EN 60529), so the equipment is splash proof and protected against dust.
Scope of Delivery Scope of Delivery Scope of Delivery The following two sections describe the scope of delivery and assembly instructions for A1040 MIRA 3D and A1040 MIRA 3D PRO. A1040 MIRA 3D A1040 MIRA 3D Scope of delivery Part...
Operating Manual inside case Hard case with custom foam inlay A1040 MIRA 3D PRO Figure 1: Scope of delivery A1040 MIRA 3D PRO Scope of delivery Part Amount Location Electronic unit A1040 MIRA 3D with 32 x A-DPC®...
Page 13
The A1040 MIRA 3D PRO tomograph leaves the factory fully assembled. Follow up the given procedure to prepare a master to work as a stand-alone tomograph. · Put the A1040 MIRA 3D PRO on the flat stable surface. The transducers lay on the surface. ·...
Keys and system LEDs Keys and system LEDs The A1040 Mira 3D Has two key -Types Type One is the key for the operational preparation of the system and to run status checks. This KEy is found on the Main Keypad Figure 3: Keypad position The Keypad has three basic functionalities.
Rechagreable Battery of the Main Device The rechargeable batteries of the Main Device shall not to be replaced by the operator, only by full trained personnel that has been authorized by ACS. For safety reasons, do not attempt to remove or replace the battery.
Rechargeable Battery of the Control Unit Rechargeable Battery of the Control Unit A1040 Mira 3D is normally equipped with a tablet device as Control Unit that is seated in a specially developed housing. The Control Units provided by ACS are selected to guarantee a save and reliable operation of the all over system.
Page 26
Rechargeable Battery of the Control Unit Rechargeable Battery of the Control Unit Do not try to control the A1040 Mira 3D with other mobile devices that are not specified and provided by ACS, even if the connectivity and capability of executing the control software is given.
For this measuring device, calibration is imperative. It should be included in the measuring equipment monitoring of the enterprise. Always keep the A1040 Mira 3D clean and maintained state. Avoid unnecessary force when using the device and always dry and clean the device before storage.
Technical Specification Technical Specification Technical Specification The technical specification is given in the A1040 Mira 3d warranty certificate, chapter 3 "Main Technical Specifications", that was given to you as part of the delivery set. The technical specification can be altered by ACS anytime by means of optimizing the device or for safety reasons.
Page 40
Remove the Control Unit from the Main Unit. Switching on the A1040 Mira 3D system Before switching on the A1040 Mira 3D system, check the status of the batteries according the chapters "Rechagreable Battery of the Main Device" and "Rechargeable Battery of the Control Unit"...
Page 61
The operator can use the farm view in both linear or matrix array mode. Transducer assignment The A1040 MIRA 3D addresses the transducer with a column and a column number separated by a dot, for example, column.row. The traducer with a minor address 1.1 has its position at the upper left corner of the instrument.
Page 64
A-Scan 21.2 Farm View: A1048 MIRA 3D PRO The farm view of A1040 MIRA 3D PRO works equally to the farm view of A1040 MIRA 3D. The A1040 MIRA 3D PRO addresses the transducers of the master and the slave throughout (see Figure 13 ).
Page 69
Instrument configuration section. The operator can specify one of the following configurations: A1040 MIRA 3D or A1040 MIRA 3D PRO . The configuration A1040 MIRA 3D PRO is currently not available. Sub-panel informs the operator about current settings for the master Sub-panel informs the operator about current settings for the slave Indicator shows whether the master/slave is currently connected.
Page 71
23.1 Instrument configuration After the app installation or re-installation, the operator must configure the instrument. The configuration is also necessary if the operator plans to change the configuration type between A1040 MIRA 3D and A1040 MIRA 3D PRO. 23.1.1 A1040 MIRA 3D To configure A1040 MIRA 3D follow up the following steps.
Page 74
Setup Setup Before starting the configuration of the A1040 MIRA 3D PRO refer to the configuration details of A1040 MIRA 3D given on the page A1040 MIRA 3D PRO configuration Switch on the tomograph, establish the connection to the hotspot and start the app.
Page 75
Setup Setup Figure 24: A1040 MIRA 3D PRO connection 23.2 Automatic transducer testing The delivery kit includes a test block. The block is a plexiglass plate with a slightly rough surface. It is used to perform the functionality test of each array transducer.
Page 81
· Charging (Usb) - charging over USB interface Instrument section. This section shows the charge level state of the master and the slave. In the case of A1040 MIRA 3D, section shows the same information as section (only the master is present). In the case of <% INSTRUMENT PRO%>, this section lists the charge levels of both the master...
Page 82
A1040 MIRA 3D PRO. 25.1 WiFi Hotspot The A1040 MIRA 3D and A1040 MIRA 3D PRO establish the connection with a Tablet-PC over WiFi Hotspot. The connection is successful if the parameters of the Tablet-PC WiFi Hotspot and the instrument are correct. This chapter describes the details on the parameters of the WiFi Hotspot connection.
Page 83
USB cable to the instrument TYPE C plug. The update indicator is on now. It confirms the device's readiness for the update. · The PC detects A1040 MIRA 3D storage device. Please, open it in the file explorer. · Copy file net.txt to the PC and open it with an text file editor ·...
Page 84
To read more about combinations, refer to the page If you need to update the A1040 MIRA 3D PRO WiFI Hotspot name, use the A1040 MIRA 3D procedure for the master and the slave separately. A1040 MIRA 3D...
Page 85
Getting started Getting started Getting started 26.1 App installation This chapter describes the steps for Installation or reinstallation of A1040 MIRA 3D android application Copying installation package · Copy installation package a1040_mira_3d-app-X.X.XX.apk to "acs-solutions apps" tablet internal storage folder. The name of apk file contains X.X.XX symbols.
Page 90
Follow up steps described at page including app installation and activation. Firmware update The A1040 MIRA 3D app informs the operator about the necessity of the firmware update. The update to be performed as follows: · Switch off the A1040 MIRA 3D ·...
Page 91
Wait until the update indicator stops blinking. · Disconnect the A1040 MIRA 3D from the PC. The instrument is updated now. · Start the instrument, connect the device to the app, and finally, check the firmware version in the setup mode.
Page 93
Licensing Licensing Combinations The operator can link or unlink electronic units of different A1040 MIRA 3D and A1040 MIRA 3D PRO tomographs to achieve different measurement goals. The A1040 MIRA 3D can operate in the following combinations with any A1040...
Page 95
File format File format File format The file format of A1040 MIRA 3D and A1040 MIRA 3D PRO is open for any user. This chapter describes user parameters, reading procedures and saving principles. 28.1 File list The application saves the data relevant to the measurement in several files placed in the folder named by the user during the save procedure.