November 2007 Warranty/Remedy Honeywell warrants goods of its manufacture as being free of defective materials and faulty workmanship. Contact your local sales office for warranty information. If warranted goods are returned to Honeywell during the period of coverage, Honeywell will repair or replace without charge those items it finds defective.
About This Document Abstract This manual provides description, specification, installation, mounting, maintenance and troubleshooting information for the Honeywell DL5000 Dissolved Oxygen (D.O.) Probe. Contacts World Wide Web http://hpsweb.honeywell.com Telephone Contact us by telephone at the numbers listed below. Organization Phone Number...
Contents INTRODUCTION ....................1 Overview ............................1 Probe Description ........................... 1 Operating Principal......................... 2 SPECIFICATIONS ....................3 INSTALLATION ....................4 Mounting and Wiring with DirectLine................... 4 Differences between the Quick Disconnect and Integral Cable options ........7 Mounting and Wiring with UDA2182 ................... 7 Probe System Check........................
Page 5
Table 3-2 Remote Mounting and Wiring Procedure for DL5000 Probes _________________________ 5 Table 3-1 Wiring DL5000 Probes with Integral Cable to UDA2182 ____________________________ 7 Table 3-1 Wiring DL5000 Probes with Quick Disconnect Cable to UDA2182 ____________________ 9 Table 5-1 Online Diagnostic Errors ____________________________________________________ 17...
1.2 Probe Description The Honeywell probe is housed in either PVC or Stainless steel casing. One end has a 1” NPT male thread for mounting. The sensor extends from the housing and is covered by a protective guard (of the same material as the casing) which allows sample entry while preventing physical damage.
1.3 Operating Principal When the probe is placed into the sample stream, oxygen diffuses through the membrane and is reduced at the cathode, and an equal amount of oxygen is generated at the anode. The diffusion continues until the oxygen partial pressure on both sides of the membrane is equal and a balance exists. The electrical circuit is designed such that the current necessary to maintain this equilibrium is converted to read out the dissolved oxygen concentration in the solution.
Probe membrane must be kept wetted both in the process and during storage. See Storage under Maintenance section. The silicone membrane surrounding the D.O. probe is permanent. If this membrane is removed by user for any reason, the warranty on the D.O. probe is voided 11/07 DL5000 Dissolved Oxygen (D.O.) Probe – User’s Manual...
Installation 3. Installation 3.1 Mounting and Wiring with DirectLine 3.1.1 DL424/425 The following section describes the installation of the Honeywell D.O. probe with the DL424/425 Integral Mounting. Table 3-1 Integral Mounting Procedure Step Procedure Connect the probe to the process source (using the appropriate mounting from those supplied for the DL5000).
When the DL424 or DL425 module is specified for remote mounting, a remote connector assembly (part number 51500768-005) is supplied loose. The remote cable connector is used to connect the DL5000 probe cable to the DL424/425 module. Table 3-2 gives the mounting procedure.
AC power wiring. After the analyzer and probe have been securely mounted, the probe should be connected to the analyzer. 3.3.1 Wiring DO Probe with Integral Cable Table 3-3 Wiring DL5000 Probes with Integral Cable to UDA2182 Step Procedure (Refer to Figure 3-3)
Installation CAUTION When installing the probe the wiring must be done in the order shown in Table 3-3, even if the analyzer is not powered. This is because the battery backup card is continuously supplying a voltage (bias) to the terminals. Figure 3-3 Connect/Disconnect Sequence for the DO Probe with an Integral Cable and the UDA2182 DL500 Dissolved Oxygen (D.O.) Probe –...
Installation 3.3.2 Wiring DO Probe with Quick Disconnect Option Table 3-4 Wiring DL5000 Probes with Quick Disconnect Cable to UDA2182 Step Procedure (Refer to Figure 3-4) Remove protective insulator tape from the green lead Loosen screws holding the lock bar and slide the lock bar away from the DO input card...
Installation Wire Wire Signal Signal Color Color Name Name Cable shield (Violet) to chassis ground screw Coax Cathode CONNECTING: Orange Reference Violet Shield wire first, then follow this order: Yellow Yellow Anode Orange Black Guard Coax Guard (Black) Green Green RTH Low RTH Low RTH Low...
Page 17
The recommended calibration schedule for ppb – µg/L, and ppm - mg/L applications is monthly. If individual plant maintenance schedules require it, the probe may be calibrated more frequently. 11/07 DL5000 Dissolved Oxygen (D.O.) Probe – User’s Manual...
Installation 3.5 Probe Mountings 3.5.1 Inline flow mounting In applications where the probe goes directly into the process pipeline, the in-line flow mounting is available. This mounting is diagramed in Figure 3-5. 30669860-017 O-ring 3/4” NPT (REF) Customer Installation Kit # 51452226-001 (SST) 51452226-002 (PVC) Figure 3-5 In-Line Mounting Kit...
Installation 3.5.3 Mounting Recommendation for Easy Calibration in Boiler Control Loop A "Trap" arrangement should be plumbed in the incoming flow to the DO probe so that if flow is shut off or an "Upstream" pipe opened, water will remain around the tip of the probe. Keeping it moist on the outlet side, an arrangement such as a "TEE"...
After these actions are taken, a cleaning schedule should be established. ATTENTION Never use a brush to clean a Honeywell D.O. probe as it can cause a tear in the probe membrane rendering it inoperable. This type of abuse is not covered under probe warranty.
Short term The Honeywell probe arrives from the factory with a red cap on the end of the probe. This cap contains a 3% solution of Isopropyl alcohol and water. The water keeps the probe wetted until it is put into the process.
, the probe may be left in the process and the Bias can be manually adjusted. Application related shift in probe bias voltage Remove the probe from the process and do an Air Calibration 11/07 DL5000 Dissolved Oxygen (D.O.) Probe – User’s Manual...
Page 24
Troubleshooting What you see What it is What to do These errors may occur during Air, Sample and Probe Bias calibration and abort the calibration process. FAIL These error messages are preceded by the Press Mode to return to online message “FAIL”...
Consult the DO Probe and Analyzer Tests section of the Manual 14.9 Appendix H for a comprehensive range of checks. In addition, the appendix section of the Honeywell UDA2182 Operator manual provides information on various subjects related to the D.O. probe, including the following topics: •...
Troubleshooting 5.4 Leak Detection in PPB- μg/L Applications Before performing and air leak detection, it is necessary to determine that both the probe and analyzer are working properly. First, check to see that the probe contains an O-ring. Per the probe directions, an O-ring must go into a probe that is used in ppb applications.