  • Page 1 FD-11614...
  • Page 2: Specifications

    SPECIFICATIONS FD-11614 ™ 16-Channel Temperature Input Device for FieldDAQ Conditions Specifications are typical and valid at -40 °C to 85 °C unless otherwise noted. Input Characteristics Number of channels 16 isolated thermocouple channels, 4 CJC Isolation Galvanic isolation between channels and to...
  • Page 3 Best 60 Hz rejection 89 dB Differential input impedance 5.34 MΩ Input noise High resolution 85 nV RMS Best 50 Hz rejection, 150 nV RMS best 60 Hz rejection High speed 1 μV RMS 2 | ni.com | FD-11614 Specifications...
  • Page 4: Temperature Measurement Accuracy

    Measurement sensitivity is a function of noise and represents the smallest change in temperature that a sensor can detect. The values assume the maximum of the full measurement range of the standard thermocouple sensor according to NIST Monograph 175. FD-11614 Specifications | © National Instruments | 3...
  • Page 5 0.36 0.72 1.58 0.42 0.92 400 °C 0.37 0.77 1.65 0.44 0.97 2.47 700 °C 0.43 0.96 1.97 0.54 2.83 900 °C 0.49 1.13 2.28 0.63 1.41 1000 °C 0.53 1.22 2.44 0.68 1.52 4 | ni.com | FD-11614 Specifications...
  • Page 6 Figure 1. Thermocouple Type K Errors (High Resolution/Best 50 Hz Rejection/Best 60 Hz Rejection) Typ 5 °C to 40 °C Max 5 °C to 40 °C Max –40 °C to 85 °C –200 1000 1200 1400 1600 Measured Temperature ( °C ) FD-11614 Specifications | © National Instruments | 5...
  • Page 7 1.29 0.35 0.76 2.07 300 °C 0.35 0.71 1.48 0.41 2.26 400 °C 0.37 0.78 1.58 0.45 0.98 2.38 700 °C 0.39 0.87 1.68 0.49 1.09 2.43 900 °C 0.43 1.89 0.56 1.24 2.65 6 | ni.com | FD-11614 Specifications...
  • Page 8 Figure 3. Thermocouple Type J Errors (High Resolution/Best 50 Hz Rejection/Best 60 Hz Rejection) Typ 5 °C to 40 °C Max 5 °C to 40 °C Max –40 °C to 85 °C –200 1000 1200 1400 1600 Measured Temperature ( °C ) FD-11614 Specifications | © National Instruments | 7...
  • Page 9 0.69 1.46 0.39 0.88 2.24 300 °C 0.33 1.41 0.39 0.89 2.08 400 °C 0.34 0.73 1.44 0.41 0.92 700 °C 0.38 0.88 1.66 0.48 2.32 900 °C 0.43 1.02 1.88 0.55 1.26 2.57 8 | ni.com | FD-11614 Specifications...
  • Page 10 Figure 5. Thermocouple Type N Errors (High Resolution/Best 50 Hz Rejection/Best 60 Hz Rejection) Typ 5 °C to 40 °C Max 5 °C to 40 °C Max –40 °C to 85 °C –200 1000 1200 1400 1600 Measured Temperature ( °C ) FD-11614 Specifications | © National Instruments | 9...
  • Page 11 0.33 0.72 400 °C 0.28 1.16 0.34 0.75 Error drift, thermocouple Type T (-100 °C to 400 °C) 5 °C to 40 °C 0.09 °C/10 °C -40 °C to 85 °C 0.19 °C/10 °C 10 | ni.com | FD-11614 Specifications...
  • Page 12 Figure 8. Thermocouple Type T Errors (High Speed) Typ 5 °C to 40 °C Max 5 °C to 40 °C Max –40 °C to 85 °C –200 1000 1200 Measured Temperature ( °C ) FD-11614 Specifications | © National Instruments | 11...
  • Page 13 Figure 9. Thermocouple Type E Errors (High Resolution/Best 50 Hz Rejection/Best 60 Hz Rejection) Typ 5 °C to 40 °C Max 5 °C to 40 °C Max –40 °C to 85 °C –200 1000 1200 1400 1600 Measured Temperature ( °C ) 12 | ni.com | FD-11614 Specifications...
  • Page 14 1700 °C 0.46 1.27 2.34 0.64 1.63 Error drift, thermocouple Type B (500 °C to 1800 °C) 5 °C to 40 °C 0.12 °C/10 °C -40 °C to 85 °C 0.13 °C/10 °C FD-11614 Specifications | © National Instruments | 13...
  • Page 15 Figure 12. Thermocouple Type B Errors (High Speed) Typ 5 °C to 40 °C Max 5 °C to 40 °C Max –40 °C to 85 °C 1000 1200 1400 1600 1800 2000 2200 Measured Temperature (°C) 14 | ni.com | FD-11614 Specifications...
  • Page 16 0.63 1.63 2.74 0.82 1.99 3.27 Error drift, thermocouple Type R/S (100 °C to 1400 °C) 5 °C to 40 °C 0.11 °C/10 °C -40 °C to 85 °C 0.17 °C/10 °C FD-11614 Specifications | © National Instruments | 15...
  • Page 17 Typ 5 °C to 40 °C Max 5 °C to 40 °C Max –40 °C to 85 °C –200 1000 1200 1400 1600 1800 2000 2200 Measured Temperature (°C) Time-Based Triggers Type Start Trigger 16 | ni.com | FD-11614 Specifications...
  • Page 18: Timing And Synchronization

    I/O synchronization is system-dependent. Assumes a system containing one hop. For information about achieving high accuracy synchronization, visit ni.com/info and enter Info Code fdsync This allows for line topologies or network redundancy. FD-11614 Specifications | © National Instruments | 17...
  • Page 19: Power Requirements

    Maximum cabling distance 100 m/segment Maximum hops per line Power Requirements The protection provided by the FD-11614 can be impaired if it is used in a Notice manner not described in the FD-11614 User Guide. Voltage input range 9 V DC to 30 V DC...
  • Page 20: Environmental Characteristics

    0.6 N · m (5.31 lb · in.) Ethernet) Calibration Calibration interval 1 year Environmental Characteristics Refer to the FD-11614 User Guide for more information about meeting these specifications. Temperature and Humidity Temperature Operating -40 °C to 85 °C Storage -40 °C to 100 °C...
  • Page 21: Environmental Standards

    IEC 60068-2-30 Damp heat cyclic (12 + 12h cycle) • IEC 60068-2-64 Random operating vibration To verify marine approval certification for a product, refer to the product Note label or visit and search for the certificate. ni.com/certification 20 | ni.com | FD-11614 Specifications...
  • Page 22: Safety Voltages

    Do not connect the FD-11614 to signals or use for measurements within Warning Measurement Categories II, III, or IV, or for measurements on MAINs circuits or on circuits derived from Overvoltage Category II, III, or IV which may have transient overvoltages above what the product can withstand.
  • Page 23: Electromagnetic Compatibility Standards

    Ethernet cables. CE Compliance This product meets the essential requirements of applicable European Directives, as follows: • 2014/35/EU; Low-Voltage Directive (safety) • 2014/30/EU; Electromagnetic Compatibility Directive (EMC) • 2011/65/EU; Restriction of Hazardous Substances (RoHS) 22 | ni.com | FD-11614 Specifications...
  • Page 24: Environmental Management

    NI products in your region, visit ni.com/environment/weee. 电子信息产品污染控制管理办法(中国 RoHS) National Instruments 符合中国电子信息产品中限制使用某些有害物 中国客户 质指令(RoHS)。关于 National Instruments 中国 RoHS 合规性信息,请登录 。(For information about China RoHS ni.com/environment/rohs_china compliance, go to ni.com/environment/rohs_china FD-11614 Specifications | © National Instruments | 23...
  • Page 25 For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your patents.txt ni.com/patents software, the file on your media, or the National Instruments Patent Notice at . You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer ni.com/legal/export-compliance...