Important. Throughout this manual the following words are given a specific meaning and the reader should be aware of their significance. Attention; Used where there is a risk of damaging the DAPcheck Plus incorrect handling or operation error. Note; Used where information is provided to aid better understanding of or its use.
In order to eliminate such insulation failures, the ionization chamber may be dried at an increased temperature (max. 50°C). Disposal The DAPcheck Plus must be disposed of in accordance with the applicable national regulations or returned to the manufacturer. A WEEE manual is available from Radcal.
System description Introduction The DAPcheck Plus is designed to function as an external sensor for use with a Accu-Gold Digitizer. Accu-Gold software operates the DAPcheck Plus though a computer interface. A support stand is provided to elevate the DAPcheck Plus above structures that might add scattered radiation to a primary beam measurement.
Preparations for operation Preparation The DAPcheck Plus should be carefully removed from its packing case and inspected. If any damage to the components is found, the supplier or manufacturer should be contacted. Damaged components must not be used. Power-on test Note.
Setup Installing the software - The DAPcheck Plus is compatible with Radcal’s Accu-Gold software AG2 as well as Accu- Gold Excel. Both are included on the USB flash-drive included with your unit. Or you can download the latest version of the software from the website at: www.radcal.com/support.
Operation DAP and DAP Rate measurements with the DAPCheckPlus system Attention. The DAPcheck Plus should only be used by personnel who are authorized to use sources of diagnostic medical X-radiation and are acquainted with these instructions. Using AG2 or AGT — The DAPcheck Plus can be used with the •...
Page 11
(I6 & I7 cells) to 5.00E+02 and 2.50E+02 respectively. Note that a measurement can be stopped at any time by pressing the Escape key. For questions or troubleshooting, contact Customer Support at (626) 357-7921 x 123 or email cust_sup@Radcal.com...
Technical specifications Notes: All technical data is valid for the ambient conditions as defined in IEC 60580. Only values given with tolerance ranges or limits are guaranteed. All other values are for information only. Subject to change without prior notice. Rate Specification: Range: 100nGy-m...
Page 13
Connection to Accu-Gold Digitizer One meter disconnect cable with 2 connectors (DAPCKP) or one meter disconnect cable with 1 connector (DAPCKP+) Protection Class (acc. IEC 60529) IP 41 Weight 2.3 kg; 5.1 lbs. Dimension (L x W x H) 350 mm x 410 mm x 35 mm;...
Page 14
Support Elevation Stand Foam Elevation Support Stand fits inside the interior of the Carry Case lid...
Radcal if the repairs are warranty-applicable. This warranty excludes batteries. Radcal shall not be held liable for damages or delays caused by defects beyond making repairs or furnishing replacement parts, nor shall Radcal be liable for any defective material replaced without Radcal’s consent during the period of this warranty.
Page 16
Declaration of Conformity According to ISO/IEC 17050 and EN 45014 The Radcal declares, under our sole responsibility, that the model DAPcheck Plus Measurement System conforms to the following product specifications. EMC: EN50081-2/EN55011, Class A EN61000-4-2, EN61000-4-3, EN61000-4-8 IEC60601-1 IEC60601-1-2 QA Manager...