Page 1
* This manual excludes information on how to operate and maintain the Induction Module Setup. Please see the “IM-CRDS Setup User’s Manual.” * This manual excludes information on how to operate and maintain the Picarro Autosampler. Please see the “Picarro Autosampler | Installation and Operation Manual.”...
Page 2
However, no guarantee is given or implied that this document is error free or that it is accurate with regard to any specification. Picarro, Inc. has prepared this manual for use by its customers as a guide for the proper installation, operation and/or maintenance of the Picarro analyzer.
See Conventions for definitions of graphic icons. For explanations of acronyms, see Acronyms. Picarro water analyzers (L2140-i, L2130-i or L2120-i) can work in many different configurations, allowing a diverse application of the CRDS (Cavity Ring-Down Spectroscopy) technology. This manual includes information on how to operate, analyze...
Page 6
Requires A0211 high precision vaporizer and A0101 standards delivery module. Alternates between analyzing ambient vapor from multiple points and a continuous stream of Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 7
Refer to the Table of Contents to find the location of any of the chapters described above. See Technical Support to learn how to streamline technical support from Picarro. For warranty information, see Limited Warranty. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual...
INTRODUCTION TO TECHNOLOGY Picarro analyzers use time-based, optical absorption spectroscopy of the target gases to determine concentration. They are based on wavelength-scanned cavity ring-down spectroscopy (WS-CRDS), a technology in which light re-circulates many times through the sample, creating a very long effective path length for the light to interact with the sample, thus, enabling excellent detection sensitivity in a compact and rugged instrument.
REMINDER is a helpful hint to procedures listed in the text. is a helpful hint to procedures listed in the text. Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual...
Standards Delivery Module Wavelength Monitor Purge Port. The Port on the Analyzer the dry gas WLM Purge Port connects to. This keeps the spectroscopy accurate. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Do not attempt repairs; except the particulate filter and the vacuum pump. Do not attempt repairs; instead, report all problems to Picarro Customer Service or your local instead, report all problems to Picarro Customer Service or your local instead, report all problems to Picarro Customer Service or your local distributor.
LASER SAFETY The outside of the Picarro analyzer is classified as a Class 1 Embedded Laser Product, nalyzer is classified as a Class 1 Embedded Laser Product, nalyzer is classified as a Class 1 Embedded Laser Product, while the inside of the Picarro a...
Dry air can be used but will require a software change to account for the change to account for the calibration shift from nitrogen. Contact Picarro for details on the software change. calibration shift from nitrogen. Contact Picarro for details on the software change.
4) Make sure the temperature of the High Precision Vaporizer stabilizes at 110° C by viewing the read out on the front of the vaporizer. 5) Make sure the Picarro Autosampler has been trained, methods and jobs defined, and samples loaded. See Picarro Autosampler: Installation, Training, Operation, Maintenance User’s Manual.
Page 15
If the analyzer is already running and these parameters need to be changed If the analyzer is already running and these parameters need to be changed it will require exiting and restarting the Picarro software. it will require exiting and restarting the Picarro software.
Page 16
6 hours vapor measurement: the total cycle time is 432 minutes and consumes 1 vial. total cycle time is 432 minutes and consumes 1 vial. The Picarro Autosampler has a utosampler has a tray for up to 105 vials. Therefore, o vials.
(see Cavity Ring-Down Spectrometer Controller). 19) When in need of a direct help from Picarro, see Technical Support. For warranty information, see Limited Warranty. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual...
4) Make sure the temperature of the High Precision Vaporizer has stabilized at 110° C by viewing the read out on the front of the vaporizer. 5) Make sure the Picarro Autosampler has been trained, methods and jobs defined, and samples loaded. See Picarro Autosampler: Installation, Training, Operation, Maintenance User’s Manual.
Page 19
7) To learn about the other coordinator modes supported by the Picarro water analyzer (but in different setups), see Coordinator Modes. 8) Once launched, the coordinator will automatically start collecting data, assuming the Autosampler job has been started.
Page 20
13) To configure data file saving details, including which data elements are written to data files, digital data output (via serial port or TCP/IP), remote data delivery (via email), and general GUI properties, click on the Setup Tool icon in the Picarro Utilities folder in the desktop. For more information, see Setup Tool.
Page 21
19) When in need of a direct help from Picarro, see Technical Support. For warranty information, see Limited Warranty. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Make sure the temperature of the High Precision vaporizer has stabilized to 110 Make sure the temperature of the High Precision vaporizer has stabilized to 110° C. 5) Make sure the Picarro Autosampler has been trained, samples loaded, and methods Make sure the Picarro Autosampler has been trained, samples loaded, and methods Make sure the Picarro Autosampler has been trained, samples loaded, and methods and jobs defined.
Page 23
Each injection cycle is less than 2 minutes. See the rest of the Chapter for more information on the Coordinator. To learn about the other coordinator modes supported by the Picarro water analyzer, see Coordinator Modes. Once launched, the coordinator will automatically start collecting data, assuming the Autosampler job has been started.
Page 24
Controller software can be used as a diagnostic tool in troubleshooting the analyzer (see Cavity Ring-Down Spectrometer Controller). 19) When in need of direct help from Picarro, see Technical Support. For warranty information, see Limited Warranty. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual...
Each injection cycle takes 9 minutes. The Manual Mode Setup can operate in two coordinator modes for the L2140-i. • O17 Manual Inject: Used for semi-automated measurement of liquid water samples with maximum precision on δ...
Page 26
12) To configure data file saving details, including which data elements are written to data files, digital data output (via serial port or TCP/IP), remote data delivery (via email), and general GUI properties, click on the Setup Tool icon in the Picarro Utilities folder in the desktop. For more information, see Setup Tool.
Page 27
Controller software can be used as a diagnostic tool in troubleshooting the analyzer (see Cavity Ring-Down Spectrometer Controller). 19) When in need of further help from Picarro, see Technical Support. For warranty information, please see Limited Warranty. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual...
1) Before continuing, review the important safety notes in Safety. 2) Make sure the hardware setup is complete and the system turned on in the correct sequence (INSTALLATION | L2140-i, L2130-i or L2120-i Analyzer and Peripherals User’s Manual) 3) Once turned on, the main GUI (Graphical User Interface) of the analyzer will open up automatically on the desktop screen.
Page 29
Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual...
Page 30
Pump 1 and Pump 2 are on the left and right side of the SDM when looking from the front. Click Apply after entering the parameters. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual...
Page 31
The Picarro analyzer is temperature stabilized and has extremely The Picarro analyzer is temperature stabilized and has extremely The Picarro analyzer is temperature stabilized and has extremely low drift so calibrations should be run typically every 6 low drift so calibrations should be run typically every 6-12 hours.
Page 32
The User Editable Parameters window will pop up. Enter the names of the liquid standards on the window, and the click ok to continue. The SDM Setup can operate in one coordinator mode. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 33
140 C temperature to 140 C° . 9) To learn about the other coordinator modes supported by the Picarro water t the other coordinator modes supported by the Picarro water t the other coordinator modes supported by the Picarro water analyzer (but in different setups), see analyzer (but in different setups), see Coordinator Modes.
Page 34
The coordinator will automatically create a *.csv file containing the measured values. The .csv file can be read using Notepad++, which is provided on the Picarro CPVU (Computer Power Vacuum Unit). 15) For information on how to post process SDM Data and how to set the output size and location of SDM Data files, see the Data section of this chapter.
Page 35
These steps are in addition to the user entered sequence. There are different situations when this occurs: Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 36
C:\SyringePumpData\ProcessedData by default. The processed data is saved in *.csv format and can be viewed by double clicking on the blue hyperlinks shown in Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 37
The Syringe Pump Data Processor Syringe Pump Data Processor Window Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual...
Page 38
To limit the size of the output file: Step 1: Open PicarroSyringePumpCoordinatorV13.ini file Step 2: Use Notepad++ text editor (right click on file) Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 39
24) For troubleshooting information for the SDM Setup, please see Troubleshooting | SDM. 25) When in need of further help from Picarro, see Technical Support. For warranty information, please see Limited Warranty. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual...
Switcher Software may or may not be included. When clicked, you will be led to a window which will allow you to switch between various measurement modes. For the L2140-i this is how you switch between the normal mode for measuring only δ δ δ δ...
Page 41
• For the L2140-i, you , you must stop the instrument using the Stop Instrument stop the instrument using the Stop Instrument icon prior to switching modes.
The Picarro Mode Switcher allows users to operate the analyzer in various modes. Switching between measurement modes is accomplished with a few easy steps: • Activate the user interface by double-clicking the Picarro Mode Switcher icon on the analyzer’s deskstop.
Page 43
For the L2140-i, you must stop the instrument (software and drivers) prior to i, you must stop the instrument (software and drivers) prior to i, you must stop the instrument (software and drivers) prior to δ...
The Picarro analyzer GUI has useful features. See below for more information. The Picarro analyzer GUI has useful features. See below for more information. Figure 2 - Layout of the Picarro analyzer GUI, G2000 series. Types of data Layout of the Picarro analyzer GUI, G2000 series. Types of data Layout of the Picarro analyzer GUI, G2000 series.
• Tools Menu This menu item has three entries: 1) User Calibration: Opens the user calibration window (default password is “picarro”). The password can be reset in the QuickGui.ini file in the instrument directory: “C:\Picarro\G2000\AppConfig \Config\QuickGUI\” under the section: 2) [Authorization] UserCalPassword = Picarro Show/Hide Valve 3) Sequencer GUI: Toggles the display of the external valve sequencer window.
Page 46
The filename and path of the active data log is displayed in this pane. The indicator is grayed-out if there is no active data log (i.e., if a new data log has not been started Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual...
Page 47
This window displays instrument status messages, in the following form: “MM/DD/YYYY hh:mm:ss generic message text.” These messages include all messages sent to the front panel display. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 48
To lock or unlock the time axes of each graph during zooming, select that menu item in the ‘View’ menu. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual...
Pressure High – check vacuum pump (error) There is unusually high pressure. The pump is either weak or dying or the vacuum hose has a leak. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Flux G2311-f analyzers. o Location: C:\UserData\DataLog_User_Sync\Year\Month\Day o Naming Convention: Example: CFHADS2007-20111222-000131- DataLog_User_Sync.dat Explanations are as same as for DataLog_User. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 51
This information is generally not useful to the user, but it can be useful for diagnostic purposes. For more information, please contact Picarro. o Location: C:\Picarro\G2000\Log\Archive\DataLog_Private\Year\Month\Day Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
CD) and “Troubleshooting” chapter of this manual. This command set may also be used across a TCP/IP interface through an ethernet connection. Please contact Picarro or visit our Community website for further details. Remote Data Access Using the RemoteAccess.ini file, the analyzer can be configured to automatically: •...
Page 53
“Network Connections” in the Control Panel. The values of the keys “Username,” “Password,” and “Number” are used to make the connection. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 54
Server5=time-c.timefreq.bldrdoc.gov Server6=time.nist.gov Server7=time-nw.nist.gov UpdateClock=1 [DIALUP] ConnectionName=Picarro Dialup Access Username=user Password=password Number=14085551212 [EMAIL] Server=smtp.servername.org Directory=c:/picarro/mailbox From=instrument@picarro.com To1=recipient1@site1.com To2=recipient2@site2.com Subject=CRDS data from Silverstone instrument UseAuthentication=0 Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
If you have any questions, please contact Picarro If you have any questions, please contact Picarro or refer to Picarro Community for or refer to Picarro Community for more information.
It is recommended to include standards in the middle and end of the sample set as controls. An example is shown below, vials with the CYAN stars being the standards: STANDARD Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 57
A sample data file along with the standards and instruction files can all be found in the ChemCorrect™ main folder C: all be found in the ChemCorrect™ main folder C:\Picarro\ChemCorrectExe ChemCorrectExe Below is a screenshot of the st...
Page 58
(contain a library of the standards’ isotope values) and number of injections to be ignored. You have an option to plot additional graphs for other parameters but those are just additional features. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 59
F. When the correct source and instruction files are shown, click the “Ok” button at the bottom of the window to cue ChemCorrect™ to begin analysis. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual...
Page 60
The color code legend is also available at the end of the summary tab. C. Red/yellow rows display relative contamination due to methane, methanol, or “other” hydrocarbons in the respective columns on the right. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 61
These files can be sent to Picarro for further analysis and spectral library files. These files can be sent to Picarro for further analysis and spectral library files.
Page 62
This corresponds to approximately 2 weeks of operation. Running the post processing ensures that any spectra associated with contamination are not erased. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
• Increasing the pressure by a small fraction of a psi should reduce the next pulse’s peak concentration. • Decreasing the pressure should increase the next pulse’s peak concentration. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Seven tabs of the Setup Tool window are explained in the next pages. If you have any indow are explained in the next pages. If you have any questions about the Setup Tool, please contact Picarro or refer to Picarro Community questions about the Setup Tool, please contact Picarro or refer to Picarro Community questions about the Setup Tool, please contact Picarro or refer to Picarro Community for further details: http://www.picarro.com/community/...
Page 65
. This command set may also be used across a TCP/IP interface a TCP/IP interface through an ethernet connection. thernet connection. Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual 40035 Rev. B...
Page 66
After making the appropriate edits, click “Apply” to put changes into effect edits, click “Apply” to put changes into effect edits, click “Apply” to put changes into effect, and then “Exit” to close the window. Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual...
Page 67
After making the appropriate edits, click “Apply” to After making the appropriate edits, click “Apply” to put changes into effect put changes into effect, and then “Exit” to close the window. Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual...
After making the appropriate edits, click “Apply” to put changes into effect After making the appropriate edits, click “Apply” to put changes into effect, and then “Exit” to close the window. Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual...
After making the appropriate edits, click “Apply” to put changes into effect After making the appropriate edits, click “Apply” to put changes into effect, and then “Exit” to close the window. Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual...
Page 70
After making the appropriate edits, click “Apply” to put changes into effect After making the appropriate edits, click “Apply” to put changes into effect, and then “Exit” to close the window. Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual...
This gas should be clean and dry to down. This gas should be clean and dry to Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual...
Page 72
If power to the analyzer is cut off for any reason the analyzer will cease operation. However, when the power is reapplied, the analyzer will restart automatically. the Picarro software tools will properly close out previous files and open new files for data collection so that previously collected data, instrument diagnostics and other parameters recorded up to the time of power outage are retained.
“External Valve Sequencer,” which is available from the Tools menu in the GUI. Default Configuration For all models of Picarro analyzers, the rotary valve control is disabled in the factory default setting. The solenoid valve control, however, is ready to use by default for all solenoid valve connectors.
Page 74
Please note the 9-pin connector cable is not supplied with the instrument. The setup will be similar to that in the pictures below: Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 75
For each individual measurement the analyzer makes, the valve codes and rotary valve positions Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 76
For example, if no solenoid valves are present, the codes will be recorded regardless of whether a valve is connected or not. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual...
METHOD 1: USING THE MAIN GRAPHICS USER INTERFACE (GUI) TO CALIBRATE THE ANALYZER Since the Picarro analyzer is extremely linear, it is only necessary to use three calibration standards to calibrate each gas or isotopic species (two points define the calibration line and a third intermediate point is used for verification).
Page 78
Tools menu and entering the slope and intercept for each species from the Tools menu and entering the slope and intercept for each species (see below). This is a password-protected function, with the default protected function, with the default password “picarro." password “picarro." After the calibration is entered, it After the calibration is entered, it will take effect immediately after clicking “ok.”...
Page 79
Picarro analyzer. The Picarro Data Recalibration (“Data Recal” The Picarro Data Recalibration (“Data Recal” icon) software can be found in the “Picarro ) software can be found in the “Picarro Utilities” folder in the desktop, and it can be used to recalibra Utilities”...
Page 80
12. You have now calibrated your analyzer. The next time you do any sample measurements, the measurements will be based on the new calibration setting. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Figure 3 - Overview generalized schematic of Picarro analyzer system hardware, software, and data file generation: The samples A, B, and C are introduced into the analyzer sequentially by a sampling module, if present. The sampling module could be a set of customer-supplied valves, a valve manifold, a discrete sampling module, or other device.
Page 82
Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual...
Page 83
For the Picarro Autosampler, only one dialog box will appear. Select the sample the Picarro Autosampler, only one dialog box will appear. Select the sample the Picarro Autosampler, only one dialog box will appear.
Page 84
Coordinator Window. d. UPPER PORTION OF THE WINDOW: Each row represents the analysis results from a single injection. The types of columns are pre-selected by Picarro to include the most useful values for isotopic water analysis and for diagnostic indications.
Page 85
The output file is rewritten to use the new sample descriptions, so that the most recently loaded descriptions are always used. most recently loaded descriptions are always used. Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual...
COORDINATOR MODES The CRDS analyzer needs to be equipped with the appropriate hardware to support a coordinator mode. Unless stated otherwise, the chapters below are referenced in the INSTALLATION | L2140-i, L2130-i, or L2120-i Analyzer and Peripherals User’s Manual. SUPPORTED...
Page 87
This requires the Induction Module. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 88
IMPORTANT NOTE: The IMPORTANT NOTE: The L2140-i can operate in four instrument modes i can operate in four instrument modes (iH2O N2, iH2O Air, iH2O N2 O (iH2O N2, iH2O Air, iH2O N2 O-17 and iH2O air O-17). By default, only By default, only the ‘High Precision’...
There are two in-line, sub-micron particulate filters before the measurement cavity: • • • • The first filter is user-replaceable and is found immediately following the inlet at the back of the analyzer. Replacement filters can be purchased from Picarro to be installed by the user.
Page 90
(from the back of the analyzer) to remove it. Remove Remove this foam. this foam. Loosen this retaining nut on the input bulkhead. Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual 40035 Rev. B...
Page 91
Loosen this first nut Hold the old filter steady with 11/16” steady with 11/16” Loosen the wrench first nut with 9/16” wrench Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual 40035 Rev. B...
Page 92
7. Slide the filter slightly towards the back of the a Slide the filter slightly towards the back of the analyzer and lift it out. nalyzer and lift it out. Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual...
Page 93
The metal edge of the filter cover should be under the foam. 4. With a 2mm hex driver, re , reattach the analyzer’s top with six screws. screws. Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual 40035 Rev. B...
Wait for the Change Septum button to change to Septum Changed (see below). This will happen when the This will happen when the current injection analysis is complete. Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual...
Page 95
The analyzer will restart the vaporizer purge cycle and then wait for the button. The analyzer will restart the vaporizer purge cycle and then wait for the next sample injection. 9. Done! Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual...
If, after attempting these procedures, the problem remains If, after attempting these procedures, the problem remains If, after attempting these procedures, the problem remains unresolved, please contact Picarro Technical Support Picarro Technical Support. More troubleshooting . More troubleshooting information is available on the Community section of our website.
Page 97
3. Select “Shut down” in the drop-down box. 4. Wait for the computer and analyzer to turn off completely and the shutdown to complete normally. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 98
1. Hold down the power switch on the front panel for a few seconds until the computer and the instrument power off. 2. After another few seconds, restart the analyzer by momentarily depressing the power button. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
The temporary work around is to delete this file "chemcorrect.clt" in the ChemCorrectExe root directory. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual...
If both are operating correctly, this means excess water entered the vaporizer previously. Please contact Picarro for instructions on how to dry out the vaporizer. If the pump is running and air flow is <250 sccm then the line is leaking before the vaporizer or is blocked.
Page 101
This is illustrated in the examples below: Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 102
Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
This failure mode is more pronounced in samples with a large amount of solids, i.e. seawater. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual 40035 Rev. B...
Page 104
In any case, values under about 12 seconds are not recommended. It may also be advisable to try reducing the sample volume. Picarro L2140-i, L2130-i or L2120-i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting User’s Manual...
Page 105
Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual...
CONTROLLER On your desktop, there is an icon label On your desktop, there is an icon labelled “Picarro Controller.” Clicking on this icon will ed “Picarro Controller.” Clicking on this icon will open up a useful diagnostics panel (see image below), allowing the user to see the...
When shipping the analyzer, use its . When shipping the analyzer, use its original packaging only. original packaging only. Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual 40035 Rev. B...
(12) months from shipment. specifications for twelve (12) months from shipment. This warranty is the only warranty made by Picarro with respect to its Products and no e only warranty made by Picarro with respect to its Products and no...
Page 109
The following warranty conditions shall apply to all Picarro, Inc. products unless amended by a written instrument signed by a duly authorized officer of Picarro: ADJUSTMENT – No electrical, mechanical or optical adjustments to the product(s) are permitted. PARTS AND LABOR - New or factory-built replacements for defective parts will be supplied for twelve (12) months from date of shipment of the product.
• Email us at support@picarro.com. We will get back to you right away. We highly recommend that you attach data and/or screen shots that you feel might help us diagnose your problem to your email.
Page 111
The “LogMeIn Free” window will pop up. corner. The “LogMeIn Free” window will pop up. 2. Click the “Turn on now” button. Click the “Turn on now” button. Picarro L2140-i, L2130-i or L2120-i i Analyzer and Peripherals Operation, Maintenance, and Troubleshooting Troubleshooting User’s Manual...
Page 112
@picarro.com email). 4. After your problem has been After your problem has been solved, you can turn off Picarro’s access to your solved, you can turn off Picarro’s access to your desktop by clicking on the “Turn off” button (see screenshot above).
CONNECTING OUR CUSTOMERS Want to show off your research to the world? Apply to publish your story on our newsletter by contacting us at info@picarro.com! Is there a topic you would like to see covered in the newsletter? Let us know at info@picarro.com! Curious about what our customers around the globe are working on? Picarro’s quarterly electronic newsletter highlights stories and research around the...
Need help?
Do you have a question about the L2140-i and is the answer not in the manual?
Questions and answers