Chattanooga Vectra Neo Service Manual
Chattanooga Vectra Neo Service Manual

Chattanooga Vectra Neo Service Manual

Clinical therapy system
Hide thumbs Also See for Vectra Neo:
Table of Contents

Advertisement

Quick Links

Vectra® Neo Clinical
Therapy System
Service Manual
Service Instructions For:
• Vectra Neo Clinical Therapy
System Part #6000
Optional Accessories:
• Stim 1_2 Module - PN 70000
• Ultrasound Module - PN 70002
• Stim 3_4 Module - PN 70003
• Stim 1_2 + EMG Module - PN 70004
• Laser Module - PN 70005
• Neo Cart - PN 70001

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Vectra Neo and is the answer not in the manual?

Questions and answers

Summary of Contents for Chattanooga Vectra Neo

  • Page 1 Vectra® Neo Clinical Therapy System Service Manual Service Instructions For: • Vectra Neo Clinical Therapy System Part #6000 Optional Accessories: • Stim 1_2 Module - PN 70000 • Ultrasound Module - PN 70002 • Stim 3_4 Module - PN 70003 •...
  • Page 3: Table Of Contents

    Vectra® Neo Clinical Therapy System – Service Manual TABLE OF CONTENTS INTRODUCTION ........3 UNIT STARTUP AND FAN TESTING .
  • Page 4 Vectra® Neo Clinical Therapy System – Service Manual APPENDIX........66 COMMON TERMS .
  • Page 5: Introduction

    INTRODUCTION Vectra® Neo Clinical Therapy System – Service Manual FOREWORD PRECAUTIONARY INSTRUCTIONS This manual is intended for users of Vectra® Neo Clinical The precautionary instructions found in this section and Therapy System. It contains general information on throughout this manual are indicated by speci c symbols. operation, precautionary practices, and maintenance.
  • Page 6: General Terminology

    INTRODUCTION Vectra® Neo Clinical Therapy System – Service Manual GENERAL TERMINOLOGY The following are de nitions for the terminology used throughout this manual. Study these terms to become familiar with them for ease of system operation, and control functionality of the Vectra® Neo Clinical Therapy System. SYSTEM SOFTWARE SYMBOLS Back Arrow Stim...
  • Page 7: Description Of Device Markings

    INTRODUCTION Vectra® Neo Clinical Therapy System – Service Manual DESCRIPTION OF DEVICE MARKINGS The markings on the unit are assurance of its conformity to the highest applicable standards of medical equipment safety and electromagnetic compatibility. One or more of the following markings may appear on the device: Refer to Instructional Manual Booklet .
  • Page 8: Device Description

    DEVICE DESCRIPTION Vectra® Neo Clinical Therapy System – Service Manual PRODUCT DESCRIPTION The Vectra® Neo Clinical Therapy System is a modular system used with or without an optional Cart, allowing for the inclusion of Channel 1/2 Electrotherapy module with or without sEMG, Channel 3/4 Electrotherapy module, Laser module and Ultrasound module.
  • Page 9: Operator Interface

    DEVICE DESCRIPTION Vectra® Neo Clinical Therapy System – Service Manual OPERATOR INTERFACE The Vectra® Neo Clinical Therapy System Operator Front Controls Interface contains all the functions and controls necessary for operator access to all operator utilities, modalities, and parameters for modi cation and system set up. 1.
  • Page 10: General Warnings And Precautions

    GENERAL WARNINGS AND PRECAUTIONS Vectra® Neo Clinical Therapy System – Service Manual CAUTION CAUTION • Read, understand, and practice the precautionary and operating instructions. • Failure to use and maintain the Vectra® Neo Clinical Therapy System, its modules, Know the limitations and hazards associated with using any electrical and its accessories in accordance with the instructions outlined in this manual stimulation, Laser device or ultrasound device.
  • Page 11: Warning

    GENERAL WARNINGS AND PRECAUTIONS Vectra® Neo Clinical Therapy System – Service Manual WARNING WARNING • U.S.A. Federal Law restricts these devices to sale by, or on the order of, a • The color of skin, age of lesion, depth of lesion, sensitivity of the patient, tissue physician or licensed practitioner.
  • Page 12: Danger

    GENERAL WARNINGS AND PRECAUTIONS Vectra® Neo Clinical Therapy System – Service Manual WARNING DANGER • Stimulus delivered by the TENS waveforms of this • The laser head must be cleaned with a disinfectant cleaner device, in certain configurations, will deliver a charge of 25 (i.e.
  • Page 13: Detail Device Description

    DETAIL DEVICE DESCRIPTION Vectra® Neo Clinical Therapy System – Service Manual COMPONENTS Throughout these instructions the terms “left” and Modules “right” referring to the machine sides are from the • Stimulation Channel 1/2 perspective of a user standing in front of the unit. •...
  • Page 14: Module Slots

    DETAIL DEVICE DESCRIPTION Vectra® Neo Clinical Therapy System – Service Manual MODULE SLOTS MODULE KIT CONTENTS 1. Laser Electrotherapy Module Channels 1/2 – PN 70000 • Stimulation module 2. Stimulation (1 & 2) / Stimulation (1&2) + sEMG • Clip lead wire 3.
  • Page 15: Ultrasound Applicator

    DETAIL DEVICE DESCRIPTION Vectra® Neo Clinical Therapy System – Service Manual ULTRASOUND APPLICATOR 1. Applicator Head The component of the applicator that makes contact with the patient during Ultrasound or Combination therapy. 2. Applicator The assembly that connects to the system and incorporates the Applicator.
  • Page 16: Patient Remote/Laser Interrupt Switch

    DETAIL DEVICE DESCRIPTION Vectra® Neo Clinical Therapy System – Service Manual PATIENT REMOTE/LASER INTERRUPT SWITCH The Vectra® Neo Patient Remote/Laser Interrupt Switch buttons are described below. By default, the remote is not assigned to any treatment. When assigned, the buttons function as follows: Increase Intensity (1) Decrease Intensity (2)
  • Page 17: Neo Leg To Cart Assembly/Adjustment

    SET UP INSTRUCTIONS Vectra® Neo Clinical Therapy System – Service Manual NEO LEG TO CART ASSEMBLY/ADJUSTMENT The Neo Cart is shipped without the legs attached. To 2. There are two Cart height adjustments. Standard install or adjust the leg assemblies onto the Neo Cart, shown on the left and lowered, shown on the right.
  • Page 18: Module Specific Information

    SET UP INSTRUCTIONS Vectra® Neo Clinical Therapy System – Service Manual MODULE INSTALLATION MODULE SPECIFIC INFORMATION All modules are installed from the left side (when facing Ultrasound Cable Insertion the screen) of the Neo head unit and are each installed Shown below is the Ultrasound Cable Insertion location.
  • Page 19 SETUP INSTRUCTIONS Vectra® Neo Clinical Therapy System – Service Manual PATIENT REMOTE/LASER INTERRUPT SWITCH INSTALLATION To operate the Patient Remote/Laser Interrupt Switch, 2. Press the Remote ON/OFF toggle icon to assign or plug the remote into the device on the Rear Access Panel unassign the remote to the selected treatment.
  • Page 20: Installing The Laser Interlock Door Interrupt Switch

    SET UP INSTRUCTIONS Vectra® Neo Clinical Therapy System – Service Manual INSTALLING THE LASER INTERLOCK DOOR INTERRUPT SWITCH Diagram for Therapy Room with One Door. The Laser Interlock is an optional safety device designed to interrupt Laser therapy if the door to the therapy room is opened.
  • Page 21: Therapy System Start Up

    SETUP INSTRUCTIONS Vectra® Neo Clinical Therapy System – Service Manual THERAPY SYSTEM START UP Complete the following steps for initial setup of the 3. Select desired function on the Home Screen (shown Vectra® Neo Clinical Therapy System: below). 1. Plug the Power cord into the back of device. Plug the other end of the cord into an electrical outlet.
  • Page 22: System Specifications

    SYSTEM SPECIFICATIONS Vectra® Neo Clinical Therapy System – Service Manual SYSTEM SPECIFICATIONS AND DIMENSIONS Width Depth Height Weight Module 11.12” (28.2448 cm) 6.34” (16.1036 cm) 1.43” (3.6322 cm) 1lb (0.453592 kg) Head @ 45 degree with Base (Tabletop) 15.89” (40.3606 cm) 15.89”...
  • Page 23: Ultrasound Specifications

    SYSTEM SPECIFICATIONS Vectra® Neo Clinical Therapy System – Service Manual ULTRASOUND SPECIFICATIONS ULTRASOUND SPATIAL PATTERN The following charts represent the distribution of the Frequency ....... . . 1 MHz, ± 5%; 3.3 Mhz, ±5% Duty Cycles .
  • Page 24: Laser Specifications

    SYSTEM SPECIFICATIONS Vectra® Neo Clinical Therapy System – Service Manual LASER SPECIFICATIONS Power Output Type ........Infrared Lamp (Laser) Laser Class.
  • Page 25: Laser Applicator Technical Specifications

    SYSTEM SPECIFICATIONS Vectra® Neo Clinical Therapy System – Service Manual LASER APPLICATOR TECHNICAL SPECIFICATIONS For all single diode and cluster laser and LED applicators, the expected increase in the measured quantities after manufacture added to the values measured at the time of manufacture is 20%.
  • Page 26 SYSTEM SPECIFICATIONS Vectra® Neo Clinical Therapy System – Service Manual LASER APPLICATOR TECHNICAL SPECIFICATIONS CONTINUED Output Power Nominal Ocular Divergence Divergence Diode Wavelength(s) Treatment Area Model # Description Power Density Hazard Distance Type (nm) (Spot Size) (cm (mW) (W/cm (NOHD-in meters) (rad) (rad) 9-Diode...
  • Page 27: Laser Protective Eyewear Specifications

    SYSTEM SPECIFICATIONS Vectra® Neo Clinical Therapy System – Service Manual LASER PROTECTIVE EYEWEAR SPECIFICATIONS The graph below illustrates optical density in relation to wavelength. Each unit is shipped with laser protective eyewear that is L3 rated and approved as well as EN207 compliant.
  • Page 28: Laser Labels

    SYSTEM SPECIFICATIONS Vectra® Neo Clinical Therapy System – Service Manual LASER LABELS This serial decal label is a xed at the rear of the system. Complies with 21CFR 1040.10 21CFR 1040.11 IEC 60601-1 IEC 60601-1-2 IEC 60601-2-22 IEC 60601-2-57 IEC 62471 IEC 60825-1:2007 DJO, LLC 1430 Decision Street...
  • Page 29: Waveforms

    WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual WAVEFORMS TENS- Asymmetrical Biphasic IFC (Interferential) Traditional (4 Pole) The Asymmetrical Biphasic waveform has a short pulse duration. It is capable Interferential Current is a medium frequency waveform. Current is distributed of strong stimulation of the nerve bers in the skin as well as of muscle tissue.
  • Page 30 WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual VMS™ VMS™ FR VMS is a symmetrical biphasic waveform with a 100 µsec interphase interval. The VMS-FR version of the VMS waveform is a physiologically based channel Because the pulse is relatively short, the waveform has a low skin load, interaction in which one channel stimulates the agonist and the other the making it suitable for applications requiring high intensities, such as in muscle antagonist of the muscle group that is being exercised.
  • Page 31 WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual IFC Premodulated (Traditional 2 Pole) DC (Direct Current) Premodulated Current is a medium frequency waveform. Current comes out of DC Current is a direct current owing in one direction only. The current can be one channel (two electrodes).
  • Page 32 WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual High Voltage Pulsed Current (HVPC) The High Voltage Pulsed Current (HVPC) has a very brief pulse duration characterized by two distinct peaks delivered at high voltage. The waveform is monophasic (current ows in one direction only). Output Mode .
  • Page 33: Therapy System Testing

    FACTORY standards and ensure proper readings. 4. The troubleshooting and repair of the Vectra Neo Clinical Therapy System Modules, and Accessories E. Full Functional Tests should be performed only by authorized technicians Perform the tests found in this section to verify Full trained and certified by DJO, LLC.
  • Page 34: Visual Inspection

    WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual VISUAL INSPECTION Visually inspect the Vectra® Neo Clinical Therapy System. A visual inspection can, to an experienced technician, indicate possible abuse of the unit and internal problems. LEAKAGE TESTS Conduct all necessary leakage tests as required per “Chapter 7 Electrical Equipment”...
  • Page 35: Stimulator Test System Setup

    WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual STIMULATOR TEST SYSTEM SETUP The following tests for Stimulator Outputs will be performed on Channels 1 and 2. The performance of these same tests will apply to the Channel 3 and 4 Electrotherapy Module for four channel therapy systems.
  • Page 36: Vms™ Mode Test

    WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual VMS™ MODE TEST A. VMS™ Mode Test Procedures 1. Set Scope; Time- 100 µS, Channel- 50 V, and Trigger- 2. Press Electrotherapy Button. 3. Press VMS Button and then press Customize Button. 4.
  • Page 37: Interferential Mode Test

    WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual INTERFERENTIAL MODE TEST It is assumed that the unit is ready for tests as described in STIMULATOR TEST SYSTEM SETUP Section. If not, set up Unit per STIMULATOR TEST SYSTEM SETUP Section parts A and B prior to performing tests.
  • Page 38: Premodulated Mode Test

    WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual PREMODULATED MODE TEST Set up System per STIMULATOR TEST SYSTEM SETUP Section prior to performing test. A. Premodulated Mode Test Procedures 1. Set Scope; Time- 2.50 mS, Channel- 20 V, and Trigger- 2.
  • Page 39: Russian Mode Test

    WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual RUSSIAN MODE TEST Set up System per STIMULATOR TEST SYSTEM SETUP Section prior to performing test. A. Russian Mode Test Procedures 1. Set Scope; Time- 5 mS, Channel- 50 V, and Trigger- DC NOTE: A test of the Optional Patient Interrrupt Switch is provided within the Russian Mode Test.
  • Page 40: Microcurrent Mode Test

    WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual MICROCURRENT MODE TEST Set up System per STIMULATOR TEST SYSTEM SETUP Section prior to performing test. Place ESTI-2 Load Switch in the 10 K Micro position only LOAD SWITCH for the Microcurrent Mode Tests. See Figure 1. TO 10K MICRO A.
  • Page 41: High Voltage Pulsed Current Hvpc Mode Test

    WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual HIGH VOLTAGE PULSED CURRENT HVPC MODE TEST PROPER NEGATIVE HIGH VOLT WAVEFORM Set up System per STIMULATOR TEST SYSTEM SETUP Section prior to performing tests. A. High Voltage Pulsed Current (HVPC) Mode Test Procedures 1.
  • Page 42: Ultrasound Tests

    WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual ULTRASOUND TESTS A. Equipment Required 1. Degassed Water 2. Ohmic Instruments UPM DT 10 or DT 100 Ultrasound Power Meter. 3. Dissolved Oxygen Test Kit. Used to test oxygen level of degassed water.
  • Page 43: Ultrasound Applicator Output Test

    WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual ULTRASOUND APPLICATOR OUTPUT TEST Perform this test using the Vectra® Neo Clinical Therapy System. A. Ultrasound Applicator Output Test Procedures 1. Set up Ohmic Instruments UPM DT 10 or DT 100 Ultrasound Power Meter per Operator’s Instructions and fill test reservoir with Degassed Water.
  • Page 44: Ultrasound Duty Cycle Test

    WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual ULTRASOUND DUTY CYCLE TEST This test is performed using the Vectra® Neo Clinical Therapy System. A. Ultrasound Duty Cycle Test Procedures 1. Set up Ohmic Instruments UPM DT 10 or DT 100 Ultrasound Power Meter per Operator’s Instructions and fill test reservoir with Degassed Water.
  • Page 45: Combo Operation Test

    WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual COMBO OPERATION TEST This test is performed using the 5 cm Applicator. Select Channel 2 and set up System per STIMULATOR TEST SYSTEM SETUP Section parts A and B prior to performing tests.
  • Page 46: Semg And Semg + Electrical Stimulation Tests

    WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual SEMG AND SEMG + ELECTRICAL STIMULATION TESTS Perform this test on all Channels with sEMG. However, ATTENUATOR SCHEMATIC only one channel at a time can be tested with the Test 10 M 1% Equipment described in this section.
  • Page 47 WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual SEMG AND SEMG + ELECTRICAL STIMULATION TESTS CONTINUED 3. If icons are present, connect known good sEMG Lead Wire to Channels 1 and 2. See Figure 4. NOTE: Only one Channel at a time can be tested for sEMG. 4.
  • Page 48 WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual SEMG AND SEMG + ELECTRICAL STIMULATION TESTS CONTINUED 7. Make certain the Audio Signal Generator is set up per sEMG TEST PROCEDURES steps 1, a) through 1, g). Connect the Audio Signal Generator Test Leads from the Generator SYNC Ports to the Attenuator (make certain test leads are connected red to red and black to black).
  • Page 49 WAVEFORMS Vectra® Neo Clinical Therapy System – Service Manual SEMG AND SEMG + ELECTRICAL STIMULATION TESTS CONTINUED D. sEMG + Stim Tests 1. To Check Stim Output, conduct the Electrical Stimulator Tests. 2. Set up Signal Generator and Attenuator as described in sEMG TEST PROCEDURES.
  • Page 50: Troubleshooting

    TROUBLESHOOTING Vectra® Neo Clinical Therapy System – Service Manual TROUBLESHOOTING CODES The Vectra® Neo Therapy System incorporates information, warning, and error messages to inform the user of problems or potential problems with the system, modality, or accessories. The numbering sequence is: 100-level messages are general use information messages.
  • Page 51 TROUBLESHOOTING Vectra® Neo Clinical Therapy System – Service Manual TROUBLESHOOTING CODES CON’T Code Type Probable Cause Possible Remedies Number Message 1. If a stim module is installed in the unit, ensure module is Attempting to select an electrotherapy securely inserted into the unit Information treatment but no Stim module is installed in the 2.
  • Page 52: 300 Level Errors

    TROUBLESHOOTING Vectra® Neo Clinical Therapy System – Service Manual TROUBLESHOOTING CODES CON’T Code Type Probable Cause Possible Remedies Number Message Warning Immediately stop all use of the system and contact the dealer or DJO for service. Warnings in these 200-level categories indicate an internal problem with the system that must be tested and corrected by DJO or a code not Trained Technician before any further operation or use of the system.
  • Page 53: Ultrasound Calibration

    TROUBLESHOOTING Vectra® Neo Clinical Therapy System – Service Manual ULTRASOUND CALIBRATION A. Tools and Equipment Required 1. Vectra® Neo Clinical Therapy System and the Intelect Vet Combination Therapy System and all Vectra Genisys, Intelect Legend XT and the Intelect Vet Ultrasound Applicators associated with the System being serviced.
  • Page 54: Touch Screen Calibration

    TROUBLESHOOTING Vectra® Neo Clinical Therapy System – Service Manual TOUCH SCREEN CALIBRATION A. Calibrating With the Unit Off 1. With the Vectra® Neo Clinical Therapy System off, press and hold the Stop and Start Buttons at the same time. See Figure 1. 2.
  • Page 55: Laser Pin Code

    TROUBLESHOOTING Vectra® Neo Clinical Therapy System – Service Manual LASER PIN CODE A. Reset Pin Code Back to 1111 1. Press the Utilities Button. 2. Press and hold the Stop and Power Buttons simultaneously, refer to Figure 1. Laser Pin will reset. 3.
  • Page 56: Accessories

    ACCESSORIES Vectra® Neo Clinical Therapy System – Service Manual REPLACEMENT ACCESSORIES The following provides users of the Vectra® Neo Therapy System the necessary information to order replacement accessories commonly used with the system. This list of replacement accessories is designed for use with the Vectra®...
  • Page 57: General Accessories

    ACCESSORIES Vectra® Neo Clinical Therapy System – Service Manual REPLACEMENT ACCESSORIES CON’T GENERAL ACCESSORIES Model Number Description 70000 Electrotherapy Module Channels 1/2 70001 Neo Cart 70002 Ultrasound Module 70003 Electrotherapy Module Channels 3/4 70004 Electrotherapy Module Channels 1/2 + sEMG 70005 Laser Module 70008...
  • Page 58: Laser Applicators And Eyewear Accessories

    ACCESSORIES Vectra® Neo Clinical Therapy System – Service Manual REPLACEMENT ACCESSORIES CON’T LASER APPLICATORS AND EYEWEAR ACCESSORIES Model Number Description 27840 Single 850nm Laser 100mW 27804 Single 850nm Laser 150mW 27841 Single 850nm Laser 200mW 27803 Single 850nm Laser 40mW 27805 Single 820nm Laser 300mW 27810...
  • Page 59: Parts

    FAN TRAY 13-6789 HUB BASE DESKTOP VERSION ASSEMBLY 13-7611 CLIP LEAD WIRE MANAGER NEO 13-7611 CLIP RAIL DESKTOP NEO 13-7645 LABEL, CHATTANOOGA LOGO ROUND 13-7564 POWER SUPPLY 12-6776 HUB CHASSIS ASSY 13-7583 COVER EMPTY MODULE BAY NEO 13-7686 POWER ENTRY IEC320 ASSY...
  • Page 60: Neo Display Swivel Assembly

    PARTS Vectra® Neo Clinical Therapy System – Service Manual NEO DISPLAY SWIVEL ASSEMBLY ITEM PART DESCRIPTION NUMBER NUMBER REQ'D 13-7589 COVER SWIVEL NEO 13-7590 PLATE SWIVEL NEO 13-7659-3 SCREW, 8-32x5/16”, PH PNHD 13-7658-1 SCREW, #8-16x3/8”, TRI-SHANK, TORX, PNHD 13-7656 GROMMET EDGING 1.25"...
  • Page 61: Neo Desktop Assembly

    Vectra® Neo Clinical Therapy System – Service Manual NEO DESKTOP ASSEMBLY ITEM PART DESCRIPTION NUMBER NUMBER REQ'D 12-6779 HUB ASSEMBLY, DESKTOP 12-6775 DISPLAY SWIVEL ASSEMBLY 13-7697-1 SCREW, 6-32X 5/8” PH PNHD 13-7609 LEFT CRADLE APPLICATOR 13-7610 RIGHT CRADLE APPLICATOR 12-6774-1 DISPLAY ASSY VECTRA NEO...
  • Page 62 VECTRA NEO 40-0107 POWER CORD NEMA 5-15 MEDICAL GRADE 48095 POLYBAG 16 X 14 X 36 X .003 13-7680-1 CARTON HEAD UNIT NEO 13-7680-2 INSERT BOTTOM HEAD UNIT NEO 13-7680-3 INSERT TOP HEAD UNIT NEO 13-7647 IFU CD VECTRA NEO...
  • Page 63 Vectra® Neo Clinical Therapy System – Service Manual NEO CART ITEM PART DESCRIPTION NUMBER NUMBER REQ'D 12-6780 CHASSIS CART VECTRA NEO 13-7624 DRAWER BIN SHORT NEO 13-7624 DRAWER BIN TALL NEO 13-7658-1 SCREW, #8-16x3/8”, TRI-SHANK, TORX, PNHD 13-7618 COVER LOWER CART NEO...
  • Page 64 ITEM PART DESCRIPTION NUMBER NUMBER REQ'D 13-7640 CART LEG NEO 13-7645 LABEL, CHATTANOOGA LOGO ROUND 20898 CASTER 3” BLK W/LOCKING BRAKE 20899 CASTER 3” BLK W/O BRAKE 12-6787 NEO CART 13-7662-1 SCREW, #1/4-20x1-1/4”, SOCHD CAP SCREW, 10-24X1/2” PH PNHD SS...
  • Page 65: Maintenance

    MAINTENANCE Vectra® Neo Clinical Therapy System – Service Manual CLEANING THE VECTRA® NEO CLINICAL THERAPY FUSE INFORMATION SYSTEM With the system disconnected from the power source, clean the system with a clean, lint-free cloth moistened 3.15 A 250V AC with water and mild antibacterial soap. If a more sterile cleaning is needed, use a cloth moistened with an antimicrobial cleaner.
  • Page 66: Warranty

    WARRANTY Vectra® Neo Clinical Therapy System – Service Manual WARRANTY REPAIR/OUT OF WARRANTY REPAIR Service When the Vectra® Neo Clinical Therapy System or any of the accessory modules require service, contact the selling dealer or DJO Service Department. All Therapy System and accessory modules returned to the factory for service must include the following: 1.
  • Page 67: Warranty

    WARRANTY Vectra® Neo Clinical Therapy System – Service Manual WARRANTY COMPANY SHALL NOT BE LIABLE IN ANY EVENT FOR DJO, LLC (“Company”) warrants that the Vectra® Neo INCIDENTAL OR CONSEQUENTIAL DAMAGES. Clinical Therapy System, Channel 1/2 Electrotherapy Module, Channel 1/2 Electrotherapy+EMG Module, Some states do not allow the exclusion or limitation Channel 3/4 Electrotherapy Module, Laser Module, of incidental or consequential damages, so the above...
  • Page 68: Common Terms

    APPENDIX Vectra® Neo Clinical Therapy System – Service Manual COMMON TERMS Applicator – Hand held assembly that delivers Laser Wavelength – Wavelength is measured in nanometers energy and includes laser head, diode, and related (nm) and is the key component in obtaining e ective electronics therapy as di erent wavelengths cause di erent physiological e ects.
  • Page 69 APPENDIX Vectra® Neo Clinical Therapy System – Service Manual INTENSITY KNOB ADJUSTMENT PROCEDURE SCREEN REPLACEMENT PROCEDURE 1. Remove PCB from housing. 1. Tilt screen completely vertical. 2. Flip over PCB and press the two protruding lips. 2. Remove side wings. (6 screws) 3.
  • Page 70 DJO, LLC 1430 Decision Street Vista, CA 92081-8553 USA DJOGlobal.com © 2014 DJO, LLC. All rights reserved. Vectra® Neo Service Manual 00-1296_B...