E2648-Cl2 Rev 10.05.2023 Chlorine Chlorine is a heavy green-yellow gas with a pungent, irritating odor. It is highly toxic and corrosive. It is used in the chemical industry and for bleaching and disinfection. Chemical formula Molar weight 70.9 Relative gas density (to air) 2.44 Conversion 1 ppm = 2.90 mg/m...
E2648-Cl2 Rev 10.05.2023 Specifications Sensor type Electrochemical cell Sampling method Diffusion Typical detection range 0...10 ppm Maximum overload 100 ppm Resolution / digital unit 0.1 ppm Response time T90 <60s Sensor lifetime > 1 years Calibration interval 6 months Signal update Every 1 second <...
E2648-Cl2 Rev 10.05.2023 Product description E2648 series heavy duty detectors-transmitters belong to the PluraSens® family of multifunctional measurement instruments. The instruments utilize gas sensors of various types with excellent repeatability, stability, and long lifetime. Due to its IP66 enclosure of die-cast aluminum, the device may be used under harsh conditions.
E2648-Cl2 Rev 10.05.2023 Operating conditions The device should be used both in a non-hazardous area and in a basic electromagnetic environment, where the latter is defined in EN 61326-1. Avoid strong mechanical shock and vibrations. Avoid corrosive atmosphere and areas highly contaminated with dust, oil mist, etc.
E2648-Cl2 Rev 10.05.2023 Sensor probe handling The E2648 series devices are available with a remote probe. The remote probe is connected to the main unit with a shielded cable. The default remote probe cable length is 3 m. Wall mount remote probe with fixing clamp (default version), Remote probe with rubber flange and three self-tapping screws (on request) The sensor probes of all types are equipped with a hydrophobic microporous PTFE filter to protect the sensor from dust, dirt, and water drops.
E2648-Cl2 Rev 10.05.2023 Electrical connections Unscrew four lid screws and detach the lid from the device. Use the M16 cable gland to let in cables of the power supply and of the external devices. Attach the power cable to the device without turning it on.
Page 11
E2648-Cl2 Rev 10.05.2023 Jumpers OUT1 type (open: 4-20 mA; closed 0-10 V) OUT2 type (open: 4-20 mA; closed 0-10 V) Reset Modbus network parameters to default X4 terminals OUT1 4-20 mA / 0-10 V output OUT2 4-20 mA / 0-10 V output 0 V / 24 VAC Neutral (optional) RS485 A / Data + RS485 B / Data -...
E2648-Cl2 Rev 10.05.2023 NOTE! The outputs are not galvanically isolated from the external power supply and share a common 0V. Allowed load resistance limits are stated in the Specifications table. To power the instrument from an external power source, connect terminals 0V and +U to the source.
E2648-Cl2 Rev 10.05.2023 Operation Turn on the power. The instrument warm-up time takes about 1 minute after switching on and the final sensor stabilization time to maximum accuracy takes about 30 minutes, but depending on the sensor type, it can take up to several hours. The operating status is indicated by the LED on the PCB of the device.
E2648-Cl2 Rev 10.05.2023 Configuring A standard configuration kit includes a USB-RS485 converter, fixed flow regulator, gas tubing with applicators, and a software pack. Please contact your Seller for more information. Detectors- transmitters E2648 share all functionalities of the PluraSens® multifunctional platform.
E2648-Cl2 Rev 10.05.2023 Modbus RTU Communication RS485 communication interface Databits: 8 Supported Modbus functions: Parity: none / odd / even 03 – Read multiple registers Stop bits: 1 or 2 06 – Write a single register Protocol: Modbus RTU Communication parameters Parameter Permitted values Default...
Page 17
E2648-Cl2 Rev 10.05.2023 0x0006 6 / 40007 RW Response delay, ms 1...255 1 – No parity bit, 1 stop bit (default after factory reset) 2 – No parity bit, 2 stop bits 3 – Odd parity, 1 stop bit 0x0007 7 / 40008 RW Stop bits, parity bit * 4 –...
Page 18
E2648-Cl2 Rev 10.05.2023 0 – none 2 – gas concentration 212 / Parameter tied to relay 0x00D4 9- – control by Modbus 40213 control, state set in MHR / 40215 213 / Forced state for relay 0x00D5 0 –off, 1 – on 40214 RE1*** Forced state for relay...
Page 19
E2648-Cl2 Rev 10.05.2023 221 / LOW setpoint for relay 0x00DD 0...65535 (gas units) 40222 Specifications 222 / HIGH setpoint for relay 0x00DE 0...65535 (gas units) 40223 Specifications 223 / LOW setpoint for relay 0x00DF 0...65535 (gas units) 40224 Specifications 224 / HIGH setpoint for relay 0x00E0 0...65535 (gas units)
Page 20
E2648-Cl2 Rev 10.05.2023 257 / 0x0101 Raw gas sensor data 0...4095, ADC units 40258 259 / Gas concentration, signed integer, 0x0103 40260 gas units -32000...+32000 (gas units) 261 / 0% value User 0x0105 -32000...+32000 gas units 40262 for analog output 1 defined 262 / 100% value...
Manufacturer or damaged by customer error or negligence or if there has been an unauthorized modification. Manufacturer contacts Evikon MCI OÜ Teaduspargi 9, Tartu 50411 Estonia info@evikon.eu...