Page 3
Dear Customer, Congratulations on having chosen a quality ELBUR PB 636 IV nursing bed which will certainly fully meet your expectations. We would like to take this opportunity to thank you for showing such confidence and trust in our company and purchasing one of our products.
General information This instruction for use is part and parcel of the product and must accompany every product sold. Article name, [number] Instruction for use of ELBUR PB 636 IV, [35165] Document number, TF-13.5 version, date of issue ver.01, 18.10.2021 All rights reserved - no part of this manual may be reproduced in any form without written permission of the manufacturer.
1.0. Product description 1.1. Intended purpose and range of application ELBUR PB 636 IV is classified as a Medical Device Class 1 in accordance with the Medical Device Regulation 2017/745 on medical devices. It is developed, manufactured and tested according to the current state of technical knowledge and all relevant safety standards for medically used beds.
Page 7
Handset. The Actuators, Handset and Power Cable are connected to a Control Box, which is fixed onto Mattress Support Platform ELBUR PB 636 IV is equipped with Full-length Side Rails and Lifting Pole in the standard configuration. Instead of the Full-lenth Wooden Side Rails, ELBUR Nursing Bed described in this manual can be equipped also with Split Side Rails.
ELBUR PB 636 IV Basic UDI ELBUR nursing GS1 Company Internal Number Valid-check bed Model Prefix / Model Ref. character pair ELBUR PB 636 IV 590426134 ELBURPB636IV 590426134ELBURPB636IVHP Description Description MSP (Head Section) Back Rest Section MSP (Leg Section) Leg Rest Section...
Page 9
* The safe working load equal to 175 kg is calculated as follows (as specified by EN 60601-2-52): Maximum Patient Weight: 140 kg Mattress: 20 kg Accessories: 15 kg ** Electrical data values visible on the Name Plate of the ELBUR Nursing Bed. Elbur sp. z o.o. sp.k. 9 of 56...
ELBUR PB 636 IV 1.3. Labelling of Nursing Beds Manufacturer’s Name Plate (including serial number*) [located on the Scissor Lift inner tube] *Please quote SN on all correspondence regarding sales and customer service information or to report unexpected operation. Warning / Information label no. 35350...
ELBUR PB 636 IV Warning / Information label no. 35372 (Indication of the appropriate assembly of drive units) [located underneath the Bed] Warning / Information label no. Linak CA40 (Indication of the appropriate cables connection to Control Box channels) [located on the CA40 Control Box] zzzz 1.4.
Safety/EMC Regulations 2.0. Definition of the groups of persons involved WARNING All persons working with the ELBUR Nursing Beds must be acquainted with the safety instructions contained in this manual. Improper operation can result in personal injuries or product damage.
Any actions that are inconsistent with the manual are performed at your own risk and can result in serious injury or death. Elbur sp. z o.o. sp.k. shall not be held liable for any damage, injuries or accidents arising from unauthorised modifications, non-genuine spare parts, negligence or use that is at variance with this manual.
Page 14
ELBUR PB 636 IV WARNING manual. The operating staff must make the patient aware of the control functions that apply to the patient subject to an assessment by a professional. Two people qualified in Nursing Bed’s operating are recommended (not required) to perform the assembly process.
Page 15
2 minutes, which must not be exceeded. For reasons of safety and electromagnetic compatibility, use only original electric components from Elbur that have been released for the Nursing Bed model in order. Otherwise it can caused danger for patients and loss of Bed functionality.
Page 16
Never use area under the Bed as a storage space. Use only original Elbur accessories which are compatible with this Bed – any modification is strictly forbidden without authorization of Elbur. Only use Side Rails that are compatible with this Bed as supplied by Elbur.
Any missing parts, faults or damage must be reported immediately to the carrier and your local distributor or Elbur directly in writing. ELBUR PB 636 IV is delivered unassembled and packed in the following order in several cardboard boxes: 1) Scissor lift with Main Lift Actuator LA40,...
Page 18
ELBUR PB 636 IV Description Reference picture Scissor Lift with Main Lift Actuator LA40 There are two options to order of the Scissor Lift to configurate ELBUR PB 636 IV: Standard model of a) Standard model with total brake PB 636 IV Scissor Lift with...
30 minutes, it should be ready to re-use again. Limit switches turn off the Actuators, when they get to the limit position. ELBUR PB 636 IV is equipped with drive system from Linak, a market leader in electric linear actuator technology worldwide.
Mattress Support Platform and positions of its sections. Elbur sp. z o.o. sp.k. does not accept any liability for the use of these products on other makes. Every element of our additional equipment offer meets the special quality and safety standards of Elbur.
Various accessories can be added to assist patient and carer, but not to every product – it depends on the configuration of the ELBUR Nursing Bed. Accessories intended to use with ELBUR Nursing Beds Article No.
ELBUR PB 636 IV 8.0. Assembly and use CAUTION Before commencing assembly of the Nursing Bed, you must read, understand and follow all instructions in this manual and labelling on the bed carefully. Assembly must be carried out by suitably trained and qualified personnel.
ELBUR PB 636 IV 8.3. Assembly of Control Box and MSP Actuators Caution Take special care in assigning Back Rest and Leg Rest actuators under the Mattress Support Platform. These must not be interchanged. Otherwise it can caused damage of electric components.
ELBUR PB 636 IV Bottom View MSP Head End 1) Control Box Linak CA40; 2) Leg Rest Actuator Linak LA27; 3) Back Rest Actuator Linak LA27. MSP Foot End 8.4. Actuator cable clip protection New type of actuator cable clip Old type of actuator cable clip Elbur sp.
ELBUR PB 636 IV 8.4.1. New type of actuator cable clip To perform correct assembly insert the actuator cable to appropriate channel (A) and push the retaining clip down (B). Removing the retaining clip from the base of actuator is done by squeezing the clip on either side (C) then lifting the clip (D).
ELBUR PB 636 IV holder with the strain relief provided on the cable, otherwise it might be damaged by stretching, running over or squeezing. The handset cable must be routed from the Control Box underneath the Mattress Support Platform. It may come out from the MSP on either side, depending on user requirements.
(e.g. ceramic floor tiles, linoleum, cast floors). ELBUR PB 636 IV with the standard scissor lift (total brake for each castor) Each of the four castors is equipped with a separate brake. 1) To apply the brake press down on the surface of the lower tilting lever downwards by foot as far as it goes (A), until it brakes the castor and locks in the ‘down’...
ELBUR PB 636 IV 1) To apply the brake press down on the surface of the lower tilting lever downwards by foot as far as it goes (A), until it brakes the castor and locks in the ‘down’ position. 2) To release the brake press horizontally the surface of the top tilting lever backward by foot (B), until it springs back into the ‘up’...
Back Rest and Leg Rest sections, either individually or together (auto contour), within a predetermined range. To change the position of the ELBUR PB 636 IV it is necessary to use the Handset. Press and hold one of the buttons on the keypad on the front side of the Handset until desired position is reached.
Page 30
ELBUR PB 636 IV RAISING LOWERING Raise the Back Rest Lower the Back Rest Section Section Raise the Mattress Lower the Mattress Support Platform Support Platform Raise the Leg Rest Lower the Leg Rest Section Section Raise the Back Rest and...
ELBUR PB 636 IV 8.10. Additional adjustment of the Lower Leg Rest WARNING Do not put your fingers between moving parts of the Mattress Support Platform – it can caused their entrapment and injuries! If the Leg Rest is raised using the Handset, its both sections are adjusted in that direction: Upper Leg Rest and Lower Leg Rest.
220 mm a risk assessment must be performed to assure equivalent safety. If the ELBUR Nursing Bed type PB is configurated ELBUR Split Side Rails, gap between: - Split Side Rails and Head End or Foot End,...
Page 33
ELBUR PB 636 IV WARNING Position of drilled holes in the wooden slats of Full-length Side Rails are not the same. Upper wooden slat have two Lower wooden slat have two holes drilled offset downwards. holes drilled offset upwards. does not let to achieve...
Page 34
ELBUR PB 636 IV ATTENTION With every set of ELBUR Split Side Rails Instruction Manual of installation and use is supplied. Please refer to all instructions included in that document, when ELBUR Nursing Bes is configurated with them. The ELBUR Nursing Bed described in this manual is equipped with Full-length Side Rails, made of wooden bars with plastic end caps.
(depends on model they are intended to work with). They cannot be used on any other medically used beds manufactured by other companies. Do not use the ELBUR Side Boards to lift or move the ELBUR Nursing Bed they are fitted on. ATTENTION With every set of ELBUR Side Boards an Assembly Manual is supplied.
Page 36
ELBUR PB 636 IV 1) Attach ELBUR Side Board - Model Standard [R1] to the external side of Mattress Support Platform, matching holes in them at the same time. Insert three M6x30 Button Head Screws through holes in ELBUR Side Boards and Mattress Support Platform from inside of MSP.
Page 37
2) Assembly of Split Side Rails by putting threaded pins through holes in ELBUR Side Boards and Mattress Support Platform. Short Split Side Rail is assembled in the Head End of the ELBUR Nursing Bed. Long Split Side Rail is assembled in the Foot End of the ELBUR Nursing Bed.
4) Position the top of Lifting Pole so that it is above the centre of ELBUR nursing Bed. Rotate the Lifting Pole from side to side until the cutout locates in the bar in the vertical tube.
Page 39
ELBUR PB 636 IV Pole. If it does rotate, lift the Lifting Pole up a little, then lower it again and attempt to locate the cutout on the bar. 5) Strap should be placed over the top of the Lifting Pole so its rubber pad (A) rests on the top of the Lifting Pole between two lugs (B and C) on the top of the horizontal part of the Lifting Pole.
ELBUR PB 636 IV 9.0. Movements of Nursing Beds WARNING This bed is not adapted for patient’s transport, but it can be moved within patient room for cleaning or patient access. Due to force the propulsion, it has to be done by two people. Patient has to stay in flat position and nursing bed should be adjusted in the lowest height.
Page 41
ELBUR PB 636 IV CAUTION Check all cable ties have been removed (there is one cable tie for attaching Back Rest to MSP Head Section and one cable tie for attaching Leg Rest Section to MSP Leg Section for transport).
Elbur cannot be liable for any damage or risk of damage if inappropriate cleaning or disinfectant agents are used. Observing the instructions described in this chapter will retain the usability of the ELBUR nursing bed for a long time to come and will keep its nice appearance. Cleaning and disinfecting procedure: 1) Empty the nursing bed - remove the mattress and all accessories.
All functions must be tested and approved by a competent person after troubleshooting. ATTENTION Even if the nursing bed is used properly, a technical problem may occur. In this case, always contact your distributor or Elbur Customer Service directly. DEFECT POSSIBLE CAUSES...
ELBUR PB 636 IV DEFECT POSSIBLE CAUSES SOLUTION Individual Actuator, Control Box Notify operator Elbur Actuators or Handset defective Customer Service one direction only Opposite functions Mixed cables Check the plug-in connection in when operating the connection Control Box, follow...
14.0. Disposal CAUTION If the ELBUR Nursing Bed is to be scrapped, the synthetic and metallic parts are to be separated and disposed of properly. The operator must ensure that all components of the Bed are to be disposed of are not...
Page 46
To ensure bed user or bed patient safety and to prolong the lifetime of the ELBUR nursing bed in safe condition, the company Elbur sp. z o.o. sp.k. as manufacturer specifies an inspection interval, which stipulates that a safety- technical inspection is to be executed at least once annually, before passing on to a new user and after every repair.
Page 47
ELBUR PB 636 IV Inspection Protocol for ELBUR Nursing Beds type PB (Form Number: PSF-08.9-01 ; date of issue: 17.05.2021) Customer Name / Medical Facility Address Model designation ELBUR PB ________ Reference Number Serial Number Electrical appliance class Date of manufacture Manufacturer Elbur sp.
Page 48
ELBUR PB 636 IV 1. Visual Inspection Description Are the Name Plate and all Warning/Information labels present, readable and easy to understand? Is the Instruction for Use available? Are the Maximum Patient Weight and Safe Working Load observed and not exceeded?
Page 49
Is the area around, above and below the bed clear of possible obstruction? 2. Performance check Description Is it possible to operate the ELBUR bed as per its intended purpose without unexpected noise or motion? Are all electrical positioning functions, available to...
Page 50
ELBUR PB 636 IV 2. Performance check Description Is it possible to lock/unlock individual functions on the Handset with a special Locking Key? Are drive units working properly (allow the smooth operation without any abnormal noises)? Are the Mattress Support Platform and Bed Ends mechanically sound i.e.
Page 51
Results of Inspection Date of inspection Inspected by Signature PASS FAIL In the event the Inspection result did not pass and ELBUR Repair nursing bed did not meet the safety requirements Take out of use Additional comments / Description of defects...
16.0. Electromagnetic Compatibility statement Guidance and manufacturer’s declaration – Electromagnetic emissions The ELBUR nursing bed is intended for use in the electromagnetic environment specified below. The customer or the user of the bed should assure that it is used in such an environment.
Page 53
ELBUR PB 636 IV Guidance and manufacturer’s declaration - Electromagnetic immunity The ELBUR nursing bed is intended for use in the electromagnetic environment specified below. The customer or the user of the bed should assure that it is used in such environment.
ELBUR nursing bed The ELBUR nursing bed is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the ELBUR nursing bed can help prevent electromagnetic...