Notices on Cleaning
Sterilisation of the Table
7.2.2
Disinfecting the Device
20.12.2023
To clean the surfaces, use a soft cloth, ideally lint-free. This applies
in particular to the operating panel. Disinfect using commercially avail-
able disinfectant if necessary. Only use permitted auxiliary materials
for cleaning and disinfection. After cleaning, use a cloth to dry the
device.
If necessary, the table incl. guide rollers can be sterilised at tempera-
tures up to 120 °C. In contrast, the basic unit and operating panel must
not be sterilised.
Only use quaternary ammonium compounds for wipe-down disinfec-
tion. As a tried as tested disinfectant, we recommend Fermacidal D2.
NOTICE
Heat (temperatures above 80 °C), aggressive disinfectants such as
chlorine bleach and UVC radiation can damage the device and signif-
icantly limit the function and service life of the device.
We recommend against using UV lamps for disinfection of the
device because the UV rays can massively damage the housing in
case of multiple applications.
Celltron - Operating manual
Cleaning and Maintenance
Page 41 of 56