TEXIO ASR-RK Series Instruction Manual

Ac/dc power source

Advertisement

Quick Links

AC/DC POWER SOURCE
ASR902-351RK
ASR133-351RK
ASR243-351RK
This instruction manual is a supplementary manual for the ASR-RK series.
For detailed operation information, please refer to the instruction manual for
the ASR-4.5k / 6k series.
INSTRUCTION MANUAL
ASR-RK SERIES
ASR123-351RK
ASR183-351RK

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ASR-RK Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for TEXIO ASR-RK Series

  • Page 1 INSTRUCTION MANUAL AC/DC POWER SOURCE ASR-RK SERIES ASR902-351RK ASR123-351RK ASR133-351RK ASR183-351RK ASR243-351RK This instruction manual is a supplementary manual for the ASR-RK series. For detailed operation information, please refer to the instruction manual for the ASR-4.5k / 6k series.
  • Page 2 ■ About Brands and Trademarks “TEXIO” is the product brand name of our industrial electronic devices. All company names and product names mentioned in this manual are the trademark or the registered trademark of each company or group in each country and region.
  • Page 3: Table Of Contents

    Table of Contents USING THE PRODUCT SAFELY ..............I ..............1 1. GETTING STARTED 1-1. ASR Series Overview ................1 ................ 1 1-1-1. Series lineup ..............2 1-1-2. Operating Area ................7 1-1-1. Accessories 1-2. Appearance ..................... 8 ................8 1-2-1.
  • Page 4: Using The Product Safely

    USING THE PRODUCT SAFELY ■ Preface To use the product safely, read instruction manual to the end. Before using this product, understand how to correctly use it. If you read the manuals but you do not understand how to use it, ask us or your local dealer. After you read the manuals, save it so that you can read it anytime as required.
  • Page 5 USING THE PRODUCT SAFELY WARNING CAUTION ■ Do not remove the product's covers and panels Never remove the product's covers and panels for any purpose. Otherwise, the user's electric shock or fire may be incurred. ■ Warning on using the product Warning items given below are to avoid danger to user's body and life and avoid the damage or deterioration of the product.
  • Page 6 USING THE PRODUCT SAFELY ■ Warning item on Grounding If the product has the GND terminal on the front or rear panel surface, be sure to ground the product to safely use it. ■ Warnings on Installation environment ● Operating temperature and humidity Use the product within the operating temperature indicated in the “rating”...
  • Page 7 USING THE PRODUCT SAFELY ■ Input / Output terminals Maximum input to terminal is specified to prevent the product from being damaged. Do not supply input, exceeding the specifications that are indicated in the "Rating" column in the instruction manual of the product. Also, do not supply power to the output terminals from the outside.
  • Page 8: Getting Started

    GETTING STARTED This chapter describes the ASR-RK series power supply in a nutshell, including its main features and front / rear panel introduction. This manual is a supplementary manual for the ASR-RK series. For detailed operation information, please refer to the ASR-4.5k / 6k instruction manual.
  • Page 9: Operating Area

    Model Name Power Rating Max. Output Current Max. Output Voltage ASR902-351RK 6000VA 30 / 15A 700Vrms / 1000Vdc ASR123-351RK 8000VA 40 / 20A 700Vrms / 1000Vdc ASR133-351RK 9000VA 45 / 22.5A 700Vrms / 1000Vdc ASR183-351RK 12000VA 60 / 30A 700Vrms / 1000Vdc ASR243-351RK 16000VA 80 / 40A...
  • Page 10 ASR902-351RK DC Mode Operating Area ASR123-351RK AC Mode Operating Area...
  • Page 11 ASR123-351RK DC Mode Operating Area ASR133-351RK AC Mode Operating Area...
  • Page 12 ASR133-351RK DC Mode Operating Area ASR183-351RK AC Mode Operating Area...
  • Page 13 ASR183-351RK DC Mode Operating Area ASR243-351RK AC Mode Operating Area...
  • Page 14: Accessories

    ASR243-351RK DC Mode Operating Area Accessories 1-1-1. Before using the ASR-RK power source unit, check the package contents to make sure all the standard accessories are included. Standard Part number Description Accessories Safety guide 62SR-6KDSC201 Input terminal cover 62SR-6KDSC301 62SR-6KDSC501 Output terminal cover 62SR-6KDSC601 GTL-246...
  • Page 15: Appearance

    1-2. Appearance 1-2-1. Front Panel...
  • Page 16 Index Description Power switch button USB interface connector (A Type) LCD screen Function keys (blue zone) Menu key Test key Preset key Scroll wheel Range key / Output mode key Arrow keys Output key Shift key Cancel key Enter key Irms / IPK-Limit button Lock / Unlock button F / F-Limit button...
  • Page 17 Function Keys Assigned to the functions displayed on the right side of the screen. Menu Key Enters the Main menu or goes back to one of the display modes. Test Key Puts the instrument into the Sequence and Simulation control mode. Preset Key Puts the instrument into Preset mode.
  • Page 18 Cancel Key Used to cancel function setting menus or dialogs. Enter Key Confirms selections and settings. Irms Used for setting the maximum output current. IPK-Limit Used to set the peak output current limit value. Lock/Unlock Key Used to lock or unlock the front panel keys except output key.
  • Page 19 Local Mode Switches operation back to local mode from remote mode. IPK CLR Used to clear peak output current value. ALM CLR Clears alarms. Hardcopy Key Used to take a screenshot. Make sure an USB flash disk in well inserted before the action.
  • Page 20: Rear Panel

    1-2-2. Rear Panel...
  • Page 21 Item Index Description Output terminal AC power input terminal Remote sensing input terminal External I/O connector External IN/OUT connection in parallel function RS232 connector Ethernet (LAN) connector USB interface connector (B Type) Optional interface Slot ▪ GPIB card (ASR-003) Item Description Output Terminal Output terminal...
  • Page 22: Operation

    OPERATION We take the illustration of 3P4W Input Connection here for example. Please refer to page 16 of the Input Terminal Connection chapter for the detailed information covering the 2 different connection methods. 2-1. Power Up and Procedure Steps Connect the AC power cords to the AC input terminals.
  • Page 23: Input Terminal Connection

    Press the POWER key. The welcome screen of TEXIO will be displayed followed by self-checking procedure before the continuous mode screen appears with the settings loaded. ⚫ If the warning message of “Parallel Communication Error” appears in the screen display, turn Off both POWER key and...
  • Page 24 Install the protective lid of power input terminals followed by fastening the single screw to fix the lid firmly into place. The AC power cords of 3P3W input are connected with the AC input terminals equipped with protective lid completely. Note ⚫...
  • Page 25: Output Terminal Connection

    Input Terminal 3P4W Connection Steps Connect the AC power cords to the AC input terminals. ⚫ Red⇒L3 Green⇒L2 ⚫ ⚫ Yellow⇒L1 ⚫ Blue⇒Neutral Install the protective lid of power input terminals followed by fastening the single screw to fix the lid firmly into place.
  • Page 26 Be aware of dangerous voltages. Ensure that the power to the instrument is disabled before handling the power supply output WARNING terminals. Failing to do so may lead to electric shock. After configuring phase settings via the front panel, please make sure the cords connection on the rear panel is corresponding to the set CAUTION configuration.
  • Page 27 Note Grounded Neutral Output for 1P2W output only: ASR-RK allows for a grounded return on the neutral output. It is suit for the medical industry that required between ground with neutral is 0 V essentially. And possible to mitigate ground loops that is ideal for reduce ground noise and isolate sensitive equipment from the effects of ground loops.
  • Page 28 Install the protective lid of power output terminals followed by fastening the single screw to fix the lid firmly into place. The AC power cords of 1P3W output are connected with the AC output terminals equipped with protective cover and lid completely. Note ⚫...
  • Page 29: Remote Sensing, Ext I/O And Interface Connection

    Install the protective lid of power output terminals followed by fastening the single screw to fix the lid firmly into place. The AC power cords of 3P4W output are connected with the AC output terminals equipped with protective cover and lid completely. Note ⚫...
  • Page 30: Appendix

    APPENDIX 3-1. Firmware Update Background The ASR series firmware can be upgraded using the USB A port on the front panel. See your local distributor or the TEXIO website for the latest firmware information. https://www.texio.co.jp ⚫ Both Master and Slave ASR units are required to be plugged in...
  • Page 31 7. Insert USB flash drives into the USB A-type ports on front panel of both Master and Slave units. The USB drives should include the texio_sb6.upg file in a directory name“TEXIO”. Press the Menu key on the Master unit. and...
  • Page 32: Function Difference Table

    2~3 units flexible Fixed 3-3. Factory Default Settings The following default settings are the factory configuration settings for the ASR-RK series. For details on how to return to the factory default settings, refer to the User Manual of ASR- 4.5k / 6k.
  • Page 33 Syc Phs Freq Limit 2000 Hz Vrms Limit 175.0 Vrms VPK+ Limit + 250 V VPK- Limit - 250 V +120.0 +360.0 +160.0 +480.0 +180.0 +540.0 +240.0 +720.0 +320.0 +960.0 IPK+ Limit -120.0 -360.0 -160.0 -480.0 -180.0 -540.0 -240.0 -720.0 -320.0 -960.0 IPK- Limit...
  • Page 34 ASR902-351RK ASR123-351RK ASR133-351RK ASR183-351RK ASR243-351RK DHCP ASR902-351RK ASR123-351RK ASR133-351RK ASR183-351RK ASR243-351RK USB Device Speed Full Mode ASR902-351RK ASR123-351RK ASR133-351RK ASR183-351RK ASR243-351RK RS232C Baudrate 9600 Databits 8bits Parity None Stopbits 1bit ASR902-351RK ASR123-351RK ASR133-351RK ASR183-351RK ASR243-351RK GPIB Address ASR902-351RK ASR123-351RK ASR133-351RK ASR183-351RK ASR243-351RK Sequence Mode...
  • Page 35: Error Messages & Messages

    Trig Out ON Phs Free OFF Phs Free Fixed(0) 120 Fixed(0) 120 Fixed(0) 120 Fixed(0) 120 Fixed(0) 120 Phase ASR902-351RK ASR123-351RK ASR133-351RK ASR183-351RK ASR243-351RK Sequence Mode Step Initial Repeat Time 0.1000 S Term Free Code Item 0.00 0.00 0.00 0.00 0.00 Fset 50.00...
  • Page 36 Limited check allowance set range Message Only Setting Phase Setting ON/OFF Phase Limited Display Limited Message Only Setting Duty Limited Setting Duty be limited Display Message Only Invalid with Output Invalid with Output ON Display Message Only Rear USB Port Rear USB port connected to PC Display Connected To PC...
  • Page 37 Message Only ARB# Deleted ARB1~ARB253 Deleted Display Message Only Save All Data Ready to save all data (Preset0~9 + SEQ0~9 + Display SIM0~9 + ARB1~253) Message Only All Data Saved All data are saved successfully (Preset0~9 + Display Success SEQ0~9 + SIM0~9 + ARB1~253) Message Only Recall All Data Ready to recall all data (Preset0~9 + SEQ0~9 +...
  • Page 38 Sequence Message Only Invalid in This Invalid Operation In This Simulate Display Simulate Message Only SEQ#Deleted SEQ0~SEQ9 Deleted Display Message Only SIM#Deleted SIM0~SIM9 Deleted Display Message Only Cleared SEQ# Cleared SEQ0~SEQ9 Display Message Only Cleared SIM# Cleared SIM0~SIM9 Display Message Only Recalled from Recalled fromSEQ0 ~ SEQ9 Display...
  • Page 39: Specifications

    3-5. Specifications The specifications apply when the ASR-RK is powered on for at least 30 minutes. 3-5-1.Electrical specifications - ASR902-351RK / ASR123-351RK Model ASR902-351RK ASR123-351RK Input ratings Power type Three-phase, Delta or Y connection Three-phase, Delta:200V~240V±10% or Voltage range Three-phase, Y connection:380V~460V±10% Factory specified Frequency range 47Hz to 63Hz...
  • Page 40 0.00V to 0.00V to 606.2V 606.2V (sine wave (sine wave only) only) Setting Setting Resolution: Resolution: 0.01V / 0.1V 0.01V / 0.1V Maximum current 90A / 45A 30A / 15A 120A / 60A 40A / 20A Maximum peak Four times of the maximum RMS current current Load power factor 0 to 1 (leading phase or lagging phase, 45Hz to 65Hz)
  • Page 41 Model ASR902-351RK ASR123-351RK DC output (only single-phase output) Output capacity 12kW Mode Floating output, the N terminal can be grounded -250.0V to +250.0V / -500.0V to +500.0V, Setting Resolution: Setting 0.01V / 0.1V Voltage Range ± (|0.3% of set| + 0.3V / 0.6V) Accuracy Maximum current 90A / 45A...
  • Page 42 ※8.With respect to the capacitor-input rectifying load. Limited by the maximum current. ※9.External power injection or regeneration which is over short reverse power flow capacity is not available. ※10.*For 45Hz to 65Hz, the rated output voltage, no load and the resistance load for the maximum current, and the operating temperature range.
  • Page 43 Measured value display (All accuracy of the measurement function is indicated for 23 °C±5 °C.) Single-phase output Polyphase output Resolution 0.01V / 0.1V 45Hz to 65Hz and DC: 45Hz to 65Hz: ± ± (0.5% of rdg + 0.5V / (0.5 % of rdg + 0.5V / RMS value accuracy 15Hz to 1000Hz: ±...
  • Page 44 Harmonic Range Up to 100th order of the fundamental wave voltage Full Scale 200V / 400V, 100% Effective Resolution 0.01V / 0.1V, 0.1% value (rms) Percent Up to 20th: ± (0.2% of rdg + 0.5V / 1V) Accuracy (AC-INT 21th to 100th: ± (0.3% of rdg + 0.5V / 1V) and 50 / 60 Hz only) Harmonic...
  • Page 45: 3-5-2.General Specifications - Asr902-351Rk / Asr123-351Rk

    Model ASR902-351RK ASR123-351RK Others UVP, OVP, OCP, OTP, OPP, Fan Fail, Peak and RMS Current Protections Limit Parallel Not supported Display TFT-LCD, 7 inches Memory function Store and recall settings, Basic settings: 10 Number of 253 (nonvolatile) memories Waveform Arbitrary Wave 4096 words length Amplitude...
  • Page 46: 3-5-3.Electrical Specifications -Asr133-351Rk / Asr183-351Rk

    EN 61326-1 (Class A) EN 61326-2-1/-2-2 (Class A) EN 61000-3-2 (Class A, Group 1) EN 61000-3-3 (Class A, Group 1) EN 61000-4-2/-4-3/-4-4/-4-5/-4-6/-4-8/- 4-11 (Class A, Group 1) EN 55011 (Class A, Group1) Safety EN 61010-1 Operating environment Indoor use, Overvoltage Category II Operating 0°C to 40°C temperature range...
  • Page 47 Efficiency 80% or higher Maximum power 18kVA or lower 24kVA or lower consumption Model ASR133-351RK ASR183-351RK AC output Single-phase Polyphase Single-phase Polyphase Multi-phase output output output output output 1P3W: 9kVA 1P3W: 12kVA Output capacity 13.5kVA 18kVA 3P4W: 13.5kVA 3P4W: 18kVA 1P3W 1P3W Mode...
  • Page 48 Load power factor 0 to 1 (leading phase or lagging phase, 45Hz to 65Hz) AC Mode: 15.00Hz to 1000.0Hz, AC+DC Mode: 1.00Hz to Setting 1000.0Hz, Setting resolution: 0.01Hz / 0.1Hz range Frequency ± 0.01% of set Accuracy ± 0.005% Stability Output on phase 0.0°...
  • Page 49 Model ASR133-351RK ASR183-351RK Output Stability, Total Harmonic Distortion, Output voltage rising time and Ripple noise Line regulation ±0.1% or less (Phase voltage) Load regulation ±0.5V / ±1.0V (phase voltage, 0 to 100%, via output terminal) <0.3%@1Hz~100Hz, Load regulation <0.5%@100.1Hz~500Hz, <1%@500.1Hz~1000Hz Output voltage Middle: 100μs(typ.)
  • Page 50 +500V, no load, AC voltage set to 0V (AC+DC mode) and 23°C ± 5°C ※16.If the output voltage is higher than rated value, this is limited to satisfy the power capacity. If there is the AC superimposition, the active current of AC+DC satisfies the maximum current.
  • Page 51 ± (0.5% of rdg + 0.3A / of rdg + 0.15A / 0.08A) 0.15A) 15Hz to 1000Hz: ± 15Hz to 1000Hz: ± (0.7% of rdg + 0.3A / (0.7% of rdg + 0.6A / 0.15A) 0.4A) DC: ± (|0.5% of rdg| + DC: ±...
  • Page 52 ※1.In the polyphase output, the specification is for phase voltage, and the DC average value display cannot be selected. ※2.Accuracy values are in the case that the output voltage is within voltage setting range. ※3.The accuracy is for output waveform DC or sine wave only. ※4.Accuracy values are in the case that the output current is 5% to 100% of the maximum current.
  • Page 53: 3-5-4.General Specifications - Asr133-351Rk / Asr183-351Rk

    3-5-4.General Specifications - ASR133-351RK / ASR183-351RK Model ASR902-351RK ASR123-351RK Type A: Host, Type B: Slave, Speed: 2.0, USB-CDC / USB-TMC MAC Address, DNS IP Address, User Password, Gateway IP Address, Instrument IP Address, Subnet Mask Standard External Signal Input Interface External External Control I/O V/I Monitor Output...
  • Page 54: 3-5-5.Electrical Specifications -Asr243-351Rk

    Storage humidity 90% RH or less (no condensation) range Altitude Up to 2000m 598(W)×1116(H)×906(D) (not including Dimensions (mm) protrusions) Weight Approx. 200kg ⚫ A value with the accuracy is the guaranteed value of the specification. However, an accuracy noted as reference value shows the supplemental data for reference when the product is used, and is not under the guarantee.
  • Page 55 range V to 700.0 V 3P4W: 0.00V to 303.1V / 0.00V to 606.2V (sine wave only) Setting Resolution: 0.01V / 0.1V 設定分解能: 0.01V / 0.1V Maximum current 240A / 120A 80A / 40A Maximum peak Four times of the maximum RMS current current Load power factor 0 to 1 (leading phase or lagging phase, 45Hz to 65Hz)
  • Page 56 Maximum current 240A / 140A Maximum peak Four times of the maximum current current Model ASR243-351RK Output Stability, Total Harmonic Distortion, Output voltage rising time and Ripple noise Line regulation ±0.1% or less (Phase voltage) Load regulation ±1 V / ±2V (phase voltage, 0 to 100%, via output terminal) Distortion of <0.3% @1Hz to 100Hz, <0.5% @100.1Hz to 550Hz Output...
  • Page 57 ※13.For an output voltage of 50V or higher, sine wave, same load and voltage condition for all phase. ※14.In the case of the AC mode and output voltage setting to 0 V, 23°C ± 5°C ※15.For an output voltage of -250V to -10V, +10V to +250V / -500V to -20V, +20V to +500V, no load, AC voltage set to 0V (AC+DC mode) and 23°C ±...
  • Page 58 45Hz to 65Hz and DC: ± 45Hz to 65Hz: ± (0.5% (0.5% of rdg + 0.3A / of rdg + 0.15A / 0.08A) RMS value accuracy 0.15A) 15Hz to 550Hz: ± (0.7% 15Hz to 550Hz: ± (0.7% of rdg + 0.3A / 0.15A) of rdg + 0.6A / 0.4A) DC: ±...
  • Page 59 ※1.In the polyphase output, the specification is for phase voltage, and the DC average value display cannot be selected. ※2.Accuracy values are in the case that the output voltage is within voltage setting range. ※3.The accuracy is for output waveform DC or sine wave only. ※4.Accuracy values are in the case that the output current is 5% to 100% of the maximum current.
  • Page 60: 3-5-6.General Specifications - Asr243-351Rk

    3-5-6.General Specifications - ASR243-351RK Model ASR902-351RK ASR123-351RK Type A: Host, Type B: Slave, Speed: 2.0, USB-CDC / USB-TMC MAC Address, DNS IP Address, User Password, Gateway IP Address, Instrument IP Address, Subnet Mask Standard External Signal Input Interface External External Control I/O V/I Monitor Output Complies with the EIA-RS-232 232C...
  • Page 61: 3-5-7.Dimensions

    3-5-7.Dimensions ASR902-351RK / ASR123-351RK Scale=mm...
  • Page 62 ASR133-351RK / ASR183-351RK Scale=mm...
  • Page 63 ASR243-351RK Scale=mm...
  • Page 64 7F Towa Fudosan Shin Yokohama Bldg. 2-18-13, Shin Yokohama, Kohoku-ku,Yokohama, Kanagawa, 222-0033 Japan https://www.texio.co.jp/...