Description Description The JULABO Refrigerated Circulators are compact bench-top models consisting of the refrigerating bath unit and circulator. Regarding the circulator, please refer to the separate service manual. The refrigerating bath units are equipped with virtually noiseless and vibration-free cooling machines containing a sealed compressor.
Operation Operation Before operating the unit after transport, wait about one hour after setting it up. This will allow any oil that has accumulated laterally during transport to flow back down. Check to make sure that the line voltage matches the supply voltage specified on the identification plate.
General safety recommendations and repair instructions 6.2. Troubleshooting Disconnect the mains power plug before opening the unit! When the overload protection device is triggered although the condenser is clean, or the compressor does not start, check the electronics for possible damages before disassembling the cooling machine.
General safety recommendations and repair instructions 6.3. Cleanliness during assembly The greatest dangers to hermetically sealed compressors are dirt and moisture. Particular care is therefore required when assembling such units: Use only cooling coils (evaporators), condensers and tubing that are clean and dry on the inside and have been delivered airtight.
Spare parts list F12 Spare parts list F12: Print date 05.10.06 7.920.2212 POS. PART DESCRIPTION ORDER-NO. FROM FROM DATE BARCOD E NO. Wiring diagram F12 230V 50 Hz Euro 1.590.0228-V0 Wiring diagram F12 115V 60Hz 1.590.0228-V1 Cooling control board circuit diagram 7.537.3020 0S0...
Wiring diagrams Wiring diagrams 230V 115 V Wiring diagram F12 1.590.0228-V0 1.590.0228-V1 Refrigerant circuit diagram F12 1.650.0033 Vocabulary used on wiring diagrams (in alphabetical order): German expression English expression Badmantel .............Tank housing Einbaudose ............Panel socket Kompressor............Compressor Rückwand ............Rear panel Lüftermotor............Ventilator motor Netzschalter ............Mains switch...
Page 13
ZE-Bodenblech ZE-Rückwand Si oben Si unten Zeichnung Blatt Gegenstand Teile-Nummer Ansicht Datum erstellt: Kurz Dat: Verdrahtungsplan 1 v 1 F12 115V mit Sich. 1.590.0228-V1 04.05.1999 geprüft: Rother L-Kl.:...
Page 15
Kap. Maßstab / Scale Allgemeintoleranz Tolerance ISO 2768-mH Date Name RI-Fließbild F12 Editor 05.10.06 W. Lindemann Inspect. 05.10.06 RI-flow chart F12 Orig. 05.10.06 W. Lindemann Bl./sheet Ident.-No. 1.650.0033 Ind. Date Name Revision...