Julabo F12 Service Manual
Julabo F12 Service Manual

Julabo F12 Service Manual

Refrigerating bath unit

Advertisement

Quick Links

Refrigerating Bath Unit
F12
1.921.2212.0 - 04/06
19212212.doc
Service Manual
Innovative Temperature Technology
JULABO Labortechnik GmbH
77960 Seelbach / Germany
+49 (0) 7823 / 51-0
+49 (0) 7823 / 24 91
info@julabo.de
www.julabo.de
05.10.2006

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the F12 and is the answer not in the manual?

Questions and answers

Summary of Contents for Julabo F12

  • Page 1 Service Manual Refrigerating Bath Unit Innovative Temperature Technology JULABO Labortechnik GmbH 77960 Seelbach / Germany  +49 (0) 7823 / 51-0  +49 (0) 7823 / 24 91  info@julabo.de 1.921.2212.0 - 04/06  www.julabo.de 19212212.doc 05.10.2006...
  • Page 2: Important Note

    Copyright © 2005 The Content of this manual is internationally protected for JULABO Labortechnik GmbH, Seelbach, Germany. All rights are reserved. All information, including texts, pictures and other content, may not be multiplied, distributed, transmitted, stored, passed on or put to use in any other form without previous, explicit written authorization.
  • Page 3: Table Of Contents

    Safety recommendations ..................5 6.2. Troubleshooting ..................... 6 6.3. Cleanliness during assembly ................7 6.4. Leak detection and elimination ................7 6.5. Filling refrigerant ....................8 6.6. Cooling capacity data..................... 8 Spare parts list F12 ......................9 Wiring diagrams ......................12...
  • Page 4: Description

    Description Description The JULABO Refrigerated Circulators are compact bench-top models consisting of the refrigerating bath unit and circulator. Regarding the circulator, please refer to the separate service manual. The refrigerating bath units are equipped with virtually noiseless and vibration-free cooling machines containing a sealed compressor.
  • Page 5: Operation

    Operation Operation Before operating the unit after transport, wait about one hour after setting it up. This will allow any oil that has accumulated laterally during transport to flow back down. Check to make sure that the line voltage matches the supply voltage specified on the identification plate.
  • Page 6: Troubleshooting

    General safety recommendations and repair instructions 6.2. Troubleshooting Disconnect the mains power plug before opening the unit! When the overload protection device is triggered although the condenser is clean, or the compressor does not start, check the electronics for possible damages before disassembling the cooling machine.
  • Page 7: Cleanliness During Assembly

    General safety recommendations and repair instructions 6.3. Cleanliness during assembly The greatest dangers to hermetically sealed compressors are dirt and moisture. Particular care is therefore required when assembling such units:  Use only cooling coils (evaporators), condensers and tubing that are clean and dry on the inside and have been delivered airtight.
  • Page 8: Filling Refrigerant

    Refrigerant / Elektrolux GD36AA 230 V/50 Hz R134a -10,0 °C / Elektrolux GD36ME 115 V/60 Hz R134a -12,5 °C F12 USA / Elektrolux GD40ME 115 V/60 Hz R134a -3,5 °C F12 Korea / GD36AF-a ACC 230 V/60 Hz R134a -8 °C 6.6.
  • Page 9: Spare Parts List F12

    Spare parts list F12 Spare parts list F12 Example: POS. NO. 140 includes the POS. NO. 140 to 149...
  • Page 10 Spare parts list F12...
  • Page 11: Part Description

    Spare parts list F12 Spare parts list F12: Print date 05.10.06 7.920.2212 POS. PART DESCRIPTION ORDER-NO. FROM FROM DATE BARCOD E NO. Wiring diagram F12 230V 50 Hz Euro 1.590.0228-V0 Wiring diagram F12 115V 60Hz 1.590.0228-V1 Cooling control board circuit diagram 7.537.3020 0S0...
  • Page 12: Wiring Diagrams

    Wiring diagrams Wiring diagrams 230V 115 V Wiring diagram F12 1.590.0228-V0 1.590.0228-V1 Refrigerant circuit diagram F12 1.650.0033 Vocabulary used on wiring diagrams (in alphabetical order): German expression English expression Badmantel .............Tank housing Einbaudose ............Panel socket Kompressor............Compressor Rückwand ............Rear panel Lüftermotor............Ventilator motor Netzschalter ............Mains switch...
  • Page 13 ZE-Bodenblech ZE-Rückwand Si oben Si unten Zeichnung Blatt Gegenstand Teile-Nummer Ansicht Datum erstellt: Kurz Dat: Verdrahtungsplan 1 v 1 F12 115V mit Sich. 1.590.0228-V1 04.05.1999 geprüft: Rother L-Kl.:...
  • Page 14 1.590.0228_V0.JPG (2480x3508x256 jpeg)
  • Page 15 Kap. Maßstab / Scale Allgemeintoleranz Tolerance ISO 2768-mH Date Name RI-Fließbild F12 Editor 05.10.06 W. Lindemann Inspect. 05.10.06 RI-flow chart F12 Orig. 05.10.06 W. Lindemann Bl./sheet Ident.-No. 1.650.0033 Ind. Date Name Revision...