Ranger RD
3.3
3.3.1
Internal
External
3.3.2
UNIt1
UNIt2
rESOLU
UNt.rOLL
Notes
3.3.3
A-tArE
CHAIn.tr
A.CL-tr
Scale settings (SCALE)
CAL – calibration (adjustment)
This menu item is not available for certified scales without internal calibration weight.
For scales with an internal calibration weight:
1. Unload scale.
2. Activate menu item CAL with Yes. The scale calibrates with the internal cali-
bration weight. -Int CAL- appears in the display. After calibration is
completed, -donE- appears briefly in the display, and the scale automatically
returns to weighing mode.
For scales without an internal calibration weight:
1. Unload scale.
2. Activate menu item CAL with Yes. The scale determines the zero point.
–0- appears in the display. The calibration weight to be placed on the scale then
flashes in the display.
3. If necessary, change the weight value displayed with No.
4. Place the calibration weight on the scale and confirm with Yes.
The scale calibrates with the calibration weight loaded. After calibration is com-
pleted, -donE- appears briefly in the display, and the scale automatically returns
to weighing mode.
DISPLAY – weighing unit and display accuracy
Select weighing unit 1: g, kg, oz, lb, t
Select weighing unit 2: g, kg, oz, lb, t
Select readability (resolution), model-dependent
When UNT.rOLL is switched on, the weight value can be displayed in all available
units with Units.
On certified scales, the weighing units oz and lb are displayed with the symbol *.
•
On certified scales, resolutions that deviate from the scale definition are displayed
•
without a weighing unit and with the symbol *.
On dual-range/dual interval scales, resolutions marked with |<–> 1/2| are
•
divided up into 2 weighing ranges / intervals, e.g. 2 x 3000 d.
TARE – tare function
Switching on/off automatic taring
Switching on/off chain tare
Switching on/off automatic taring with automatic clearing of the tare weight when the
load is removed from scale
Settings in the menu