  • Page 1 Sepa CF Cell - Hastelloy Assembly & Operation Manual www.sterlitech.com info@sterlitech.com 1-877-544-4420 Copyright © 2024...
  • Page 2: Table Of Contents

    Hastelloy Sepa CF Cell Components ..............Hastelloy Sepa CF Cell Assembly ............Hastelloy Sepa CF Cell System Assembly ............Operation of the Hastelloy Sepa CF Cell ............Supplementary Operating Information ..........Accessory and Spare Part Ordering Information ................ Return Material Authorization ....................
  • Page 3: Introduction

    Introduction Introduction The Hastelloy Sepa CF Cell is a laboratory-scale crossflow filtration unit designed to evaluate the performance of flat sheet membranes in a variety of applications. It simulates the flow dynamics of larger, commercially available membrane elements such as industrial spiral wound membrane elements. By using a combination of shims, feed spacers, and membranes, users can vary the operating conditions and fluid dynamics over broad ranges.
  • Page 4: Hastelloy Sepa Cf Cell Components

    Cell Components Hastelloy Sepa CF Cell Components Figure 1 illustrates a typical assembly of the Sepa CF Cell. The cell is also packaged with a length of permeate tubing and membrane cutting template. Figure 1: Hastelloy Sepa CF Cell Assembly...
  • Page 5: Hastelloy Sepa Cf Cell Assembly

    (not included) to the quick disconnect fitting on the side of the cell holder. Note: The cell holder is filled with Mobil DTE 20 Series hydraulic oil. It is recommended to use Sterlitech's hand pump to match the oil type and ensure compatibility.
  • Page 6: Hastelloy Sepa Cf Cell System Assembly

    Skid System. Alternatively, Figure 2 lists the key components required to operate a Sepa cell. These components are sold separately and can be found here. Figure 2: Cross Flow System Diagram for Hastelloy Sepa CF Cell Cell Holder Pressure Gauge...
  • Page 7: Operation Of The Hastelloy Sepa Cf Cell

    Operation of the Cell Operation of the Hastelloy Sepa CF Cell After assembling the Sepa CF Cell into the cell holder and attaching the hand pump, the cell can now be operated with the system. To operate the Sepa CF Cell: Close the hand pump relief valve knob (turn clockwise) and operate the hand pump.
  • Page 8: Supplementary Operating Information

    During CIP, cleaning solutions are re-circulated and, in some cases, allowed to sit for a period of time within the cell body. In some cases, the feed pump can be used to re- circulate the cleaning solutions. Refer to the Membrane Cleaning Blog linked above for recommended cleaning instructions. Hastelloy Sepa CF Cell www.sterlitech.com info@sterlitech.com 1-877-544-4420 Copyright ©...
  • Page 9: Accessory And Spare Part Ordering Information

    This information is not intended as a license to operate under, or a recommendation to infringe upon, any patent of Sterlitech or others covering any material or use.The foregoing may not be altered except by a written agreement signed by officers of the manufacturer.
  • Page 10 Sterlitech’s benchtop cross flow membrane test system provides the versatility required to innovate. Sterlitech’s comprehensive line of products is supported by the expertise of its technical specialists who can assist with application-specific product selection, and provide customized solutions where necessary. Unique problem-solving approaches, flexibility, and consistent quality have made Sterlitech Corporation a renowned global provider of filtration products and equipment.